Hoffman LRF System

K161753 · Stryker GmbH · KTT · Nov 15, 2016 · Orthopedic

Device Facts

Record IDK161753
Device NameHoffman LRF System
ApplicantStryker GmbH
Product CodeKTT · Orthopedic
Decision DateNov 15, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The Hoffmann LRF System is intended for fixation of fractures, joint contractures, fusions, limb lengthening, deformity correction, and bone and soft tissue reconstruction in pediatric patients and adults.

Device Story

Hoffmann LRF System is a modular, ring-based external fixation system for orthopedic limb conditions. Device uses pins and wires to connect bone to rings; rings are statically placed or gradually manipulated to achieve correction. System components include stainless steel, aluminum, PEEK, and carbon fiber. Used in clinical settings by physicians for fracture fixation, limb lengthening, and deformity correction. Provides mechanical stabilization and distraction capabilities to facilitate bone healing and reconstruction. Benefits include customized treatment for complex orthopedic pathologies.

Clinical Evidence

No clinical data was performed or required for this submission. Substantial equivalence is based on non-clinical bench testing, including static bending, dynamic bending, pullout, and corrosion testing.

Technological Characteristics

Modular, ring-based external fixation system. Materials: stainless steel, aluminum, PEEK, and carbon fiber. Mechanical fixation via pins and wires. System is non-powered and mechanical in nature.

Indications for Use

Indicated for pediatric and adult patients for treatment/fixation of open/closed fractures, post-traumatic joint contracture, fractures/disease causing contracture or requiring distraction, pseudoarthrosis/non-union, limb lengthening, bony/soft tissue deformity correction, segmental defect correction, joint arthrodesis, comminuted intra-articular distal radius fractures, and bone transport. Adults only: osteotomy, revision procedures, bone reconstruction, foot fusions/replantations, Charcot foot reconstruction, and Lisfranc dislocations.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 15, 2016 Stryker GmbH Mr. Paul Nelson Regulatory Affairs Specialist Bohnackerweg 1 CH-2545 Selzach Switzerland Re: K161753 Trade/Device Name: Hoffman LRF System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT Dated: October 12, 2016 Received: October 13, 2016 Dear Mr. Nelson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K161753 Device Name Hoffmann LRF System Indications for Use (Describe) The Stryker Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of: - Open and Closed Fractures - · Post-traumatic joint contracture which has resulted in loss of range of motion - · Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction - · Pseudoarthrosis or non-union of long bones - · Limb lengthening by epiphyseal, diaphyseal, or metaphyseal distraction - · Correction of bony or soft tissue deformity - · Correction of segmental bony or soft tissue defects - · Joint arthrodesis - · Management of comminuted intra-articular fractures of the distal radius - · Bone transport The Hoffmann LRF System is indicated in adults for: - · Osteotomy - · Revision procedure where other treatments or devices have been unsuccessful - Bone reconstruction procedures - · Fusions and replantations of the foot - · Charcot foot reconstruction - · Lisfranc dislocations Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) Summary | Submitter: | Stryker GmbH<br>Bohnackerweg 1<br>CH-2545 Selzach<br>Switzerland | |-----------------------|----------------------------------------------------------------------------------------------------| | Contact Person: | Paul Nelson<br>Regulatory Affairs Specialist<br>Phone: (201) 831-5691<br>Fax: (201) 831-6500 | | Date Prepared: | June 13, 2016 | | Name of Device: | Hoffmann LRF System | | Common or Usual Name: | External Fixation Device | | Classification Name: | Single/multiple component metallic bone fixation appliances and<br>accessories (21 CFR § 888.3030) | | Regulatory Class: | Class II | | Product Code: | KTT | | Primary Predicate: | Hoffmann LRF System, K153377 | #### Device Description: The Hoffmann LRF System is a modular, ring-based, external fixation system designed to address certain orthopedic conditions of the limbs. Through a series of pins and wires, the bone is connected to this system with the rings statically placed, or gradually manipulated, depending on the type of correction needed. The modular design allows the system to be customized according to the needs of the patient and utilizes stainless steel, aluminum, PEEK, and carbon fiber. Components of the following systems may be used with this system: Monticelli-Spinelli External Fixation System, Apex Pins, Trauma Pelvic Set, Hoffmann II External Fixation System, Hoffmann 3 External Fixation System, Hoffmann II External Fixation System 90° Post, Hoffmann II Miami Post, Hoffmann II Carbon Connecting Rods, Hoffmann II MRI External Fixation System, and Hoffmann II Compact MRI External Fixation System. Use of these components does not confer MRI compatibility to the Hoffmann LRF System. ## Intended Use: The Hoffmann LRF System is intended for fixation of fractures, joint contractures, fusions, limb lengthening, deformity correction, and bone and soft tissue reconstruction in pediatric patients and adults. ## Indications for Use: The Stryker Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of: - Open and Closed Fractures - Post-traumatic joint contracture which has resulted in loss of range of motion {4}------------------------------------------------ - . Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction - . Pseudoarthrosis or non-union of long bones - . Limb lengthening by epiphyseal, diaphyseal, or metaphyseal distraction - . Correction of bony or soft tissue deformity - . Correction of segmental bony or soft tissue defects - . Joint arthrodesis - Management of comminuted intra-articular fractures of the distal radius ● - . Bone transport The Hoffmann LRF System is indicated in adults for: - . Osteotomy - . Revision procedure where other treatments or devices have been unsuccessful - . Bone reconstruction procedures - Fusions and replantations of the foot ● - Charcot foot reconstruction - . Lisfranc dislocations #### Comparison of Technological Characteristics: This submission does not introduce any new technology to the Hoffmann LRF System. The technological characteristics of the subject and predicate systems are identical. #### Performance Data: #### Non-Clinical Testing Comparative mechanical testing to the predicate system demonstrated substantial equivalence. The following tests were performed: - Static Bending - Dynamic Bending - Pullout - Corrosion ## Clinical Testing Clinical testing was not performed, or required, for this submission. #### Conclusion: With the exception of the material and design changes specified, the Hoffmann LRF System in this submission is identical to the previously cleared and primary predicate, K153377. Mechanical testing resulting from risk analyses demonstrated that the subject system performs as intended and at least as well as the predicates. Based on these attributes, the subject device is deemed substantially equivalent.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...