The OsteoRemedies Hip Spacer System consists of modular heads and stems, and optional acetabular cups. The REMEDY® components of the OsteoRemedies Hip Spacer System include gentamicin and the REMEDY SPECTRUM® GV components include gentamicin and vancomycin. The OsteoRemedies Hip Spacer System is indicated for temporary use (maximum 180 days) as an adjunct to total hip replacement (THR) in skeletally mature patients undergoing a two-stage procedure due to a septic process and where gentamicin or gentamicin/vancomycin are the most appropriate antibiotics based on the susceptibility pattern of the infecting micro-organism(s). Following removal of the existing femoral and acetabular components and radical debridement, the head and stem components are inserted into the femoral medullary canal and can mate directly with the native acetabulum or an acetabular component which is placed in the acetabular cavity. The device is intended for use in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection). The OsteoRemedies Hip Spacer System is not intended for use for more than 180 days, at which time all components must be explanted, and a permanent device implanted or another appropriate treatment performed (e.g., resection arthroplasty, fusion, etc.).
Device Story
Modular hip spacer system consisting of femoral heads, stems, and optional acetabular cups; used as temporary implant during two-stage revision arthroplasty for septic hip infection. Components include PMMA loaded with gentamicin or gentamicin/vancomycin; stems feature stainless steel reinforcement. Device implanted by orthopedic surgeons in clinical/surgical settings following radical debridement of infected hip joint. Functions as temporary spacer to maintain joint space and deliver local antibiotics while systemic therapy is administered. Must be explanted within 180 days and replaced with permanent prosthesis. Benefits include infection management and maintenance of joint function during treatment interval.
Clinical Evidence
Bench testing only. No clinical data provided. Performance assessments included wear characteristics, antibiotic elution profiles, range of motion, and interconnection integrity. Biocompatibility evaluated per ISO 10993. Sterilization validated to SAL 10^-6 per ISO 11135/11137/11138.
Technological Characteristics
Modular hip spacer components. Materials: PMMA with gentamicin or gentamicin/vancomycin; stainless steel core for stems; CoCr modular heads. Sterilization: Validated per ISO 11135, 11137, 11138 (SAL 10^-6). Mechanical/structural design consistent with predicate hip spacers.
Indications for Use
Indicated for temporary use (max 180 days) as an adjunct to total hip replacement in skeletally mature patients undergoing two-stage revision for septic processes where gentamicin or gentamicin/vancomycin are appropriate based on infecting organism susceptibility.
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum. This generic type of device includes designs which are intended to be fixed to the bone with bone cement (§ 888.3027) as well as designs which have large window-like holes in the stem of the device and which are intended for use without bone cement. However, in these latter designs, fixation of the device is not achieved by means of bone ingrowth.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 23, 2025
OsteoRemedies, LLC
% Hollace Saas Rhodes
Vice President, Orthopedic Regulatory Affairs
MCRA, LLC
803 7th St. NW, 4th Floor
Washington, District of Columbia 20001
Re: K253675
Trade/Device Name: OsteoRemedies Hip Spacer System
Regulation Number: 21 CFR 888.3360
Regulation Name: Hip Joint Femoral (Hemi-Hip) Metallic Cemented Or Uncemented Prosthesis
Regulatory Class: Class II
Product Code: KWL, KWY
Dated: November 21, 2025
Received: November 21, 2025
Dear Hollace Saas Rhodes:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253675 - Hollace Saas Rhodes
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253675 - Hollace Saas Rhodes
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert M.
Stefani -S
Digitally signed by Robert
M. Stefani -S
Date: 2025.12.23 13:43:02
-05'00"
For: Jesse Muir, Ph.D.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253675 | ? |
| Please provide the device trade name(s). | | ? |
| OsteoRemedies Hip Spacer System | | |
| Please provide your Indications for Use below. | | ? |
| The OsteoRemedies Hip Spacer System consists of modular heads and stems, and optional acetabular cups. The REMEDY® components of the OsteoRemedies Hip Spacer System include gentamicin and the REMEDY SPECTRUM® GV components include gentamicin and vancomycin.
The OsteoRemedies Hip Spacer System is indicated for temporary use (maximum 180 days) as an adjunct to total hip replacement (THR) in skeletally mature patients undergoing a two-stage procedure due to a septic process and where gentamicin or gentamicin/vancomycin are the most appropriate antibiotics based on the susceptibility pattern of the infecting micro-organism(s).
Following removal of the existing femoral and acetabular components and radical debridement, the head and stem components are inserted into the femoral medullary canal and can mate directly with the native acetabulum or an acetabular component which is placed in the acetabular cavity. The device is intended for use in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection).
