NEXLINK OCT SYSTEM, MODEL 7000 SERIES

K090060 · Abbott Spine, Inc. · KWP · Jun 2, 2009 · Orthopedic

Device Facts

Record IDK090060
Device NameNEXLINK OCT SYSTEM, MODEL 7000 SERIES
ApplicantAbbott Spine, Inc.
Product CodeKWP · Orthopedic
Decision DateJun 2, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3050
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Nex-Link OCT Occipital Cervical Plating System is intended to provide stabilization as an adjunct to fusion of the cervical spine and occipito-cervico-thoracic junction (occiput-T3) for the following indications: degenerative disc disease (neck pain of discogenic origin with degeneration of the disk as confirmed by patient history and radiological studies), spondylolisthesis, spinal stenosis, fracture/dislocation, atlanto-axial fracture with instability, occipito-cervical dislocation, revision of previous cervical spine fusion surgery and tumors. The Cancellous and Cortical Bone Screws (3.5mm and 4mm diameters; 6mm-20mm threaded lengths) are used with the NexLink OCT Occipital Cervical Plating System to allow for occipital fixation and limited to occipital fixation only. The 4mm Cannulated Side Loading Closed Screws are limited to placement in the upper thoracic spine (Tt-T3) for additional stabilization of the cervical spine for the indications specified above.

Device Story

NexLink OCT is a modular occipital cervical plating system; redesign of OctaFix system. Components include modular occipital plates, precontoured rods, hooks, and bone screws. Used by surgeons in clinical settings to provide temporary stabilization of the spine (occiput-T3) during spinal fusion procedures. Bone screws facilitate occipital fixation; cannulated side-loading screws provide additional stabilization in the upper thoracic spine (T1-T3). Device provides mechanical support to stabilize the spine, promoting solid fusion and patient recovery from degenerative disease, trauma, or tumor pathology.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Modular occipital plating system; includes precontoured rods, hooks, and bone screws (cancellous/cortical 3.5mm/4mm; cannulated 4mm). Metallic spinal implant system. Mechanical fixation principle. No software or energy source.

Indications for Use

Indicated for patients requiring stabilization as an adjunct to fusion of the cervical spine and occipito-cervico-thoracic junction (occiput-T3) due to degenerative disc disease, spondylolisthesis, spinal stenosis, fracture/dislocation, atlanto-axial fracture with instability, occipito-cervical dislocation, revision of previous fusion, or tumors.

Regulatory Classification

Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K090060 Page 1 of 2 ### ABBOTT SPINE, INC. SUMMARY OF SAFETY AND EFFECTIVENESS | SUBMITTER: | Abbott Spine, Inc. | |---------------------------------------|------------------------------------------------------------------------------------------------------------| | ESTABLISHMENT REGISTRATION<br>NUMBER: | 1649384 | | CONTACT PERSON: | David Padgett<br>Senior Specialist, Regulatory Affairs<br>Telephone: (512) 533-1998<br>Fax: (512) 258-0995 | | DATE: | 02 June 2009 | | TRADE NAME: | NEXLINK OCT® Cervical Plating System | | PRODUCT CODE: | KWP | | CLASSIFICATION NAME: | OCCIPITAL CERVICAL PLATING SYSTEM | | CLASSIFICATION REFERENCE: | 21 CFR § 888.3050 | | PREDICATE DEVICE: | OctaFix® Occipital Cervical Fixation System, K021009<br>Abbott Spine, cleared 18 June 2002. | ## Device Description: NexLink OCT is the re-design and modernization of the existing OctaFix svstem. The scope of the NextLink OCT Cervical Plating System will include 1) a modular occipital plate, 2) a selection of precontoured rods, 3) hooks and 4) a refined set of instrumentation. The scope of the project is for Next ink OCT to integrate as an option to the current Nex-Link System. The NexLink OCT Occipital Cervical Plating System components are temporary implants that are used to stabilize the spine (occiput-T3) during the development of a solid spinal fusion in patients with degenerative disease, trauma (including fractures), and tumor pathology. ## Indications: The NexLink OCT Occipital Cervical Plating System is intended to provide stabilization as an adjunct to fusion of the cervical spine and occipito-cervico-thoracic junction (occiput-T3) for the following indications: degenerative disc disease (neck pain of discogenic origin with degeneration of the disk as confirmed by patient history and radiological studies), spinal stenosis. fracture/dislocation, atlantoaxial fracture with instability, occipito-cervical dislocation, revision of previous cervical spine fusion surgery and tumors. The Cancellous and Cortical Bone Screws (3.5mm and 4mm diameters; 6mm-20mm threaded lengths) are used with the NexLink OCT Occipital Cervical Plating System to allow for occipital fixation and limited to occipital fixation only. The 4mm Cannulated Side Loading Closed Screws are limited to placement in the upper thoracic spine (T1-T3) for additional stabilization of the indications specified above. {2}------------------------------------------------ K090060 Page 2 of 2 Comparison to Predicate Device: PERFORMANCE DATA (NONCLINICAL AND/OR CLINICAL): The subject device is the result of modifications to the existing Occipital Cervical Plating System. The subject device has the same intended use and is substantially equivalent to the aforementioned predicate device. Non-Clinical Performance and Conclusions: Laboratory and bench testing results demonstrate that the proposed device is substantially equivalent to the predicate device. Clinical Performance and Conclusions: Clinical data and conclusions were not needed for this device. {3}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". Inside the circle is a stylized eagle emblem with three wing-like shapes. The eagle's head is facing left. #### Public Health Service ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Abbott Spine, Inc. c/o Mr. David Padgett Senior Specialist, Regulatory Affairs 5301 Riata Park Ct., Bldg F Austin, Texas 78727 JUN - 2 2009 Re: K090060 Trade/Device Name: NexLink OCT Cervical Plating System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: KWP Dated: May 26, 2009 Received: May 27, 2009 Dear Mr. Padgett: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/cdrh/comp/ for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark M. Milliman Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K090060 Device Name: NEX-LINK OCT® Cervical Plating System Indications for Use: The Nex-Link OCT Occipital Cervical Plating System is intended to provide stabilization as an adjunct to fusion of the cervical spine and occipito-cervico-thoracic junction (occiput-T3) for the following indications: degenerative disc disease (neck pain of discogenic origin with degeneration of the disk as confirmed by patient history and radiological studies), spondylolisthesis, spinal stenosis, fracture/dislocation, atlanto-axial fracture with instability, occipito-cervical dislocation, revision of previous cervical spine fusion surgery and tumors. The Cancellous and Cortical Bone Screws (3.5mm and 4mm diameters; 6mm-20mm threaded lengths) are used with the NexLink OCT Occipital Cervical Plating System to allow for occipital fixation and limited to occipital fixation only. The 4mm Cannulated Side Loading Closed Screws are limited to placement in the upper thoracic spine (Tt-T3) for additional stabilization of the cervical spine for the indications specified above. Prescription Use _X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (Please do not write below this line - Continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE) Page I of I (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K090066
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...