ZENITH, FEMORAL STEM ZENITH, DISTAL CENTRALIZER FOMORAL HEAD ACETABULAR INSERT
K121636 · Tgm Medical, Inc. · LPH · Oct 24, 2012 · Orthopedic
Device Facts
Record ID
K121636
Device Name
ZENITH, FEMORAL STEM ZENITH, DISTAL CENTRALIZER FOMORAL HEAD ACETABULAR INSERT
Applicant
Tgm Medical, Inc.
Product Code
LPH · Orthopedic
Decision Date
Oct 24, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ZENITH Hip System is designed for total hip arthroplasty and is intended to be used with compatible components (femoral heads and acetabular cups) of the Helicon Hip System. The indications for use are: - A. Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis. - B. Revision of failed femoral head replacement, cup arthroplasty or other hip procedures. - Proximal femoral fractures. ். - D. Avascular necrosis of the femoral head. - Non-union of proximal femoral neck fractures. E. - F. Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities. The un-coated stem in the ZENITH Hip System is indicated for cemented use while the porous coated stem is indicated for non-cemented use.
Device Story
ZENITH Hip System (ZHS) is a modular hip prosthesis for total hip arthroplasty; used in clinical settings by orthopedic surgeons. System comprises femoral stems (porous-coated Titanium alloy for non-cemented use; nonporous cobalt chrome alloy for cemented use), 36mm femoral heads (CoCr alloy), and 36mm acetabular inserts (UHMWPE). Stems feature 128° neck-shaft angle, 12/14 Morse taper trunnion, and cylindrical fluted shaft. Components are compatible with Helicon Hip System (HHS) femoral heads, acetabular components, and cancellous bone screws. Device provides structural replacement for diseased or damaged hip joints; restores joint function and mobility; reduces pain associated with arthritis, fractures, or necrosis. Surgeon selects appropriate stem/head/insert configuration based on patient anatomy and surgical approach (cemented vs. non-cemented).
Clinical Evidence
Bench testing only. No clinical data provided. Evaluation included Failure Modes and Effects Analysis (FMEA) and mechanical bench testing of worst-case configurations, including distal and proximal fatigue testing, range of motion analysis, porous coating bond strength, and modular connection testing (pull-out and fretting corrosion).
Technological Characteristics
Materials: Forged Titanium (Ti) alloy, commercially pure Ti (CPTi) beads, forged cobalt chrome (CoCr) alloy, UHMWPE, PMMA (distal centralizer). Design: Straight, collarless (porous) or collared (nonporous) stems; 128° neck-shaft angle; 12/14 Morse taper trunnion; cylindrical fluted shaft. Sizes: 8-15mm. Connectivity: Modular, compatible with Helicon Hip System components. Sterilization: Not specified.
Indications for Use
Indicated for total hip arthroplasty in patients with significantly impaired joints due to rheumatoid, osteoarthritis, or post-traumatic arthritis; revision of failed hip procedures; proximal femoral fractures; avascular necrosis; non-union of proximal femoral neck fractures; congenital dysplasia; arthrodesis conversion; coxa magna/plana/vara/valga; developmental, metabolic, or tumorous conditions; osteomalacia; osteoporosis; pseudarthrosis conversion; and structural abnormalities.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{1}------------------------------------------------
KI21636.
OCT 24 2012
#### 2. 510(k) SUMMARY
| Sponsor Name: | TGM Medical, Inc.<br>5145 Golden Foothill Parkway, Suite 175<br>El Dorado Hills, CA 95762 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact: | Prakash Pai<br>Phone: (916) 358-8835<br>Cell: (774) 277-1312<br>Email: Prakash.pai63@yahoo.com |
| Date Prepared: | Revised October 23, 2012 to support Supplement 1 and 2 based<br>on FDA AI requests for K121636 (August 15th, 2012 and<br>September 26, 2012) |
| Trade Name: | ZENITH Hip System |
| Common Name: | Hip prosthesis; porous coated and nonporous coated |
| Classification Name: | Hip joint metal/polymer/metal semi-constrained porous-coated<br>uncemented prosthesis (21 CFR 888.3358, Class II device,<br>Product Code LPH). |
| | Hip joint metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis (21 CFR 888.3353, Class II<br>device Product Code LZO) |
#### Device Description:
The ZENITH Hip System (ZHS) offers a proximally porous and nonporous femoral stem component each compatible with all femoral heads, acetabular components, and cancellous bone screws of the Helicon Hip System (HHS). The porous ZHS stems are manufactured from forged Titanium (Ti) alloy, and employ a porous-coating made from commercially pure Ti (CPTi) beads. The nonporous ZHS stems are manufactured from forged cobalt chrome (CoCr) alloy. The porous Ti stem is a straight, collarless design available in 6 sizes ranging from 10 to 15mm. The nonporous CoCr stem is a straight, collared design available in 8 sizes ranging from 8 to 15mm. Both stems feature a neck-shaft angle of 128°, a 12/14 Morse taper trunnion, neck relief, and a cylindrical fluted shaft. A distal centralizer manufactured from polymethylmethacrylate (PMMA) is available for cemented applications of the CoCr stem.
The ZHS also includes 36mm femoral heads and acetabular cups. The heads are manufactured from forged or wrought CoCr alloy. The 36mm inserts are manufactured from conventional UHMWPE. The heads are highly polished and are available in four neck offsets (-5mm, Neutral, +5mm, +10mm). The heads incorporate an identical female 12/14 Morse taper used on all predicate HHS CoCr heads. The 36mm inserts are compatible with predicate HHS acetabular shells ranging in size from 56 to 68mm. Refer to the compatibility table (new
Page #1 24 3
{2}------------------------------------------------
Attachment 1 of the September 12, 2012 Al response/Supplement 1 ). The 36mm inserts are offered in neutral or 10 degree hooded configurations. Hooded inserts include an X-ray marker manufactured from wrought Ti alloy. The 36mm heads and mating inserts are also compatible with the Helicon Hip System.
