The TL-HEX System is intended for limb lengthening by metaphyseal or epiphyseal distractions, fixation of open and closed fractures, treatment of nonunion or pseudoarthrosis of long bones and correction of bony or soft tissue defects or deformities. Within this range, indications include: Post-traumatic joint contracture which has resulted in loss of range of motion Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction Open and closed fracture fixation Pseudoarthrosis of long bones Limb lengthening by epiphyseal or metaphyseal distraction Correction of bony or soft tissue deformities Correction of bony or soft tissue defects Joint arthrodesis Infected fractures or non-unions
Device Story
Multilateral external fixation system for orthopedic reconstruction; includes aluminum rings, footplates, adjustable struts, and pins/wires. Used by physicians for limb lengthening, fracture fixation, and deformity correction. System includes web-based software to assist physicians in creating patient-specific strut adjustment schedules. Physician inputs clinical parameters into software; software calculates adjustment schedule for the six struts; physician uses output to guide manual adjustments of the hexapod frame. Benefits include precise, controlled correction of bony/soft tissue deformities and fractures. Operates in clinical settings; hardware provides mechanical stability; software facilitates treatment planning.
Clinical Evidence
Bench testing only. Mechanical testing performed per ASTM F 1541-02 (2007)e1. Software verification and validation (IQ, OQ, PQ) conducted per FDA guidance. No clinical data required.
Indicated for patients requiring limb lengthening, fracture fixation (open/closed), treatment of nonunion/pseudoarthrosis, correction of bony/soft tissue defects/deformities, joint arthrodesis, or treatment of infected fractures/non-unions, including post-traumatic joint contracture.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Orthofix TL-HEX True Lok Hexapod System (TL-HEX) (K141078)
Related Devices
K152171 — Orthofix TL-HEX True Lok Hexapod System (TL-HEX) V1.4 · Orthofix Srl · Sep 28, 2015
K170650 — Orthofix TrueLok Hexapod System (TL-HEX) V2.0 · Orthofix Srl · May 10, 2017
K141078 — ORTHOFIX TL-HEX TRUE LOK HEXALOBE SYSTEM (TL-HEX) · Orthofix Srl · Sep 2, 2014
K093047 — SMITH & NEPHEW CIRCULAR FIXATION SYSTEM · Smith & Nephew, Inc. · Sep 27, 2010
K231461 — Smart Correction System (HA Half Pins) · Wishbone Medical, Inc. · Jun 16, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling three human profiles facing right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 10, 2014
Orthofix Srl % Ms. Cheryl Wagoner Principal Consultant Wagoner Consulting LLC PO Box 15729 Wilmington, North Carolina 28408
Re: K143125
Trade/Device Name: Orthofix TL-HEX True Lok Hexapod System (TL-HEX) V1.3 Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation applications and accessories Regulatory Class: Class II Product Code: KTT, OSN Dated: November 7, 2014 Received: November 10, 2014
Dear Ms. Wagoner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
K143125 510(k) Number (if known):
Device Name: Orthofix TL-HEX True Lok Hexapod System (TL-HEX) V1.3
Indications for Use:
The TL-HEX System is intended for limb lengthening by metaphyseal or epiphyseal distractions, fixation of open and closed fractures, treatment of nonunion or pseudoarthrosis of long bones and correction of bony or soft tissue defects or deformities. Within this range, indications include:
- Post-traumatic joint contracture which has resulted in loss of range of motion
· Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction
- Open and closed fracture fixation
- · Pseudoarthrosis of long bones
- · Limb lengthening by epiphyseal or metaphyseal distraction
- · Correction of bony or soft tissue deformities
- · Correction of bony or soft tissue defects
- · Joint arthrodesis
- · Infected fractures or non-unions
Prescription Use _____________________________________(21 CFR 801 Subpart (21 CFR 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign Off) Division of Orthopedic Devices 510(k) Number: K143125
{3}------------------------------------------------
## Special 510(k) Premarket Notification Orthofix TL-HEX True Lok Hexapod System (TL-HEX) V1.3
#### 510(k) Summary (as required by 21 CFR 807.92)
| Submitter | Orthofix Srl |
|-----------|-------------------------------|
| | Via delle Nazioni, 9 |
| | 37012 Bussolengo (VR) - Italy |
| Telephone | +39 045 6719.000 |
| Fax | +39 045 6719.380 |
| Contact Person | Gianluca Ricadona<br>Quality & Regulatory Affairs Manager |
|----------------|-----------------------------------------------------------|
| Address | Via delle Nazioni, 9<br>37012 Bussolengo (VR) - Italy |
