K122910 · Nu Vasive, Incorporated · KWQ · Oct 11, 2012 · Orthopedic
Device Facts
Record ID
K122910
Device Name
ARCHON ANTERIOR CERVICAL PLATE SYSTE,
Applicant
Nu Vasive, Incorporated
Product Code
KWQ · Orthopedic
Decision Date
Oct 11, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3060
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The NuVasive Archon Anterior Cervical Plate System is intended for anterior screw fixation of the cervical spine. These implants have been designed to provide stabilization as an adjunct to cervical fusion. Indications for the use of this implant system include degenerative disc disease defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma, spinal stenosis, deformity, tumor, pseudarthrosis or failed previous fusion.
Device Story
Anterior cervical plating system consisting of titanium alloy plates and screws; provides spinal stabilization as an adjunct to cervical fusion. Used by surgeons in clinical settings to address pathologies including degenerative disc disease, trauma, and deformity. Implants are selected based on patient anatomy and pathology. System provides mechanical fixation to the anterior cervical spine; not for posterior pedicle fixation. Benefits include stabilization of the spinal column to facilitate fusion.
Clinical Evidence
Bench testing only. Static and dynamic compression and torsion testing performed per ASTM F1717. Results demonstrate the device meets or exceeds the performance of predicate devices.
Technological Characteristics
Materials: Titanium alloy (Ti-6Al-4V ELI) per ASTM F136/ISO 5832-3; optional lock screw of nickel-cobalt-chrome-molybdenum (MP35N) per ASTM F562. System includes plates, screws, and manual surgical instruments. Mechanical fixation principle. No software or electronic components.
Indications for Use
Indicated for anterior screw fixation of the cervical spine as an adjunct to fusion in patients with degenerative disc disease (neck pain of discogenic origin confirmed by history/radiography), spondylolisthesis, trauma, spinal stenosis, deformity, tumor, pseudarthrosis, or failed previous fusion. Contraindicated for posterior element (pedicle) fixation in cervical, thoracic, or lumbar spine.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
K032815 — PRECISION SURGERY LIMITED VARIABLE AND FIXED CERVICAL PLATE SYSTEM · Precision Surgery Limited · Feb 12, 2004
K181562 — Curiteva Anterior Cervical Plate System · Curiteva, LLC · Aug 2, 2018
K220334 — MONET Anterior Cervical Plate System · Ctl Medical Corporation · Nov 9, 2022
K202972 — Anterior Cervical Plate System · Nutech Spine and Biologics · Nov 24, 2020
K060491 — QUANTUM ANTERIOR CERVICAL PLATE SYSTEM · Quantum Orthopedics, Inc. · Mar 13, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
OCT 1 1 2012
## 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
#### Submitted by: A.
Cynthia Adams Regulatory Affairs Associate NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 (858) 909-1800
Date Prepared: October 9, 2012
#### B. Device Name
| Trade or Proprietary Name: | NuVasive® Archon Anterior Cervical Plate System |
|----------------------------|-------------------------------------------------|
| Common or Usual Name: | Cervical Plate and Screw System |
| Classification Name: | Spinal Intervertebral Body Fixation Orthosis |
| Device Class: | Class II |
|-----------------|------------------|
| Classification: | 21 CFR §888.3060 |
| Product Code: | KWQ |
#### Predicate Devices C.
The subject NuVasive Archon Anterior Cervical Plate System is substantially equivalent to the following devices:
- . K053581 - NuVasive, Inc. SmartPlate Gradient Plus System
- K071329 - NuVasive, Inc. Helix Anterior Cervical Plating System
- K000486 Aesculap ABC Cervical Plating System .
#### D. Device Description
The NuVasive Archon Anterior Cervical Plate System is an anterior cervical plating system that consists of a variety of implant components including screws and plates, as well as associated manual general surgical instruments. The implants are available in a variety of different shapes and sizes to suit the individual pathology and anatomical conditions of the patient. The subject device components are manufactured from titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F136 or ISO 5832-3, with a nickel-cobalt-chrome-molybdenum (MP35N) lock screw option conforming to ASTM F562.
#### E. Intended Use
The NuVasive Archon Anterior Cervical Plate System is intended for anterior screw fixation of the cervical spine. These implants have been designed to provide stabilization as an adjunct to cervical fusion. Indications for the use of this implant system include degenerative disc disease defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma, spinal stenosis, deformity, tumor, pseudarthrosis or failed previous fusion.
Page 1 of 2
{1}------------------------------------------------
KI22910
Image /page/1/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf-like shape on the left, followed by the word "NUVASIVE" in a simple sans-serif font. Below the word "NUVASIVE" is the tagline "Speed of Innovation" in a smaller, italicized font. The logo is black and white.
WARNING: The NuVasive Archon Anterior Cervical Plate System is not intended for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic or lumbar spine.
#### F. Technological Characteristics
As was established in this submission, the subject Archon Anterior Cervical Plate System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, function, and range of sizes.
#### G. Performance Data
Nonclinical testing was performed to demonstrate that the subject Archon Anterior Cervical Plate System is substantially equivalent to other predicate devices. The following testing was performed:
- Static and dynamic compression testing per ASTM F1717
- Static torsion testing per ASTM F1717 ●
The results of these studies show that the subject Archon Anterior Cervical Plate System meets or exceeds the performance of the predicate device, and the device was therefore found to be substantially equivalent.
#### H. Conclusions
Based on the indications for use, technological characteristics, performance testing, and comparison to predicate devices, the subject Archon Anterior Cervical Plate System has been shown to be substantially equivalent to legally marketed predicate devices, and safe and effective for its intended use.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OCT 11 2012
Nu Vasive, Incorporated % Ms. Cynthia Adams Regulatory Affairs Associate 7475 Lusk Boulevard San Diego, California 92121
Re: K122910
Trade/Device Name: NuVasive® Archon Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: September 20, 2012 Received: September 21, 2012
Dear Ms. Adams:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Cynthia Adams
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Evin I. Keith
o Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# K122910
### Indications for Use
510(k) Number (if known):
Device Name: NuVasive® Archon Anterior Cervical Plate System
Indications For Use:
The NuVasive Archon Anterior Cervical Plate System is intended for anterior screw fixation of the cervical spine. These implants have been designed to provide stabilization as an adjunct to cervical fusion. Indications for the use of this implant system include degenerative disc disease defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma, spinal stenosis, deformity, tumor, pseudarthrosis or failed previous fusion.
WARNING: The NuVasive Archon Anterior Cervical Plate System is not intended for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic or lumbar spine.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Col-bal
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number k122910
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.