K113346 · Ulrich GmbH & Co. KG · KWP · Feb 28, 2012 · Orthopedic
Device Facts
Record ID
K113346
Device Name
NEON SYSTEM
Applicant
Ulrich GmbH & Co. KG
Product Code
KWP · Orthopedic
Decision Date
Feb 28, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When intended to provide immobilization and stabilization as an adjunct to fusion of the cervical spine and the occipito-cervico-thoracic junction (occiput-T3), the plate/rod, rod, hook, screw, connector and crosslink components of neon™ are indicated for the treatment of: • DDD (neck pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies), • spondylolisthesis, spinal stenosis, fracture/dislocation, atlantoaxial fracture with instability, occipitocervical dislocation, revision of previous cervical spine surgery, and tumors The cortical (3.5mm) bone screws are limited to occipital fixation only. The hook, rod and crosslink component placement is indicated for the cervical/upper thoracic (C1-T3) spine. The use of neon screws (with easy fits), polyaxial connectors (with inlays and optional spacers) is limited to placement in the upper thoracic spine (T1-T3) for anchoring of the system. Screws are not intended to be placed in the cervical spine.
Device Story
Modular posterior spinal fixation system; used for surgical stabilization of occipito-cervico-thoracic regions. Components include longitudinal members (hybrid plate/rods, straight rods), anchors (self-drilling/self-tapping screws, hooks), and interconnecting devices (anchor-to-rod connectors, rod-to-rod crosslinks). Operated by surgeons in clinical/OR settings. Provides mechanical immobilization to facilitate spinal fusion. Benefits patients by stabilizing spine segments affected by degenerative disease, trauma, or tumors.
Clinical Evidence
Bench testing only. Mechanical testing of worst-case constructs included static and dynamic compression bending per ASTM F1717, and static and dynamic compression bending and torsion per ASTM F2076. Results demonstrate performance equivalent or superior to predicate devices.
Technological Characteristics
Materials: Titanium (Ti Grade 2, ASTM F67) or titanium alloy (Ti-6Al-4V, ASTM F136). Design: Modular rod-based system with screw/hook anchors, connectors, and crosslinks. Dimensions comparable to predicate systems. No software or energy source.
Indications for Use
Indicated for patients requiring immobilization/stabilization as adjunct to fusion of cervical spine and occipito-cervico-thoracic junction (occiput-T3) for DDD, spondylolisthesis, spinal stenosis, fracture/dislocation, atlantoaxial fracture with instability, occipitocervical dislocation, revision of previous cervical spine surgery, or tumors. Cortical screws limited to occipital fixation; other screws limited to T1-T3.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
{0}------------------------------------------------
K113346
FEB 2 8 2012
Image /page/0/Picture/2 description: The image shows the logo for Ulrich medical USA. The word "Ulrich" is in a bold, sans-serif font, with the "U" being partially obscured by a black square. Below "Ulrich" is the text "medical USA" in a smaller, sans-serif font.
