K050251 · Biomet, Inc. · KWA · May 27, 2005 · Orthopedic
Device Facts
Record ID
K050251
Device Name
BALANCE HIP SYSTEM
Applicant
Biomet, Inc.
Product Code
KWA · Orthopedic
Decision Date
May 27, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3330
Device Class
Class 3
Attributes
Therapeutic
Indications for Use
Balance® Hip System Standard Femoral stems: Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. Rheumatoid arthritis. Correction of functional deformity. Correction of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques. Revision of previously failed total hip arthroplasty. The Balance® Hip System Standard Femoral stems are intended for uncemented use only. Balance® Hip System Microplasty® stems: Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. Rheumatoid arthritis. Correction of functional deformity. Correction of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques. Revision of previously failed femoral head resurfacing component. The Balance® Hip System Microplasty® stems are intended for uncemented use only.
Device Story
Balance Hip System consists of collarless, porous-coated femoral stems; available in standard and Microplasty geometries. Designed for implantation into the metaphysis of the femur; provides immediate three-point fixation (anterior-proximal, inferior-medial, distal-lateral). Stems feature a proximal threaded hole for inserter/extractor tool attachment. Used by orthopedic surgeons in clinical settings during total hip arthroplasty or revision procedures. Device provides structural support for hip joint replacement; facilitates patient mobility; restores joint function. Compatible with various acetabular components, constrained liners, and salvage/oncology systems.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Collarless, porous-coated femoral stems. Proximal geometry designed for three-point fixation. Threaded proximal hole for instrumentation. Uncemented application. Materials and design are technically equivalent to predicate femoral hip stems.
Indications for Use
Indicated for skeletally mature patients requiring total hip arthroplasty due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, proximal femur fractures (non-union, neck, trochanteric), or revision of failed hip arthroplasty/resurfacing. Intended for uncemented use.
Regulatory Classification
Identification
A hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis is a two-part device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral and an acetabular component, both made of alloys, such as cobalt-chromium-molybdenum. The femoral component is intended to be fixed with bone cement. The acetabular component is intended for use without bone cement (§ 888.3027).
{0}------------------------------------------------
MAY 27 2005 K050251 P 1/2
#### 510(k) Summary
| Applicant/Sponsor: | Biomet Manufacturing Corp.<br>56 East Bell Drive<br>P.O. Box 587<br>Warsaw, Indiana 46581-0587 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Gary Baker<br>Regulatory Specialist<br>Biomet Manufacturing Corp.<br>P.O. Box 587<br>Warsaw, Indiana 46581-0587<br>Phone: (574) 267-6639<br>FAX: (574) 372-1683 |
| Proprietary Name: | Balance <sup>®</sup> Hip System |
| Common Name: | Femoral Hip Stems |
| Classification Name: | The Balance <sup>®</sup> Hip System femoral stems and compatible shells, heads, and<br>liners included in this submission have the following classifications: |
1. Hip joint metal/polymer constrained cemented or uncemented prosthesis (21 CFR §888.3310)
2. Hip joint metal/metal semi-constrained, with a cemented acetabular component, prosthesis (21
- CFR §888.3320) 3. Hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis (21 CFR §888.3330)
- (21 CFR 9886.3530)
Hip joint metal/polymer semi-constrained cemented prosthesis (21 CFR §888.3350) 4.
- Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis 5. (21 CFR §888.3353)
- (21 CFR 9666.3333)
Hip joint metal/polymer/metal semi-constrained, porous-coated, uncemented prosthesis (21 CFR 6. 888.3358)
176
{1}------------------------------------------------
KOSOGI 1 2/2
## Legally Marketed Devices To Which Substantial Equivalence Is Claimed:
Mallory - Head® Total Hip System - Biomet Inc. (K921181, K030055) Mayo Conservative Hip Prosthesis - Zimmer Inc. (K943230) Bi-Metric® Total Hip System - Biomet Inc. (K921224, K030055)
Device Description: The Balance® Hip System Femoral stems are collarless, porous coated stems available Device Description: The Dalance "The Bygths. The proximal geometry of the Balance" Hip System IN 00th Standard and Micropiasy - Stein Teligaal Tilling of the metaphysis and provide immediate three point (anterior-proximal, inferior-medial, and distal-lateral) fixation.
The top of the stems have a threaded hole to accommodate the inserter / extractor tool.
Indications for Use: Balance® Hip System Standard Femoral stems
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular 1 . necrosis.
- Rheumatoid arthritis 2.
- Correction of functional deformity 3.
- Correction of ranchenal developed neck fracture, and trochanteric fractures of the 4. proximal femur with head involvement, unmanageable using other techniques.
- Revision of previously failed total hip arthroplasty. న్.
The Balance® Hip System Standard Femoral stems are intended for uncemented use only.
Indications for Use: Balance® Hip System Microplasty® stems
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular 1. necrosis.
