K220492 · Heraeus Medical GmbH · KWY · Mar 23, 2022 · Orthopedic
Device Facts
Record ID
K220492
Device Name
COPAL exchange G hip spacer
Applicant
Heraeus Medical GmbH
Product Code
KWY · Orthopedic
Decision Date
Mar 23, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3390
Device Class
Class 2
Attributes
Therapeutic
Intended Use
COPAL® exchange G hip spacer (polymethylmethacrylate / gentamicin) is indicated for temporary use (maximum of 180 days) as a total hip replacement (THR) in skeletally mature patients undergoing a two-stage procedure due to a septic process. The device is inserted into the femoral medullary canal and acetabular cavity following removal of the existing implant and radical debridement. The device is assigned to be used in conjunction with systemic antibiotic therapy (standard treatment approach to an infection). COPAL® exchange G hip spacer is not intended for use for more than 180 days, at which time it must be explanted, and a permanent device implanted, or another appropriate treatment performed (e.g., resection arthroplasty, fusion etc.). COPAL® exchange G hip spacer is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, canes) throughout the implantation period.
Device Story
Temporary hip spacer implant for two-stage septic endoprosthesis revision; consists of radiopaque PMMA containing gentamicin and inner stainless steel (AISI 316L) reinforcing structure; colored with chlorophyll-copper-complex (E141) for surgical visibility. Device inserted into femoral medullary canal and acetabular cavity by orthopedic surgeon following implant removal and debridement. Gentamicin released into surrounding joint fluid to reduce bacterial colonization risk. Used in conjunction with systemic antibiotic therapy and mobility assist devices. Single-use, sterile. Modification involves updated dimensions compared to predicate.
Clinical Evidence
No clinical testing conducted. Bench testing performed, including fatigue performance testing per ISO 7206-6 and endotoxin limit verification.
Technological Characteristics
Radiopaque PMMA (polymethylmethacrylate) with gentamicin; inner stainless steel (AISI 316L) reinforcement; chlorophyll-copper-complex (E141) colorant; carbonate X-ray contrast medium. Sterile, single-use. Dimensions modified from predicate. Fatigue testing per ISO 7206-6.
Indications for Use
Indicated for skeletally mature patients undergoing two-stage total hip replacement due to septic process. Requires use of mobility assist devices (crutches, canes, walkers) during implantation. Contraindicated for use exceeding 180 days.
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis is a two-part device intended to be implanted to replace the head and neck of the femur. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a snap-fit acetabular component made of an alloy, such as cobalt-chromium-molybdenum, and ultra-high molecular weight polyethylene. This generic type of device may be fixed to the bone with bone cement (§ 888.3027) or implanted by impaction.
K234028 — COPAL® exchange G hip spacer; COPAL® G+V · Heraeus Medical GmbH · Feb 15, 2024
K191016 — COPAL exchange G Hip and Knee Spacers · Heraeus Medical GmbH · Nov 1, 2019
K031841 — SPACER G TEMPORARY HIP PROSTHESIS · Exactech, Inc. · Jan 22, 2004
K252326 — InterSpace GV Hip Spacer · Tecres S.P.A. · Aug 6, 2025
K253675 — OsteoRemedies Hip Spacer System · Osteoremedies, LLC · Dec 23, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
March 23, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".
Heraeus Medical GmbH Heike Gustke Senior Manager Regulatory Affairs Philipp-Reis-Str. 8/13 Wehrheim, Hessen 61273 Germany
Re: K220492
Trade/Device Name: COPAL® exchange G hip spacer Regulation Number: 21 CFR 888.3390 Regulation Name: Hip Joint Femoral (Hemi-Hip) Metal/Polymer Cemented Or Uncemented Prosthesis Regulatory Class: Class II Product Code: KWY, KWL Dated: February 14, 2022 Received: February 22, 2022
Dear Heike Gustke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220492
Device Name COPAL® exchange G hip spacer
### Indications for Use (Describe)
COPAL® exchange G hip spacer (polymethylmethacrylate / gentamicin) is indicated for temporary use (maximum of 180 days) as a total hip replacement (THR) in skeletally mature patients undergoing a two-stage procedure due to a septic process. The device is inserted into the femoral medullary canal and acetabular cavity following removal of the existing implant and radical debridement. The device is assigned to be used in conjunction with systemic antibiotic therapy (standard treatment approach to an infection). COPAL® exchange G hip spacer is not intended for use for more than 180 days, at which time it must be explanted, and a permanent device implanted, or another appropriate treatment performed (e.g., resection arthroplasty, fusion etc.). COPAL® exchange G hip spacer is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, canes) throughout the implantation period.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Heraeus Medical GmbH. The word "Heraeus" is in a bold, sans-serif font in gray. Below it, in a smaller, black, sans-serif font, is the text "Heraeus Medical GmbH."