## 510(k) Summary
ું
| Date: | 10 November 2011 |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | ulrich GmbH & Co. KG<br>Buchbrunnenweg 12<br>89081 Ulm<br>Germany<br>Phone: +49 (0) 731-9654-1304<br>Fax: +49 (0) 731-9654-2802 |
| Contact Person: | Hans Stover<br>ulrich medical USA, Inc.<br>612 Trade Center Blvd.<br>Chesterfield, MO 63005<br>(636) 519-0268 Office<br>(636) 519-0271 Fax |
| Proposed Trade<br>Name: | neon™ |
| Device Classification | Class II |
| Classification Name: | Spinal interlaminal fixation orthosis |
| Regulation: | 888.3050 |
| Device Product<br>Code: | KWP |
| Device Description: | neon™ is a modular, posterior system used for the surgical<br>stabilization and fixation of the occipito-cervico-thoracic regions of<br>the spine. If necessary, the occipito-cervical junction as well as the<br>cervico-thoracic junction may be included. The system components<br>include longitudinal members such as hybrid plate/rods and straight<br>rods, anchors such as self-drilling and self-tapping screws and<br>hooks, and interconnecting devices such as anchor-to-rod<br>connectors and a rod-to-rod crosslink. |
| Intended Use: | When intended to provide immobilization and stabilization as an<br>adjunct to fusion of the cervical spine and the occipito-cervico-<br>thoracic junction (occiput-T3), the plate/rod, rod, hook, screw,<br>connector and crosslink components of neon™ are indicated for the<br>treatment of:<br>• DDD (neck pain of discogenic origin with degeneration of the<br>disc as confirmed by patient history and radiographic<br>studies),<br>• spondylolisthesis, |
| | spinal stenosis, fracture/dislocation, atlantoaxial fracture with instability, occipitocervical dislocation, revision of previous cervical spine surgery, and tumors |
| | The cortical (3.5mm) bone screws are limited to occipital fixation only. |
| | The hook, rod and crosslink component placement is indicated for the cervical/upper thoracic (C1-T3) spine. |
| | The use of neon screws (with easy fits), polyaxial connectors (with inlays and optional spacers) is limited to placement in the upper thoracic spine (T1-T3) for anchoring of the system. Screws are not intended to be placed in the cervical spine. |
| Materials: | neonTM components are manufactured from titanium (Ti Grade 2, ASTM F67) or titanium alloy (Ti-6Al-4V) as described by ASTM F136: |
| Predicate Devices: | Ascent POCT System (K033480 and K073420)<br>Nex-Link OCT and Cervicothoracic Spinal System (K031985, K052566, K062505 and K090060)<br>CerviFix® System (K030377 and K011966) |
| Technological<br>Characteristics: | neonTM possesses the same technological characteristics as one or more of the predicate devices. These include: intended use (as described above) basic design (rod-based having screw and/or hook anchors), material (titanium / titanium alloy), sizes (dimensions are comparable to those offered by the predicate systems) and The fundamental scientific technology of neonTM is the same as previously cleared devices. |
| Performance Data: | Mechanical testing of the worst case neonTM constructs included static and dynamic compression bending according to ASTM F1717, and static and dynamic compression bending and torsion according to ASTM F2076.<br>The mechanical test results demonstrate that neonTM performs as well as or better than the predicate devices and therefore that the device is as safe and as effective as the predicates. |
sheet I of 2
Page 8
{1}------------------------------------------------
sheet 2 of 2
:
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ulrich GmbH & Co., KG % Ulrich Medical USA. Inc. Mr. Hans Stover 612 Trade Center Boulevard Chesterfield, Missouri 63005
FEB 2 8 2012
Re: K113346 Trade/Device Name: neon™ Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: II Product Code: KWP Dated: February 11, 2012 Received: February 14, 2012
Dear Mr. Stover:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
Page 2 - Mr. Hans Stover
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.html
Sincerely yours.
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## K113346
## Indications for Use Statement
510(k) Number:
Device Name: neon™
Indications for Use:
When intended to provide immobilization and stabilization as an adjunct to fusion of the cervical spine and the occipito-cervico-thoracic junction (occiput-T3), the plate/rod, rod, hook, screw, connector and crosslink components of neon™ are indicated for the treatment of
- DDD (neck pain of discogenic origin with degeneration of the disc as confirmed by . patient history and radiographic studies),
- spondylolisthesis, ●
- spinal stenosis, ●
- fracture/dislocation, .
- atlantoaxial fracture with instability, .
- occipitocervical dislocation, .
- revision of previous cervical spine surgery, and .
- tumors ●
The cortical (3.5mm) bone screws are limited to occipital fixation only.
The hook, rod and crosslink component placement is indicated for the cervical/upper thoracic (C1-T3) spine.
The use of neon screws (with easy fits), polyaxial connectors (with inlays and optional spacers) is limited to placement in the upper thoracic spine (T1-T3) for anchoring of the system. Screws are not intended to be placed in the cervical spine.
| Prescription Use X | AND/OR |
|------------------------|--------|
| (21 CFR 801 Subpart D) | |
Over-the-Counter Use_
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K113346
Page 6
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.