- Rheumatoid arthritis 2.
- Correction of functional deformity 3.
- Corroon of non-union, femoral neck fracture, and trochanteric fractures of the 4. Proximal femur with head involvement, unmanageable using other techniques.
- Revision of previously failed femoral head resurfacing component. ર.
The Balance® Hip System Microplasty® stems are intended for uncemented use only.
Summary of Technologies: The Balance® Hip System stems are materials and utilize the Summal y of Technologics. The Datanoo Trip 55000 ses as the predicates. Testing determined that the stems are substantially equivalent to the predicate stems.
Reference literature and performance data demonstrate that the Balance® Hip Non-Clinical Testing: Nor Chilleur Technically equivalent to the predicate femoral hip stems.
All trademarks are property of Biomet, Inc.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features three abstract human figures connected by flowing lines. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 27 2003
Mr. Gary Baker Regulatory Specialist Biomet Incorporated P.O. Box 587 Warsaw, Indiana 46581-0587
Re: K050251
Trade/Device Name: Balance® Hip System Microplasty® Stems Regulation Number: 21 CFR 888.3310, 888.3320, 888.3330, 888.3350, 888.3358 Regulation Name: Hip joint metal/polymer constrained cemented or uncemented prosthesis; Hip joint metal/metal semi constrained, with a cemented acetabular component, prosthesis; Hip joint metal/ polymer semi-constrained cemented prosthesis; Hip joint metal/polymer/metal semi-constrained porous-coated, uncemented prosthesis Regulatory Class: III Product Code: KWZ, JDL, KWA, JDI, LZO, MEH, LPH, LZY, KWY Dated: May 6, 2005 Received: May 9, 2005
Dear Mr. Baker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Page 2 - Mr. Gary Baker
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0210. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
A. Stephen Purdy
Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications For Use
510(k) Number (IF KNOWN): K05025
Device Name: Balance Hip System Standard Femoral Stems
Indications for Use:
- ations for Use.
1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- 2. Rheumatoid arthritis
- 3. Correction of functional deformity
- 3. Correction of non-union. femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
- 5. Revision of previously failed total hip arthroplasty.
Balance® Hip System Femoral Stems are intended for uncemented use only.
Specific indications for compatible components that can be used with the above femoral stems include:
### Constrained Liners (K030047)
Constrained liners are intended for general use in skeletally mature individuals undergrand in the bone loss, soft Constrained liners are mielided for general use in skelcians valistocation, joint or bose loss, soft revision surgery at high risk of intra-operative instability and for whom all other options to constrained acetabular components have been considered.
# OSS / Salvage Systems / Total Femur (K974558. K002757. K021380. K033871)
Salvage/Oncology Hip and Total Femur components are also indicated for cases of ligament deficiency, Salvage Oncology Trip and Total Pennal Virginian of unsuccessful osteotomy or arthrodesis.
### Interlocking Stems (K990830, K042774)
Interlocking hip stens are indicated for non-cemented application in cases of revision, tracture, Interlocking mp siems are moreated for non econential.
oncology or other situations where severe proximal bone loss may compromise the fixation and stability of a standard type hip replacement prosthesis.
X Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801 Subpart D) AND OR
Over-the-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE
tion (ODE)
Hypt Rurla
Image /page/4/Picture/25 description: The image shows handwritten text. The text on the right side of the image reads 'K050251'. The text on the left side of the image is partially visible and blurry, but it appears to be a word.
{5}------------------------------------------------
### Indications For Use
510(k) Number (IF KNOWN): K050251
Device Name: Balance® Hip System Microplasty® Stems
Indications for Use:
- 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- Rheumatoid arthritis 2.
- Correction of functional deformity 3.
- Correction of functional deformily
Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur ব with head involvement, unmanageable using other techniques.
- while it not on of previously failed femoral head resurfacing components. న్న
Balance & Hip System Microplasty Stems are intended for uncemented use only.
Specific indications for compatible components that can be used with the above femoral stems include:
### Constrained Liners (K030047)
Constrained liners are intended for general use in skeletally mature individuals undergoung port on home loss, soft Constrained liners are mielided for general use in skercars non or brinn of bone loss, soft revision surgers at lightifisk of life distance date instability and for whom all other options to constrained acetabular components have been considered.
# OSS / Salvage Systems / Total Femur (K974558, K002757, K021380, K033871)
Salvage/Oncology Hip and Total Femur components are also indicated for cases of ligament deficiency, Salvage Oneology Trip and Total ions insuccessful osteotomy or arthrodesis.
### Interlocking Stems (K990830, K042774)
Interlocking hip stems are indicated for non-cemented application in cases of revision, tracture, Interlocking hip stems are indicated for non-centence approamer the fixation and stability of a standard type hip replacement prosthesis.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
(Division Sign-Off)
and Neurological Devices
510(k) Number kos0251
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.