HELICON HIP SYSTEM, FEMORAL STEM, HELICON HIP SYSTEM, FEMORAL HEAD, COCR, HELICON HIP SYSTEM, ACETABULAR SHELL, HELICON
Applicant
Tgm Medical
Product Code
LPH · Orthopedic
Decision Date
Sep 6, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HELICON™ Hip System is designed for total hip arthroplasty. The indications for use are: - Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis. - Revision of failed femoral head replacement, cup arthroplasty or other hip procedures. - Proximal femoral fractures. - Avascular necrosis of the femoral head. - Non-union of proximal femoral neck fractures. - Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities The HELICON™ hip stem is indicated for cementless use. The HELICON™ acetabular shell component is indicated for cemented or cementless use.
Device Story
HELICON Hip System (HHS) is a semi-constrained, porous-coated hip prosthesis for primary or revision total hip arthroplasty. Components include a monolithic titanium alloy femoral stem (dual wedge geometry, 135° neck-shaft angle, 12/14 Morse taper), CoCr femoral head, acetabular shell (titanium alloy), and UHMWPE insert. Stem features proximal CPTi plasma porous coating; acetabular shell features CPTi beaded porous coating. Titanium bone screws provide optional fixation. Used in clinical settings by orthopedic surgeons. Device replaces damaged hip joint components to restore function and mobility. Substantial equivalence based on identical design, materials, and geometry to predicate Consensus TaperSet Hip System.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing performed on the predicate TaperSet Hip System, including stem fatigue testing, range of motion analysis, porous coating characterization, polyethylene insert retention testing, and modular connection (fretting/corrosion) analysis.
Indicated for total hip arthroplasty in patients with significantly impaired joints due to rheumatoid, osteoarthritis, or post-traumatic arthritis; revision of failed hip procedures; proximal femoral fractures; avascular necrosis; non-union of femoral neck fractures; or structural/developmental conditions including dysplasia, coxa vara/valga, and metabolic/tumorous conditions.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{1}------------------------------------------------
TGM Medical, Inc Premarket Traditional 510(k) Notification: HELICON Hip System
Page 1 of 3
| 510(k) SUMMARY | | K 111472 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Sponsor Name: | TGM Medical, Inc.<br>5145 Golden Foothill Parkway, Suite 175 & 180<br>El Dorado Hills, CA 95762 | SEP - 6 20 |
| 510(k) Contact: | Dennis Crane<br>Emerson Consultants, Inc<br>12701 Whitewater Drive, Suite 120<br>Minnetonka, MN 55343<br>Phone: 952.374.6012 / Fax: 952.374.6001<br>dennisc@emersonconsultants.com | |
| Date Prepared: | August 29, 2011 | |
| Trade Name: | HELICON Hip System (HHS) | |
| Common Name: | Porous-coated hip prosthesis for cementless use | |
| Classification Name: | Hip joint metal/polymer/metal semi-constrained porous-coated<br>uncemented prosthesis (21 CFR 888.3358, Class II device,<br>Product Code LPH). | |
#### Device Description:
The HELICON Hip System (HHS) is a porous-coated, semi-constrained, hip prosthesis designed for either primary or revision hip arthroplasty. The HHS includes a femoral stem, a CoCr femoral head, non-crosslinked UHMWPE insert, acetabular shell and bone screws.
The HHS stem is a monolithic, titanium alloy, tapered hip stem available with a proximal CPTi plasma porous coating. The stem has a dual wedge geometry and is available in both standard and 7mm lateral offsets in sizes ranging from 7.5mm to 24mm. The stems feature a neck shaft angle of 135° and a 12/14 Morse taper trunnion. HHS stems are compatible with the HHS CoCr femoral head and HHS acetabular insert. The HHS CoCr head is highly polished and available in multiple offsets and diameters (22, 28, and 32mm). The HHS acetabular components include a titanium alloy shell and a mating UHMWPE insert. The shell is a hemispherical design, is available with or without screw holes, and employs a CPTi beaded, porous coating. The UHMWPE insert is available in 20° hooded or nonhooded (neutral) configurations in inner diameters of 22, 28, and 32mm. Titanium alloy bone screws are available for additional fixation.
#### Indications for Use:
The HELICON™ Hip System is designed for total hip arthroplasty.
The indications for use are:
- A. Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic arthritis.
{2}------------------------------------------------
#### TGM Medical, Inc Premarket Traditional 510(k) Notification: HELICON Hip System
Page 2 of 3
- B. Revision of failed femoral head replacement, cup arthroplasty or other hip procedures.
- C. Proximal femoral fractures.
- D. Avascular necrosis of the femoral head.
- E. Non-union of proximal femoral neck fractures.
- F. Other indications such as congenital dysplasia, arthrodesis conversion, coxa magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities
The HELICON™ hip stem is indicated for cementless use. The HELICON™ acetabular shell component is indicated for cemented or cementless use.