| 7 | 510(k) Summary |
|---|----------------|
|---|----------------|
| I. | <b>SUBMITTER</b> |
|-----------------|---------------------------------------------|
| Manufacturer: | Heraeus Medical GmbH |
| | Philipp-Reis-Str. 8/13 |
| | 61273 Wehrheim |
| | Germany |
| Contact Person: | Dr. Heike Gustke |
| | Senior Manager Regulatory Affairs |
| | Phone: +49 (6181) 35-2942 |
| | Email: heike.gustke@heraeus.com |
| | Additional Email: Regulatory_HM@heraeus.com |
| Date Prepared: | March 21, 2022 |
| II. | <b>DEVICES</b> |
|-------------------------|-----------------------------------------------------------------------------------|
| Name of Device: | COPAL® exchange G hip spacer |
| Device Common Name: | KWL: Prosthesis, Hip, Hemi-, Femoral, Metal |
| | KWY: Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented or Uncemented |
| Regulation Description: | KWL: Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis |
| | KWY: Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis |
| Regulation Number: | KWL: 21 CFR 888.3360 |
| | KWY: 21 CFR 888.3390 |
| Regulatory Class: | Class II |
| Product Code(s): | KWL, KWY |
### PREDICATE DEVICES lll.
COPAL® exchange G hip and knee spacers (K191016)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Heraeus Medical GmbH. The word "Heraeus" is in a large, bold, sans-serif font, and it is in gray. Below the word "Heraeus" is the text "Heraeus Medical GmbH" in a smaller, black, sans-serif font.
#### DEVICE DESCRIPTION IV.
A Special 510(k) submission is being supplied to the U.S. FDA to gain clearance for a modification to the existing COPAL® exchange G hip spacer previously cleared in K191016. The change includes a modified dimension of COPAL® exchange G hip spacer.
The intended use and indications for use of existing COPAL® exchange G hip spacer previously cleared in K191016 remain unchanged.
COPAL® exchange G hip spacer is a temporary hip spacer implant as part of two-stage septic endoprosthesis revision based on bone cement. COPAL® exchange G hip spacer contains gentamicin. Gentamicin reduces the risk for bacterial colonization of the device and is released into the fluid surrounding the joint. COPAL® exchange G hip spacer is intended for single-use and is supplied sterile.
COPAL® exchange G hip spacer is made of fully formed polymethylmethacrylate (radiopaque PMMA with gentamicin) and contains an inner stainless steel (AISI 316L) reinforcing structure. The mass used in the filling of the molds (the PMMA unformed resin) is prepared from a powder component and a liguid component. It contains the X-ray contrast medium carbonate. To improve visibility in the surgical field, it has been colored with chlorophyll-copper-complex (E141).
COPAL® exchange G hip spacer will be used with COPAL® exchange G hip trials (510(k) exempt devices).
#### INDICATIONS FOR USE >
COPAL® exchange G hip spacer (polymethylmethacrylate / gentamicin) is indicated for temporary use (maximum of 180 days) as a total hip replacement (THR) in skeletally mature patients underqoing a two-stage procedure due to a septic process. The device is inserted into the femoral medullary canal and acetabular cavity following removal of the existing implant and radical debridement. The device is assigned to be used in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection). COPAL® exchange G hip spacer is not intended for use for more than 180 days, at which time it must be explanted, and a permanent device implanted, or another appropriate treatment performed (e.g., resection arthroplasty, fusion etc.). COPAL® exchange G hip spacer is only indicated for patients who will consistently use traditional mobility assist devices (e.q. crutches, walkers, canes) throughout the implantation period.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE VI. DEVICE
Device comparison demonstrated that COPAL® exchange G hip spacer is substantially equivalent to the previously cleared COPAL® exchange G hip spacer (K191016) regarding intended use, indications for use, technological characteristics (design features, material, and performance) as well as operating principle.
At a high level, the subject device and predicate devices are based on the following same technological elements:
- Temporary joint prostheses
- . Preformed PMMA hip spacer containing gentamicin with an inner stainless-steel core
- Spacer design (diameter and shape)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Heraeus Medical GmbH. The word "Heraeus" is in a large, bold, sans-serif font and is colored in gray. Below the word "Heraeus" is the text "Heraeus Medical GmbH" in a smaller, bold, sans-serif font and is colored in black.
The following technological differences exist between the subject and predicate devices:
- Spacer design (length) o
#### PERFORMANCE DATA VII.
The submitter of this Special 510(k) is the manufacturer of the predicate device.
A risk analysis was performed as per DIN EN ISO 14971 to assess the impact of the modification on the device. The records of risk analysis process are retained in design history file. The evaluation demonstrated that the subject device did not present a new worst case and that the same verification and validation methods were applied to the subject device in comparison to the previously cleared predicate device (K191016). The risk analysis demonstrated that the subject device is as safe and effective as the predicate device.
The fatigue performance testing as per ISO 7206-6 was performed to address the modified dimension of COPAL® exchange G hip spacer. The same protocol as the original submission was used for collecting and assessing the data. The acceptance criteria were not altered from those used for the original device. COPAL® exchange G hip spacer meets endotoxin limits. It has been indicated that the subject device is as safe and effective as the predicate device.
#### CLINICAL TESTING VIII.
No clinical testing of COPAL® exchange G hip spacer has been conducted.
#### IX. CONCLUSION
COPAL® exchange G hip spacer is substantially equivalent to the predicate device (COPAL® exchange G hip spacer (K191016)) identified in this premarket notification
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.