The OsteoRemedies Hip Spacer System is not intended for use for more than 180 days, at which time all components must be explanted, and a permanent device implanted or another appropriate treatment performed (e.g., resection arthroplasty, fusion, etc.). | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253675 - Page 1
# 510(k) Summary
Device Trade Name: OsteoRemedies Hip Spacer System:
- REMEDY POLY+PLUS Acetabular Cup
- REMEDY CoCr Modular Femoral Head
Manufacturer: OsteoRemedies, LLC
6800 Poplar Avenue, #120
Germantown, TN 38138
Phone: (901) 453-3141
Email: info@osteoremedies.com
Prepared by: MCRA, LLC
803 7th Street, NW, 4th Floor
Washington, DC 20001
Office: 202.552.5800
Date Prepared: December 19, 2025
Classifications: 21 CFR 888.3360, Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis
Class: II
Product Codes: KWL, KWY
Subject & Predicate Devices:
| Subject Devices | Predicate Devices |
| --- | --- |
| REMEDY POLY+PLUS Acetabular Cup | REMEDY Acetabular Cup (K192995, K191981 and K173967) |
| REMEDY CoCr Modular Femoral Head | REMEDY Hip Spacer (K112470) – Originally cleared under the trade name, 2GC Hip/Knee Modular Spacer
REMEDY SPECTRUM GV Hip Spacer (K172906) – Originally cleared under the trade name, REMEDY PLUS Hip Spacer
XS REMEDY and REMEDY SPECTRUM GV Modular Heads |
The subject devices are substantially equivalent to the previously cleared components of the OsteoRemedies Hip Spacer System with respect to intended use, materials, design, and function. The information summarized in the Design Control Activities Summary demonstrates that the subject components met the pre-determined acceptance criteria for the verification activities.
{5}
K253675 - Page 2
## Reference Device:
DePuy Prostalac Hip System (H000004, converted to a 510(k) in February 2003)
## Indications For Use:
The OsteoRemedies Hip Spacer System consists of modular heads and stems, and optional acetabular cups. The REMEDY® components of the OsteoRemedies Hip Spacer System include gentamicin and the REMEDY SPECTRUM® GV components include gentamicin and vancomycin.
The OsteoRemedies Hip Spacer System is indicated for temporary use (maximum 180 days) as an adjunct to total hip replacement (THR) in skeletally mature patients undergoing a two-stage procedure due to a septic process and where gentamicin or gentamicin/vancomycin are the most appropriate antibiotics based on the susceptibility pattern of the infecting micro-organism(s).
Following removal of the existing femoral and acetabular components and radical debridement, the head and stem components are inserted into the femoral medullary canal and can mate directly with the native acetabulum or an acetabular component which is placed in the acetabular cavity. The device is intended for use in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection).
The OsteoRemedies Hip Spacer System is not intended for use for more than 180 days, at which time all components must be explanted, and a permanent device implanted or another appropriate treatment performed (e.g., resection arthroplasty, fusion, etc.).
## Device Description:
The OsteoRemedies' REMEDY Hip Spacer and REMEDY SPECTRUM GV Hip Spacer families (K112470, K172906, K191981, K192995) are legally marketed devices that are used in skeletally mature patients undergoing a two-stage procedure due to a septic process.
The 510(k)-cleared Hip Spacer is a temporary hip spacer device consisting of independent components (femoral stem, head, and cup components). All hip spacer constructs include component(s) manufactured from PMMA with gentamicin (REMEDY Hip Spacers) or PMMA with gentamicin and vancomycin (REMEDY SPECTRUM GV Hip Spacers); the stems are structurally reinforced with a stainless steel core.
The subject REMEDY POLY+PLUS Acetabular Cup and REMEDY CoCr Modular Femoral Head Components are intended to be a line extension to the REMEDY Hip Spacer and REMEDY SPECTRUM GV Hip Spacer families (K112470, K172906, K191981, K192995).
## Performance Testing Summary:
Sterilization and Shelf-Life:
The devices are sterilized using standard methods and the sterilization cycles have been validated following international standards. The shelf life has been established through
{6}
K253675 - Page 3
stability studies. The sterilization validation complies with ISO 11135, ISO 11137, ISO 11138. Sterilization as validated with a sterility assurance level of $10^{-6}$.
## Biocompatibility
Biocompatibility evaluation has been performed to show the device materials are safe, biocompatible and suitable for their intended use. Both ISO 10993 and FDA Draft Guidance “Use of International Standard ISO 10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process” have been taken into account to evaluate the biocompatibility of the device materials.
## Performance Testing:
The following performance characteristics of the device have been assessed:
- Wear Characteristics
- Elution Profile of the System
- Range of Motion Assessment
- Interconnection Assessment
## Substantial Equivalence:
The subject devices were demonstrated to be substantially equivalent to the predicates with respect to indications, design materials, function, manufacturing, and performance.
## Conclusion:
Based on the information provided within this $510(\mathrm{k})$, the subject device is determined to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.