The OsteoRemedies Hip Spacer System consists of modular heads and stems, and optional acetabular cups. The REMEDY components of the OsteoRemedies Hip Spacer System include gentamicin and the REMEDY SPECTRUM GV components include gentamicin and vancomycin. The OsteoRemedies Hip Spacer System is indicated for temporary use (maximum 180 days) as an adjunct to total hip replacement (THR) in skeletally mature patients undergoing a two-stage procedure due to a septic process and where gentamicin or gentamicin or gentamicin vancomycin are the most appropriate antibiotics based on the infecting micro-organism(s). Following removal of the existing femoral and acetabular components and radical debridement, the head and stem components are inserted into the femoral medulary canal and can mate directly with the native actabular component which is placed in the acetabular cavity. The device is intended for use in conjunction with systemic antibiotic therapy (standard treatment approach to an infection). The OsteoRemedies Hip Spacer System is not intended for use than 180 days, at which time all components must be explanted, and a permanent device implanted or another appropriate treatment performed (e.g., resection arthroplasty, fusion, etc.).
Device Story
Modular temporary hip spacer system; includes femoral stems, femoral heads, and acetabular cups. Components composed of PMMA impregnated with gentamicin (REMEDY) or gentamicin and vancomycin (REMEDY SPECTRUM GV). Used by orthopedic surgeons in clinical settings during two-stage revision hip arthroplasty. Device inserted into femoral medullary canal following radical debridement of infected joint. Provides mechanical support, maintains joint space, allows partial weight-bearing, and preserves soft tissue. Antibiotics elute from PMMA to inhibit bacterial adhesion. Temporary implant; must be explanted within 180 days and replaced with permanent prosthesis. Benefits include infection management, maintenance of limb length/function, and facilitation of subsequent definitive reconstruction.
Clinical Evidence
Bench testing only. Performance verified via ISO 7206-4 (stem fatigue), ISO 7206-6 (neck fatigue), and ISO 7206-13 (head/stem disassembly). Additional testing included chemical/physical properties of cement, antibiotic elution kinetics, biocompatibility, sterilization, pyrogenicity, bacterial endotoxin, and shelf life.
Technological Characteristics
Modular hip spacer components made of PMMA with gentamicin or gentamicin/vancomycin. Mechanical function via structural support and antibiotic elution. Standards: ISO 7206-4, ISO 7206-6, ISO 7206-13. Sterile, single-use. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients undergoing two-stage total hip replacement due to septic process. Requires systemic antibiotic therapy. Contraindicated for use exceeding 180 days.
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum. This generic type of device includes designs which are intended to be fixed to the bone with bone cement (§ 888.3027) as well as designs which have large window-like holes in the stem of the device and which are intended for use without bone cement. However, in these latter designs, fixation of the device is not achieved by means of bone ingrowth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 10, 2020
OsteoRemedies, LLC % Hollace Rhodes Vice President, Orthopedic Regulatory Affairs MCRA. LLC 1050 K Street NW Suite 1000 Washington, District of Columbia 20001
Re: K192995
Trade/Device Name: OsteoRemedies Hip Spacer System Regulation Number: 21 CFR 888.3360 Regulation Name: Hip Joint Femoral (Hemi-Hip) Metallic Cemented Or Uncemented Prosthesis Regulatory Class: Class II Product Code: KWL. KWY Dated: December 10, 2019 Received: December 11, 2019
Dear Ms. Rhodes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laura C. Rose, Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
### 510(k) Number (if known) K192995
#### Device Name
OsteoRemedies Hip Spacer System
#### Indications for Use (Describe)
The OsteoRemedies Hip Spacer System consists of modular heads and stems, and optional acetabular cups. The REMEDY components of the OsteoRemedies Hip Spacer System include gentamicin and the REMEDY SPECTRUM GV components include gentamicin and vancomycin.
The OsteoRemedies Hip Spacer System is indicated for temporary use (maximum 180 days) as an adjunct to total hip replacement (THR) in skeletally mature patients undergoing a two-stage procedure due to a septic process and where gentamicin or gentamicin or gentamicin vancomycin are the most appropriate antibiotics based on the infecting micro-organism(s).
Following removal of the existing femoral and acetabular components and radical debridement, the head and stem components are inserted into the femoral medulary canal and can mate directly with the native actabular component which is placed in the acetabular cavity. The device is intended for use in conjunction with systemic antibiotic therapy (standard treatment approach to an infection).
The OsteoRemedies Hip Spacer System is not intended for use than 180 days, at which time all components must be explanted, and a permanent device implanted or another appropriate treatment performed (e.g., resection arthroplasty, fusion, etc.).
