SYNTHES CANNULATED ANGLE BLADE PLATE (CABP) SYSTEM

K974537 · Synthes (Usa) · KTT · Feb 18, 1998 · Orthopedic

Device Facts

Record IDK974537
Device NameSYNTHES CANNULATED ANGLE BLADE PLATE (CABP) SYSTEM
ApplicantSynthes (Usa)
Product CodeKTT · Orthopedic
Decision DateFeb 18, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

Synthes CABP System is a plate and screw system intended to treat fractures of the proximal humerus and distal tibia. Proximal humerus fractures include two-part greater tubercle fractures and fracture dislocations, two-part surgical neck fractures and fracture dislocations, three-part fractures or fracture dislocations, fractures in osteopenic bone, and nonunions and malunions. Distal tibia fractures include acute fractures, fractures in osteopenic bone, and nonunions and malunions.

Device Story

Cannulated Angle Blade Plate (CABP) system for orthopedic fracture fixation; used in proximal humerus and distal tibia. Device consists of straight plate with integrated blade; cannulated blade allows guide wire placement and adjustment. Undersurface cuts reduce bone contact area to preserve cortical blood supply and minimize damage-induced porosis; undercuts also improve bending uniformity for contouring. System includes round and dynamic compression screw holes; accepts 4.5 mm and 6.5 mm screws. Used by surgeons in clinical/OR settings to stabilize bone fragments; facilitates healing in acute fractures, nonunions, and malunions.

Clinical Evidence

No clinical data provided; substantial equivalence based on design, material, and intended use comparison to predicate device.

Technological Characteristics

Materials: stainless steel and commercially pure titanium. Design: straight plate with cannulated blade; undersurface undercuts for bone contact reduction and uniform bending. Fixation: accepts 4.5 mm and 6.5 mm screws; round and dynamic compression holes. Form factor: multiple blade lengths.

Indications for Use

Indicated for patients with proximal humerus fractures (two-part greater tubercle, two-part surgical neck, three-part fractures/dislocations, osteopenic bone, nonunions, malunions) and distal tibia fractures (acute, osteopenic bone, nonunions, malunions).

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K974537 Feb. 18, 1998 # SYNTHES ## Attachment VIII: Summary of Safety and Effectiveness Info. [510(k) Summary] ### SUBMITTER Synthes (USA) 1690 Russell Road Paoli, PA 19301 (610) 647-9700 Contact: Sheri L. Musgnung ### COMMON OR USUAL NAME: Appliance, Fixation, Nail/Blade/Plate Combination, Single Component ### DEVICE CLASSIFICATION: Class II, 21 CFR 888.3030; 888.3040 ### PREDICATE DEVICE: Synthes Cannulated Angle Blade Plate System (K954289) ### DESCRIPTION: The CABP is a straight plate with a blade at the head to allow for better fixation in the head of the humerus or in the distal tibia. The blade of the plate is cannulated to fit over a guide wire, allowing for the adjustment of the wire placement several times without adversely affecting the final result. There are cuts in the undersurface of the plate to reduce the surface area of the plate in contact with bone. The area of contact between the plate and the bone is decreased in an effort to reduce damage to the cortical blood supply under the plate, and resultant reduction in damage-induced porosis and remodeling of the bone near the plate undersurface. Plate undercuts also make the bending properties of the plate more uniform, which facilitate contouring. The plate has round and dynamic compression screw holes, accepts 4.5 mm and 6.5 mm screws, is available in multiple blade lengths, and is manufactured from stainless steel and commercially pure titanium. ### INTENDED USE: Synthes CABP System is a plate and screw system intended to treat fractures of the proximal humerus and distal tibia. Proximal humerus fractures include two-part greater tubercle fractures and fracture dislocations, two-part surgical neck fractures and fracture dislocations, three-part fractures or fracture dislocations, fractures in osteopenic bone, and nonunions and malunions. Distal tibia fractures include acute fractures, fractures in osteopenic bone, and nonunions and malunions. 1C {1} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB 18 1998 Ms. Sheri L. Musgnung Regulatory Affairs Associate Synthes (USA) 1690 Russell Road P.O. Box 1766 Paoli, Pennsylvania 19301 Re: K974537 Synthes (USA) Cannulated Angle Blade Plate (CABP) System Regulatory Class: II Product Code: KTT Dated: December 2, 1997 Received: December 3, 1997 Dear Ms. Musgnung: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800-to-895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2} Page 2 - Ms. Sheri L. Musgnung This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3} SYNTHES SYNTHES (USA) 1690 Russell Road Post Office Box 1766 Paoli, Pennsylvania 19301 Telephone 610-647-9700 Page 1 of 1 510(k) Number (if known): k974537 Device Name: Synthes (USA) Cannulated Angle Blade Plate System Indications For Use: The Synthes CABP System is a plate and screw system intended to treat fractures of the proximal humerus and distal tibia. Proximal humerus fractures include two-part greater tubercle fractures and fracture dislocations, two-part surgical neck fractures and fracture dislocations, three-part fractures or fracture dislocations, fractures in osteopenic bone, and nonunions and malunions. Distal tibia fractures include acute fractures, fractures in osteopenic bone, and nonunions and malunions. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ☐ ![img-1.jpeg](img-1.jpeg) Original Instruments and Implants of the Association for the Study of Internal Fixation—AO/ASIF
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...