The device is intended to be implanted with bone cement in cases of: 1. Painful and disabled joint resulting from avascular necrosis, osteoarthritis, rheumatoid arthritis, traumatic arthritis 2. Correction of varus, valgus or post traumatic deformity 3. Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement 4. Ligament deficiencies 5. Tumor resections 6. Treatment of non-unions, femoral neck and trochanteric fracture of the proximal femur with head involvement, unmanageable using other techniques 7. Revision of previously failed total joint replacement 8. Trauma
Device Story
IM Total Femur is an intramedullary rod system connecting proximal hip replacement components to knee femoral replacement components. Rods are slightly bowed to mimic natural femur anatomy; feature male tapers at both ends for modular connection to hip and knee components; secured via locking screw after impaction. Available in two taper configurations (Impact/OSS or Impact/Impact) with varying lengths. Used in orthopedic surgery for total hip and knee joint reconstruction. Surgeon selects appropriate rod based on distal knee component taper. Device articulates with previously cleared tibial bearings, base-plates, and femoral components. Provides structural support for patients with severe joint disease, trauma, or tumor resection; restores joint function and alignment.
Clinical Evidence
No clinical data provided. Bench testing only; mechanical testing confirmed taper integrity under expected loading conditions.
Technological Characteristics
Intramedullary rod system; modular design with male tapers; slightly bowed geometry. Materials, design, and indications are similar to predicate oncology/salvage systems. Cemented fixation. No software or electronic components.
Indications for Use
Indicated for patients requiring total hip and knee joint replacement due to avascular necrosis, osteoarthritis, rheumatoid arthritis, traumatic arthritis, varus/valgus/post-traumatic deformity, failed osteotomy/arthrodesis/joint replacement, ligament deficiencies, tumor resections, non-unions, or femoral neck/trochanteric fractures.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
APR 1 6 2004
K033871 (pg 1 of 2)
BIOMET
### 510(k) Summary
Applicant/Sponsor: Biomet Manufacturing Corp.
Patricia Sandborn Beres Contact Person: Senior Regulatory Specialist
Proprietary Name: IM Total Femur
Common or Usual Name: Oncology, Salvage Hip/Knee
#### Classification Name:
- 1. Prosthesis, Hip, Semi-constrained, Metal/Polymer, Cemented (21 CFR Part 888.3350)
- 1. Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer, (21 CFR Part 888.3510)
- 3. Poosthesis, Knee, PatelloFemorotibial, Semi-constrained, Polymer/Metal/Polymer, cemented (21 CFR Part 888.3560)
Legally Marketed Devices To Which Substantial Equivalence Is Claimed: Oncology Salvage System (K002757), Salvage/Oncology Hip and Total Femur System (K974558), Barrago Cystem (K915132), Biomet SCK Knee System (K015132), Biomet SCK Knee System (K003296), AVL Hinged Knee (K010774).
Device Description: The IM Total Femur is, in simplest terms, an intramedullary (IM) rod which connects a proximal hip replacement component to a knee femoral replacement component.
The IM rods are slightly bowed to mimic the natural femur and find appropriate alignment. Both ends of the rod have a male taper for connection to the hip proximal femur and a knee femoral component. Following impaction, a locking screw further secures the parts. Two types of IM rods are available based on the tapers of the distal knee component selected by the surgeon. IM total femur rods with an "Impact" taper at one end and an "OSS" taper at the other come in 1cm increments for replacements of 35cm to 45cm. The second type of rod have an "Impact" taper at both ends and come in 3cm increments for replacement from 34 to 46cm.
> MAILING ADDRESS 120 Box 5247 Warsaw. IN 46581 058 ?
SHIPPING ADDRESS 56 E. Bell Drive Warsaw. IN 46582
(DEFFICT) 57 4 207 (16 (16) 34)
145
E Min Biomers bionnet.son
{1}------------------------------------------------
K033871 (pg 2 of 2)
IM Total Femur 510(k) Summary Page 2
Femoral knee components are available with both "OSS" and "Impact" taper bores. The r omoral know components articulate with previously cleared tibial bearings, base-plates and forhordi oomponents. The device is compatible with any of Biomet's proximal modular hip components.
Intended Use: The device is intended to be implanted with bone cement in cases of:
- 1. Painful and disabled joint resulting from avascular necrosis, osteoarthritis, rheumatoid arthritis, traumatic arthritis
- 2. Correction of varus, valgus or post traumatic deformity
- 3. Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement
- 4. Ligament deficiencies
- 5. Tumor resections
- 6. Treatment of non-unions, femoral neck and trochanteric fracture of the proximal femur with head involvement, unmanageable using other techniques
- 7. Revision of previously failed total joint replacement
- 8. Trauma
Summary of Technologies: The IM Total Femur components (materials, design and indications) are similar or identical to the predicate devices.
Non-Clinical Testing: Mechanical testing has shown the tapers capable of withstanding expected loading conditions.
Clinical Testing: None provided
All tradernarks are property of Biomet, Inc.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of an eagle or other bird-like figure.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# APR 1 6 2004
Ms. Patricia Sanborn Beres Senior Regulatory Specialist Biomet Manufacturing Corp. P.O. Box 587 Warsaw, Indiana 46581-0587
Re: K033871 Trade/Device Name: IM Total Femur Regulation Number: 21 CFR 888.3350, 21 CFR 888.3510, 21 CFR 888.3560 Regulation Name: Hip joint metal/polymer semi-constrained cemented prosthesis, Knee joint femorotibial metal/polymer constrained cemented prosthesis, K nee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: II
Product Code: JDI, KRO, JWH Dated: March 22, 2004 Received: March 23, 2004
Dear Ms. Beres:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The gencral controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Codc of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Scctions 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
## Page 2 - Ms. Patricia Sanborn Beres
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sincerely yours,
Mark N. Millican
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K033871
Device Name: IM Total Femur
Indications For Use:Total hip and knee joint replacement in cases of:
- Indications For Southing from avascular necrosis, osteoarthritis, theumatoid arthritis, traumatic arthritis
- 2. Correction of varus, valgus or post traumatic deformity
- 2. Corroction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement
- 4. Ligament deficiencies
- 5. Tumor resections
- 6. Treatment of non-unions, femoral neck and trochanteric fracture of the proximal femur with head involvement, unmanageable using other techniques
- 7. Revision of previously failed total joint replacement
- 8. Trauma
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for Mark N. Wilkerson
Division of General, Restorative. Page 1 of _1 and Neurological Devices
510(k) Number K033871
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.