#### Indications for Use:
The ZENITH Hip System is designed for total hip arthroplasty and is intended to be used with compatible components (femoral heads and acetabular cups) of the Helicon Hip System.
The indications for use are:
- A. Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis.
- B. Revision of failed femoral head replacement, cup arthroplasty or other hip procedures.
- Proximal femoral fractures. ்.
- D. Avascular necrosis of the femoral head.
- Non-union of proximal femoral neck fractures. E.
- F. Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities.
The un-coated stem in the ZENITH Hip System is indicated for cemented use while the porous coated stem is indicated for non-cemented use.
#### Substantial Equivalence:
#### Technological Characteristics/Substantial Equivalence:
Consensus Orthopedics, Inc. (COI) licensed two of their Consensus Hip System (CHS) femoral stem designs along with their CHS distal centralizer to TGM Medical for use as the ZHS femoral components. These CHS stem designs were modified from the predicate CHS stem components cleared under K935193 and K922561 to incorporate neck reliefs. The distal centralizer is identical to that cleared under K922561. All ZHS implant components are compatible with all HHS femoral heads, acetabular components, and cancellous bone screws cleared under K111472 (Consensus had licensed their TaperSet Hip to TGM for the HHS).
COI licensed the predicate CHS 36mm CoCr femoral head design cleared under K070061 and a mating 36mm UHMWPE insert design to TGM Medical for use as part of the ZHS. The ZHS 36mm heads are identical in material and geometry as the CHS 36mm CoCr heads cleared under K070061. The mating ZHS 36mm inserts are identical in geometry as the CHS 36mm inserts cleared under K070061. The ZHS 36mm inserts employ identical material and outer dome/locking mechanism geometry as the predicate HHS 32mm inserts. The ZHS 36mm head/insert components are compatible with all ZHS and HHS implant components of the appropriate size.
Because the subject ZHS stems and 36mm head/insert components employ the same material composition and strength requirements, have the same or similar technological
Page #2 28-3
{3}------------------------------------------------
## X171636
characteristics, and have similar indications as their predicate devices, the ZHS stems and 36mm head/insert components are substantially equivalent to legally marketed predicates (Table 2.1).
| 510(k)<br>Number | Trade Name | 510(k) holder | 510(k)<br>Clearance |
|------------------|--------------------------------------------------------------------|-----------------------------|---------------------|
| K922561 | Consensus Total Hip System | U.S. Medical Products, Inc. | 07/21/1993* |
| K935193 | Consensus Hip System - Porous<br>Coated Titanium Femoral Stem | U.S. Medical Products, Inc. | 08/18/1994* |
| K032396 | RingLoc 36mm Liners and Modular<br>Femoral Heads | Biomet, Inc. | 08/21/2003 |
| K070061 | 36mm CoCr Femoral Head and<br>36mm [crosslinked] Acetabular Insert | Hayes Medical, Inc. | 01/31/2007** |
| K060918 | Excia Total Hip | Aesculap, Inc. | 05/26/2006 |
| K111472 | Helicon Hip System | TGM Medical, Inc. | 09/06/2011 |
Table 2.1: Legally marketed predicates to which substantial equivalence is claimed.
Notes: * Cleared prior to the purchase of U.S. Medical Products by Hayes Medical in 1996. ** Cleared prior to the change in company name from Hayes Medical to Consensus Orthopedics in 2008.
#### Non-Clinical Performance Data:
COl provided rights to reference their 510(k)s and supporting performance testing. The ZHS components were evaluated using a Failure Modes and Effects Analysis (FMEA). Nonclinical bench testing was carried out when geometry and/or material presented a new worstcase in comparison to predicate devices.
Bench testing included distal and proximal fatigue testing of the worst-case stem/head configurations, Range of Motion analysis, porous coating bond strength, and testing of the modular connection including pull-out and fretting corrosion. All of the observed results for the porous Ti and nonporous CoCr stem designs licensed to TGM medical demonstrate that the ZHS is as safe, as effective, and performs at least as safely and effectively as legally marketed predicates.
The 36mm head/insert components are proven through the use of predicate devices using the same head size, design, and material (K070061), the same mating insert size and design (K070061), and the same mating insert material composition (K032396). Therefore, no additional testing was deemed necessary over that used in support of the predicates based on identical design and materials.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
TGM Medical, Incorporated % Mr. Prakash Pai Vice President, Global Quality and Regulatory Affairs 5145 Golden Foothill Parkway, Suite 175 El Dorado Hills, California 95762
Re: K121636
Trade/Device Name: ZENITH Hip System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, LZO
Dated: October 16, 2012 Received: October 18, 2012
Dear Mr. Pai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
OCT 2 4 2012
{5}------------------------------------------------
#### Page 2 - Mr. Prakash Pai
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
ercly yours,
N. Melkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# (171636
### 1. INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Device Name: ZENITH Hip System
#### Indications for Use:
The ZENITH Hip.System is designed for total hip arthroplasty and is intended to be used with compatible components (femoral heads and acetabular cups) of the Helicon Hip System.
The indications for use are:
- Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic A. arthritis.
- Revision of failed femoral head replacement, cup arthroplasty or other hip procedures. B.
- Proximal femoral fractures. C.
- Avascular necrosis of the femoral head. D.
- Non-union of proximal femoral neck fractures. E.
- Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, F. coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities.
The un-coated stem in the ZENITH Hip System is indicated for cemented use while the porous coated stem is indicated for non-cemented use.
Prescription Use (21 CFR Part 801 Subpart D) Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Af
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number k121636
Page 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.