| Telephone | + 39 045 6719.000 |
| Fax | + 39 045 6719.380 |
| email | gianlucaricadona@orthofix.com |
Date Prepared 10/30/2014
| Trade Name | Orthofix TL-HEX True Lok Hexapod System (TL-HEX) V1.3 |
|---------------------|-----------------------------------------------------------------------------|
| Common Name | Multilateral Fixators and Accessories |
| Panel Code | Orthopaedics/87 |
| Classification Name | Single/multiple component metallic bone fixation appliances and accessories |
| Class | Class II |
| Regulation Number | 21 CFR 888.3030 |
| Product Code | KTT, OSN |
| Name of Predicate Device | 510(k) # | Manufacturer |
|--------------------------------------------------|----------|--------------|
| Orthofix TL-HEX True Lok Hexapod System (TL-HEX) | K141078 | Orthofix Srl |
| Description | The Subject device is a multilateral external fixation system. The System<br>can also be used with a web-based software component that is designed<br>to be used to assist the physician in creating a patient adjustment<br>schedule that assists in adjusting the six struts. The System can also be<br>used with other existing Orthofix external fixation components and screws<br>(such as TrueLok or X Caliber).<br><br>Components of the system include:<br>Full, 5/8 and 3/8 aluminum Rings Double Row Footplates Adjustable struts Aluminum strut clips (number and direction) Stainless steel instrumentation such as hex drivers, wrenches,<br>and pliers Implantable components such as half pins and wires Web-based software |
|-------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications and<br>Intended Use | The TL-HEX System is intended for limb lengthening by metaphyseal or<br>epiphyseal distractions, fixation of open and closed fractures, treatment<br>of non-union or pseudoarthrosis of long bones and correction of bony or<br>soft tissue defects or deformities. Within this range, indications include: |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{4}------------------------------------------------
| • Post-traumatic joint contracture which has resulted in loss of range of motion |
|-----------------------------------------------------------------------------------------------------------------------------------------|
| • Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction |
| • Open and closed fracture fixation |
| • Pseudoarthrosis of long bones |
| • Limb lengthening by epiphyseal or metaphyseal distraction |
| • Correction of bony or soft tissue deformities |
| • Correction of bony or soft tissue defects |
| • Joint arthrodesis |
| • Infected fractures or non-unions |
| Technological<br>Characteristics<br>and Substantial<br>Equivalence | Documentation was provided to demonstrate that the subject device is<br>substantially equivalent to the Predicate Orthofix TL-HEX True Lok<br>Hexapod System (TL-HEX) (K141078). The subject device is<br>substantially equivalent to the predicate device in intended use,<br>indications for use, technological characteristics, and labeling. The<br>subject device includes an extended range of rings, footplates, struts, and<br>a consequent update of the software. |
|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Data | The potential hazards have been evaluated and controlled through a Risk<br>Management Plan.<br><br>All testing met or exceeded the requirements as established by the test<br>protocols and applicable standards. A review of the mechanical data<br>indicates that the hardware components of the Subject device are capable<br>of withstanding expected loads without failure. Additionally, software<br>verification and validation testing was completed in conformance with<br>FDA's guidance document entitled "General Principles of Software<br>Validation; Final Guidance for Industry and FDA Staff." The results of<br>software testing indicate that the software performed as intended. The<br>Subject device was therefore found to be substantially equivalent to the<br>Predicate. Clinical data was not needed to support the safety and<br>effectiveness of the Subject Device.<br><br>The following mechanical and software testing were performed:<br>ASTM F 1541-02 (2007)e1 "Standard Specification and Test<br>Methods for External Skeletal Fixation Devices" Software IQ, OQ, PQ |
|------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion | Based on the indications for use, technological characteristics, materials,<br>and comparison to predicate devices, the Subject Orthofix TL-HEX True<br>Lok Hexapod System (TL HEX) has been shown to be substantially |
for its intended use.
equivalent to legally marketed predicate device, and is safe and effective
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.