SYNTHES SMALL EXTERNAL FIXATION SYSTEM, SYNTHES LARGE EXTERNAL FIXATION SYSTEM

K122455 · Synthes · KTT · Apr 11, 2013 · Orthopedic

Device Facts

Record IDK122455
Device NameSYNTHES SMALL EXTERNAL FIXATION SYSTEM, SYNTHES LARGE EXTERNAL FIXATION SYSTEM
ApplicantSynthes
Product CodeKTT · Orthopedic
Decision DateApr 11, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

Synthes External Fixation Devices are intended for use in the construction of an external fixation frame for treatment of various fracture types that require external fixation.

Device Story

Synthes Small and Large External Fixation Systems consist of metallic clamps, rods, bars, and rings used to construct external fixation frames for orthopedic fracture treatment. This submission establishes MR Conditional status for these existing devices. The system is used by surgeons in clinical settings to stabilize bone segments, assist in reduction, or facilitate bone transport/lengthening. The device provides mechanical stabilization of fractures, allowing for healing or secondary procedures. The MR Conditional designation ensures safety during MRI procedures by meeting specific displacement, torque, RF heating, and image artifact standards.

Clinical Evidence

Bench testing only. No clinical data provided. Performance testing included displacement and torque testing (ASTM F2052, ASTM F2213), RF heating evaluation (ASTM F2182), and image artifact assessment (ASTM F2119).

Technological Characteristics

Metallic bone fixation components (clamps, rods, bars, rings). MR Conditional. Testing standards: ASTM F2052 (displacement), ASTM F2213 (torque), ASTM F2182 (RF heating), ASTM F2119 (image artifact). Mechanical strength analysis performed for Bridging Rod Component.