### Substantial Equivalence:
## Technological Characteristics/Substantial Equivalence:
The HHS is equivalent to predicate systems in design geometry, materials, performance testing and indications for use. Consensus Orthopedics, Inc. (COI) licensed their previously cleared hip components to TGM Medical, Inc. The hip components consist of the previously cleared TaperSet Hip System femoral stem (K102399), CoCr femoral heads (K922561, K960339, and K960156), UHMWPE inserts (K922561 and K020153), acetabular shell (K922561) and titanium bone screws (K922561). The HHS hip components therefore have the same technological characteristics to those used in the Consensus Taper Set Hip System. Based on the identical material, characterization data, geometry and mechanical testing, the HHS system is substantially equivalent to legally marketed predicates (Table 2.1).
| 510(k)<br>Number | Trade Name | 510(k) holder | 510(k)<br>Release Date |
|------------------|------------------------------------------|-----------------------------|------------------------|
| K922561 | Consensus Total Hip System | U.S. Medical Products, Inc. | 07/21/1993 |
| K960339 | Consensus 22mm CoCrMo Femoral<br>Head | U.S. Medical Products, Inc. | 02/21/1996 |
| K960156 | Consensus 32mm CoCrMo Femoral<br>Head | U.S. Medical Products, Inc. | 02/21/1996 |
| K960340 | Consensus Apex Dome Hole Plug | U.S. Medical Products, Inc. | 03/15/1996 |
| K102399 | Consensus TaperSet Hip System | Consensus Orthopedics, Inc. | 12/02/2010 |
| K100933 | CS2 Plus Acetabular Insert | Consensus Orthopedics, Inc. | 10/06/2010 |
| K020153 | Consensus Acetabular Shell, Ti<br>Coated | Hayes Medical, Inc. | 04/15/2002 |
Table 2.1: Legally marketed devices to which substantial equivalence is claimed:
## Non-Clinical Performance Data:
COI provided rights to reference their 510(k)s and supporting performance testing. The HELICON Hip System (HHS) implant components use the identical materials, design construct features (e.g., including locking mechanisms of the polyethylene insert to the metal shell and the exterior coatings) of the predicate components. No changes were made to the predicate components that required additional testing. The HHS implant components
+
{3}------------------------------------------------
### TGM Medical. Inc Premarket Traditional 510(k) Notification: HELICON Hip System
Page 3 of 3
-111472
were evaluated using a Failure Modes and Effects Analysis (FMEA). No additional nonclinical bench testing was deemed necessary over that used in support of the predicates based on identical design and materials.
Predicate bench testing included distal and proximal stem fatigue testing of the worst-case stem consistent with the "Guidance for Industry and FDA Staff Non-clinical Information for Femoral Stem Prostheses;" Range of Motion analysis; CPTi plasma sprayed and CPTi beaded metallic coating characterization per the "Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement;" CPTi plasma spray characterization per the "Guidance for Industry on the Testing of Metallic Plasma Sprayed Coatings on Orthopedic Implants to Support Reconsideration of Postmarket Surveillance Requirements;" polyethylene insert retention mechanism testing; and bone screw mechanical strength testing. Additionally, modular connection analyses including fretting and corrosion of the metallic femoral heads for compatibility and pin-on-flat testing of the non-crosslinked UHMWPE/CoCr material combination were also performed. Both the CPTi plasma sprayed coating and the CPTi beaded coating characterizations meet the regulatory definition of porous coating for the hip construct per 21 CFR 888.3358.
All of the observed results for the predicate TaperSet Hip System had been found substantially equivalent prior to devices currently marketed. The subject Helicon Hip System is identical in material and design to that of the TaperSet Hip System. Therefore, the subject device is as safe, as effective, and performs at least as safely and effectively as legally marketed predicates.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an emblem of an eagle with its wings spread, with three bars across its body.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
TGM Medical. Inc. % Emerson Consultants. Inc. Mr. Dennis Crane 12701 Whitewater Drive, Suite 120 Minnetonka, Minnesota 55343
SEP - 6 2011
Re: K111472
Trade/Device Name: HELICON Hip System (HSS) Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH Dated: August 3, 2011 Received: August 5, 2011
Dear Mr. Crane:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
#### Page 2 - Mr. Dennis Crane
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Eric Keith
fs Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# 1. INDICATIONS FOR USE STATEMENT
K 11472
510(k) Number (if known):
Device Name: HELICON Hip System
## Indications for Use:
The HELICON™ Hip System is designed for total hip arthroplasty.
The indications for use are:
- Significantly impaired joints resulting from rheumatoid, osteo, and post-traumatic A. arthritis.
- Revision of failed femoral head replacement, cup arthroplasty or other hip B. procedures.
- Proximal femoral fractures. C..
- D. Avascular necrosis of the femoral head.
- Non-union of proximal femoral neck fractures. E.
- Other indications such as congenital dysplasia, arthrodesis conversion, coxa F. magna, coxa plana, coxa vara, coxa valga, developmental conditions, metabolic and tumorous conditions, osteomalacia, osteoporosis, pseudarthrosis conversion, and structural abnormalities
The HELICON™ hip stem is indicated for cementless use. The HELICON™ acetabular shell component is indicated for cemented or cementless use.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ X (21 CFR Part 801 Subpart D) Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
for M Melkeem
(Division Sign Division signical (Division Sign-Dff)
Division of Surgical, Orthopedic, Orthopedic,
Controlectorerive Devices Division of Surgion in Surgices
and Restorative Devices and Restorative Devices
K111472
510(k) Number -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.