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Submitter: | OsteoRemedies, LLC<br>6800 Poplar Avenue, #120<br>Memphis, TN 38138<br>Phone: (901) 453-3141<br>Email: info@osteoremedies.com |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Correspondent: | Ms. Hollace Saas Rhodes<br>Vice President, Orthopedic Regulatory Affairs<br>MCRA, LLC<br>1050 K Street NW, Suite 1000<br>Washington, DC 20001<br>Phone: (202) 552-5807<br>Email: hrhodes@mcra.com |
| Date Prepared: | January 10, 2020 |
| Trade Names: | OsteoRemedies Hip Spacer System:<br>- REMEDY® Modular Head 40mm,<br>- REMEDY SPECTRUM® GV Modular Femoral Head – 40mm,<br>- REMEDY® Acetabular Cup 40mm ID/48mm OD |
| Common Name: | Temporary femoral head and acetabular cup spacer with antibiotic |
| Classification: | Class II |
| Regulation: | 21 CFR 888.3360, Hip joint femoral (hemi-hip) metallic cemented<br>or uncemented prosthesis |
| Product Codes: | KWL KWY |
### Subject & Predicate Devices:
| Subject Devices | Predicate Devices |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------|
| XS REMEDY Modular Head | REMEDY Hip Spacer (K112470) – Originally cleared under the<br>trade name, 2GC Hip/Knee Modular Spacer |
| XS REMEDY SPECTRUM GV<br>Modular Head | REMEDY SPECTRUM GV Hip Spacer (K172906) – Originally<br>cleared under the trade name, REMEDY PLUS Hip Spacer |
| XS REMEDY Acetabular Cup | REMEDY Acetabular Cup (K191981 and K173967) |
The subject devices are substantially equivalent to the previously cleared components of the OsteoRemedies Hip Spacer System with respect to intended use, materials, design, and function. The information summarized in the Design Control Activities Summary demonstrates that the subject components met the pre-determined acceptance criteria for the verification activities.
{4}------------------------------------------------
# Device Description:
The OsteoRemedies Hip Spacer System is a modular temporary spacer device inclusive of femoral stems, femoral heads, and acetabular cups. The OsteoRemedies Hip Spacer System is sterile and single-use, intended for temporary use (maximum 180 days) for joint replacement. This submission expands the existing system to include additional sizes of the acetabular cup and femoral head. The system components are made of fully formed polymethylmethacrylate (PMMA) with gentamicin (REMEDY®) and gentamicin with vancomycin (REMEDY SPECTRUM® GV).
The OsteoRemedies Hip Spacer System provides a functional-mechanical mode of action; the system provides patients a temporary implant allowing for a natural range of motion and partial weight-bearing during treatment. The system is designed to preserve soft tissue to prevent further complications, such as muscular contraction, and to facilitate the subsequent joint replacement procedure. The REMEDY® components are protected from bacterial adhesion due to the presence of gentamicin. The REMEDY SPECTRUM® GV components are protected from bacterial adhesion due to the presence of gentamicin and vancomycin.
## Indications for Use:
The OsteoRemedies Hip Spacer System consists of modular heads and stems, and optional acetabular cups. The REMEDY® components of the OsteoRemedies Hip Spacer System include gentamicin and the REMEDY SPECTRUM® GV components include gentamicin and vancomycin.
The OsteoRemedies Hip Spacer System is indicated for temporary use (maximum 180 days) as an adjunct to total hip replacement (THR) in skeletally mature patients undergoing a two-stage procedure due to a septic process and where gentamicin/vancomycin are the most appropriate antibiotics based on the susceptibility pattern of the infecting micro-organism(s).
Following removal of the existing femoral and acetabular components and radical debridement, the head and stem components are inserted into the femoral medullary canal and can mate directly with the native acetabular or an acetabular component which is placed in the acetabular cavity. The device is intended for use in conjunction with systemic antibiotic therapy (standard treatment approach to an infection).
The OsteoRemedies Hip Spacer System is not intended for use for more than 180 days, at which time all components must be explanted, and a permanent device implanted or another appropriate treatment performed (e.g., resection arthroplasty, fusion, etc.).
# Performance Testing:
The following testing was performed on the predicate devices and is applicable to the subject components of the OsteoRemedies Hip Spacer System:
- Stem fatigue per ISO 7206-4 ●
- Neck fatigue per ISO 7206-6 ●
{5}------------------------------------------------
- Femoral head/stem disassembly based on ISO 7206-13
- Chemical and physical properties of cement ●
- Antibiotic elution kinetics ●
- Clinical data ●
- o Biocompatibility data
- Sterilization, pyrogenicity, bacterial endotoxin, and shelf life testing ●
## Substantial Equivalence:
Analyses of static and fatigue strength of the resin, fatigue strength of the head/stem construct, range of motion, risk of dislocation, surface roughness, and wear were performed on the XS REMEDY Head, XS REMEDY SPECTRUM GV Head, and XS REMEDY Acetabular Cup to support their substantial equivalence. In addition, the extra-small head and cup components contain less antibiotics than larger, previously cleared components of the OsteoRemedies Hip Spacer System.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.