Indications for Use

Indicated for stabilization and treatment of fractures of small bones (hand, wrist, forearm, foot, ankle) and long bone/pelvic fractures. Small system uses: preliminary fixation before ORIF, unstable distal radius fractures, open/comminuted bilateral fractures, fractures with soft tissue/bone/vascular/neural injury, fracture dislocations, failed closed reduction, pediatric open fractures/osteotomies. Large system uses: soft tissue/fracture stabilization, polytrauma, pelvic fractures, arthrodeses, osteotomies, total joint failures, neutralization of limited internal fixation, non-unions, intra-operative reduction, bone segment transport, leg lengthening.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 11/2245 5 **** ↓ ﻢ ﻬﻢ :1 # Attachment A - Updated 510(k) Summary Date Prepared: March 4, 2013 | Sponsor: | Synthes<br>Angela F. Lassandro<br>1301 Goshen Parkway<br>West Chester, PA 19380<br>(610) 719-6854 | |-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | Synthes Small and Large External Fixation Systems - MR Conditional | | Classification: | Classification: Class II, §888.3030, Single/multiple component metallic<br>bone fixation appliances and accessories.<br>Product Code: KTT | | Predicate Device: | Synthes Small External Fixation System (K031724, K033158, K961350,<br>K031724, K050631, K090658, K082650, K952296, K040701)<br>Synthes Large External Fixation System (K082650, K033158, K030390,<br>K031428, K962913, K952296, K950384, K043039, K914558) | | Device Description: | Synthes Reprocessed External Fixation Devices consist of various clamps,<br>rods, bars and rings which are used to construct an external fixation frames<br>in the treatment of various types of fractures. | | Intended Use: | Synthes External Fixation Devices are intended for use in the construction of<br>an external fixation frame for treatment of various fracture types that require<br>external fixation. | | Indications for Use: | The Synthes Small External Fixation System is intended to stabilize and provide<br>treatment for fractures of the small bones, such as the hand, wrist, forearm, foot,<br>and ankle. Specifically, the components can be used for:<br>· Preliminary fixation before ORIF<br>· Unstable fractures of the distal radius (both intra- and extra-articular)<br>· Open and/or comminuted bilateral fractures<br>· Fractures in combination with extensive soft tissue injury, bone loss, and<br>vascular and/or neural involvement<br>· Fracture dislocations<br>· Failed closed reduction with casting resulting in secondary deformity<br>(radial shortening and angulations)<br>· Pediatric open fractures with bone loss and osteotomies | | | The Synthes Large External Fixation Systems is intended to provide treatment<br>for long bone and pelvic fractures that require external fixation. Specifically, the<br>components can be used for:<br>• Stabilization of soft tissues and fractures<br>• Poyltrauma/multiple orthopedic trauma<br>• Vertically stable pelvic fractures, or as a treatment adjunct for vertically<br>unstable pelvic fractures<br>• Arthrodeses and osteotomies with soft tissue problems; failures of total<br>joints<br>• Neutralization of fractures stabilized with limited internal fixation<br>• Non-unions/septic non-unions<br>• Intra-operative reductions/stabilization tool to assist with indirect<br>reduction<br>• Unilateral rectilinear bone segment transport or leg lengthening | | Substantial<br>Equivalence: | Performance testing has been completed to demonstrate the use of Synthes<br>Small and Large External Fixation Systems in the MR environment.<br>Displacement and torque testing has met the requirements in ASTM F2052<br>and ASTM F2213, respectively. The devices were shown to have acceptable<br>heating when tested for RF Heating in accordance with ASTM F2182.<br>Image artifact for the devices was determined through testing in accordance<br>with ASTM F2119. Additionally, an engineering analysis was conducted to<br>assess mechanical strength of the Bridging Rod Component of the Synthes<br>Large External Fixation System. | . {1}------------------------------------------------ .122455 : : and the comments of the comments of the comments of {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, represented by three curved lines that suggest the head, body, and tail feathers. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 Synthes % Angela Lassandro Regulatory Affairs Manager 1301 Goshen Parkway West Chester, Pennsylvania 19380 Letter dated: April 11, 2013 Re: K122455 Trade/Device Name: Synthes Small External Fixation System - MR Conditional, Synthes Large External Fixation System - MR Conditional Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliance and Accessories Regulatory Class: Class II Product Code: KTT Dated: March 4, 2013 Received: March 5, 2013 Dear Ms. Lassandro: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {3}------------------------------------------------ #### Page 2 – Angela Lassandro CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mark N.Melkerson-S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indications for Use Statement 510(k) Number (if known): K122455 Device Name: Synthes Small External Fixation System - MR Conditional Indications for Use: The Synthes Small External Fixation System is intended to stabilize and provide treatment for fractures of the small bones, such as the hand, wrist, forearm, foot, and ankle. Specifically, the components can be used for: - · Preliminary fixation before ORIF - · Unstable fractures of the distal radius (both intra- and extra-articular) - · Open and/or comminuted bilateral fractures - · Fractures in combination with extensive soft tissue injury, bone loss, and vascular and/or neural involvement - · Fracture dislocations - Failed closed reduction with casting resulting in secondary deformity (radial shortening and angulations) - · Pediatric open fractures with bone loss and osteotomies Prescription Use X (Per 21 CFR 801.109) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Image /page/4/Picture/17 description: The image shows the name "Michael Owens" in a stylized font. The first name, "Michael," is written in a clear, bold font. The last name, "Owens," is also in bold, but the letters are stylized with decorative elements. There are some symbols between the first and last name. Division of Orthopaedic Devices p. 1 of 2 {5}------------------------------------------------ ### Indications for Use Statement 510(k) Number (if known): K122455 Device Name: Synthes Large External Fixation System - MR Conditional Indications for Use: The Synthes Large External Fixation Systems is intended to provide treatment for long bone and pelvic fractures that require external fixation. Specifically, the components can be used for: - · Stabilization of soft tissues and fractures - · Poyltrauma/multiple orthopedic trauma - · Vertically stable pelvic fractures, or as a treatment adjunct for vertically unstable pelvic fractures - · Arthrodeses and osteotomies with soft tissue problems; failures of total joints - · Neutralization of fractures stabilized with limited internal fixation - · Non-unions/septic non-unions - · Intra-operative reductions/stabilization tool to assist with indirect reduction - · Unilateral rectilinear bone segment transport or leg lengthening Prescription Use X (Per 21 CFR 801.109) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Image /page/5/Picture/18 description: The image shows the name "Michael Owens" in a bold, sans-serif font. The letters are black against a white background. There is some decorative imagery between the first and last name. Division of Orthopaedic Devices
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...