3-PIECE PROXIMAL HUMERAL REPLACEMENT SYSTEM

K020045 · Biomet Orthopedics, Inc. · KWT · Jan 30, 2002 · Orthopedic

Device Facts

Record IDK020045
Device Name3-PIECE PROXIMAL HUMERAL REPLACEMENT SYSTEM
ApplicantBiomet Orthopedics, Inc.
Product CodeKWT · Orthopedic
Decision DateJan 30, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3650
Device ClassClass 2
AttributesTherapeutic

Intended Use

The indications for use for the 3-Piece Proximal Humeral Replacement System include noninflamatory degenerative joint disease including osteoarthritis and avascular necrosis, theumatoid arthritis, correction of severe functional deformity, revision procedures where other treatments or devices have failed, treatment of acute fractures with humeral head involvement which are unmanageable using other treatment methods and oncology applications. The 3-Piece Proximal Humeral Replacement System is for use with bone cement.

Device Story

Total shoulder prosthesis humeral component; used for severe proximal bone loss due to tumor, trauma, or revision. System consists of modular proximal body, segment, and stem components. Distal stem cemented into remaining humeral bone; proximal body/segment reside above natural bone. Used by orthopedic surgeons in clinical/surgical settings. Provides structural replacement for humeral head to restore joint function.

Clinical Evidence

No clinical testing was provided as a basis for substantial equivalence; bench testing only.

Technological Characteristics

Modular shoulder prosthesis; metal/polymer construction; cemented fixation. Components include proximal body, segment, and stem. Designed for non-constrained use per 21 CFR 888.3650.

Indications for Use

Indicated for patients with non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, severe functional deformity, failed prior treatments/devices, acute humeral head fractures, and oncology applications requiring proximal humeral replacement.

Regulatory Classification

Identification

A shoulder joint metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's: (i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” (ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” (iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,” (iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and (v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,” (2) International Organization for Standardization's (ISO): (i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,” (ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,” (iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,” (iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,” (v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,” (vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and (vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and (3) American Society for Testing and Materials': (i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,” (ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,” (iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,” (iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,” (v) F 1108-97 “Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,” (vi) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings,” (vii) F 1378-97 “Specification for Shoulder Prosthesis,” and (viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K020045 page 1 of 2 JAN 3 0 2002 UARTERS CORPORATE ## SUMMARY OF SAFETY AND EFFECTIVENESS | Applicant or Sponsor: | Biomet Orthopedics, Inc.<br>56 East Bell Drive<br>Warsaw, IN 46581-0587 | |-----------------------|-------------------------------------------------------------------------| | Contact Person: | Dalene T. Binkley<br>Telephone: (219) 267-6639 | | Proprietary Name: | 3-piece Proximal Humeral Replacement System | | Common Name: | Shoulder prosthesis | Classification: Prosthesis, shoulder, non-constrained, metal/polymer cemented (21 CFR 888.3650) Device Classification: Class II Legally Marketed Device to which Substantially Equivalence is Claimed: The 3piece Proximal Humeral Replacement System has been developed by modifying the Proximal Humeral Replacement System (K925613). Device Description: The 3-piece Proximal Humeral Replacement System is a total shoulder prosthesis humeral component in cases where there has been severe proximal bone loss due to a tumor, trauma, or previous device placement. The modified proximal humeral replacement device can be a combination of a proximal body and stem or proximal body, segment, and stem. The device is designed so that the distal portion of the stem is cemented into the remaining humeral bone with the stem collar, segment, and proximal body residing above the natural bone. Indications for Use: The indications for use for the 3-Piece Proximal Humeral Replacement System include noninflamatory degenerative joint disease including osteoarthritis and avascular necrosis, theumatoid arthritis, correction of severe functional deformity, revision procedures where other treatments or devices have failed, treatment of acute fractures with humeral head involvement which are unmanageable using other treatment methods and oncology applications. The 3-Piece Proximal Humeral Replacement System is for use with bone cement. MAILING ADDRESS P.O. Box 587 Warsaw, IN 46581-0587 SHIPPING ADDRESS 56 E. Bell Drive Warsaw, IN 46582 nanan OFFICE 219.267.6639 FAX = E-MAIL E = {1}------------------------------------------------ Summary of Technologies: The 3-piece Proximal Humeral Replacement System Productions of indications of indications ore similar or identical to Summary of Technologies: "The 5-picce Froximal Prailers - " components-the materials, design, sizing, and indications are similar or identical to the predicate devices. Non-Clinical Testing: Mechanical Testing with Engineering Justifications determined Non-Clinical Testing: Mechanical Testing with Englacements presented no new that the 3-piece Proximal Humeral Replacement to the reading e device that the 3-piece Proximal Humoral Repliceae in the predicate device. Clinical Testing: No clinical testing was provided as a basis for substantial equivalence. page 2 of 2 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a stylized caduceus symbol. The caduceus is represented by a staff entwined with a snake, symbolizing medicine and health. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 3 0 2002 Ms. Dalene T. Binkley Regulatory Affairs Specialist Biomet Orthopedics, Inc P.O. Box 587 Warsaw, Indiana 46581 Re: K020045 Trade/Device Name: 3-Piece Proximal Humeral Replacement System Regulation Number: 21 CFR 888.3650 Regulation Name: Prosthesis, shoulder, non-constrained, metal/polymer cemented Regulatory Class: II Product Code: KWT Dated: December 19, 2001 Received: January 7, 2002 Dear Ms. Binkley: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {3}------------------------------------------------ Page 2 -- If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and II you desire specific acvice for your do rise career, please contact the Office of Compliance at additionally 807.10 for m viro diagnosions on the promotion and advertising of your device, (301) 594-4037. Additionally, for questions of (301) 594-4639. Also, please note the cegulation prease contact the Office of Comphanee at (501) of CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small information on your responsionates ander and in 11000 638-2041 or at (301) 443-6597, or at its Manufacturers "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page of 510 (k) NUMBER (IF KNOWN): `k020045 DEVICE NAME: 3-Piece Proximal Humeral Replacement System INDICATIONS FOR USE: The indications for use for the 3-Piece Proximal Humeral Replacement System include The indications for use for the 3 1 rosease including osteoarthritis and avascular necrosis, nomillianalory degencrative John abouse finational deformity, revision procedures where rheumatoid arthritis, correction of severe functional deformity, which humars' head Theumatord arthirtis, corroction of so drivent of acute fractures with humeral head other treatments of doviets nave kast, sing other treatment methods and oncology applications. This device is for use with bone cement. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.) | Concurrence of CDRH, Office of Device Evaluation (ODE) | | | | |---------------------------------------------------------------------------------------------------------------------------------------------|---------|--------------------------------------------------|--| | Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter-Use<br>(Optional Format 1-2-96) | | | <div> <img alt="Checkmark" src="prescription_use_checkmark.png"/> </div> | | | | | <div> <img alt="Signature" src="signature.png"/> </div> (Division Sign-Off)<br>Division of General, Restorative<br>and Neurological Devices | | | | | 510(k) Number | K020045 | | |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...