ALLIANCE MEDICAL CORPORATION REPROCESSED EXTERNAL FIXATION DEVICES
K052064 · Alliance Medical Corp. · KTT · Aug 25, 2005 · Orthopedic
Device Facts
Record ID
K052064
Device Name
ALLIANCE MEDICAL CORPORATION REPROCESSED EXTERNAL FIXATION DEVICES
Applicant
Alliance Medical Corp.
Product Code
KTT · Orthopedic
Decision Date
Aug 25, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Intended Use
External Fixation Devices are intended to be used for the fixation of supracondylar, or condylar fractures of the femur; for fusion of a joint; for surgical procedures that involve cutting the bone, for fixation of bone fractures; bone reconstruction; as a guide pin for insertion of other implants; or may be implanted through the skin so that a pulling force or traction may be applied to the skeletal system; and others may be used for fixation of bone fractures, for bone reconstructions, as a quide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Device Story
Reprocessed external fixation devices consist of frames, clamps, rods, couplings, pins, posts, fasteners, washers, nuts, hinges, sockets, and screws. Used in orthopedic surgery for bone fracture management, reconstruction, and deformity correction. Devices are implanted through the skin to provide skeletal fixation or traction. Operated by physicians in clinical/surgical settings. Reprocessed devices are provided non-sterile; hospital-based sterilization required prior to use. Functionality relies on mechanical stability provided by metal alloys, plastics, and composites. Output is physical stabilization of bone segments to facilitate healing.
Clinical Evidence
Bench testing only. No clinical data provided. Performance demonstrated through laboratory validation of reprocessing and functional testing.
Technological Characteristics
Components include metal alloys, plastics, and composites. Mechanical fixation via frames, clamps, rods, and fasteners. Non-sterile delivery; requires hospital-based sterilization. No software or electronic components.
Indications for Use
Indicated for patients requiring external skeletal fixation and treatment of fractures, osteotomy, arthrodesis, correction of deformities, fracture revision, bone reconstruction procedures, limb lengthening, correction of bony or soft tissue deformities, and segmental bony or soft tissue defects.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
K052062 — ALLIANCE MEDICAL CORPORATION REPROCESSED EXTERNAL FIXATION DEVICES · Alliance Medical Corp. · Aug 25, 2005
K012634 — REPROCESSED EXTERNAL FIXATION DEVICE · Alliance Medical, Inc. · Jun 24, 2002
K052065 — ALLIANCE MEDICAL CORPORATION REPROCESSED EXTERNAL FIXATION DEVICES · Alliance Medical Corp. · Aug 22, 2005
K032058 — MODIFICATION TO REPROCESSED SYNTHES EXTERNAL FIXATION DEVICES · Alliance Medical Corp. · Jul 30, 2003
K051554 — ALLIANCE MEDICAL CORPORATION REPROCESSED EXTERNAL FIXATION DEVICES · Alliance Medical Corp. · Jun 29, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
K052064
## PART B: 510(k) SUMMARY
- Submitter: Alliance Medical Corporation 10232 South 51st Street Phoenix, Arizona 85044 Contact: Moira Barton Requlatory Affairs Manager (480) 763-5300 (o) (480) 763-5310 (f)
- Date of preparation: July 15, 2005
Trade/Proprietary Name: Reprocessed External Fixation Name of device: Devices Common or Usual Name: External Fixation Devices, Fixation Appliance, Single/Multiple Component Classification Name: Single/Multiple Component Metallic Bone Fixation Appliances and Accessories and Smooth or Threaded Metallic Bone Fixation Fastener
mbarton@alliance-medical.com
## Predicate device(s):
- K012645 Reprocessed EBI® External Fixation Devices
- External fixation devices are specially designed frames, clamps, Device description: rods, rod-to-rod couplings, pins, posts, fasteners, wire fixations, fixation bolts, washers, nuts, hinges, sockets, connecting bars and screws used for the management of bone fractures and reconstructive, as well as corrective, orthopedic surgery. Materials used include metal alloys, plastic and composites. These materials are chosen to address a wide range of fractures and applications as well as to allow for the appropriate amount of rigidity and stability.
- Intended use: External Fixation Devices are intended to be used for the fixation of supracondylar, or condylar fractures of the femur; for fusion of a joint; for surgical procedures that involve cutting the bone, for fixation of bone fractures; bone reconstruction; as a guide pin for insertion of other implants; or may be implanted through the skin so that a pulling force or traction may be applied to the skeletal system; and others may be used for fixation of bone fractures, for bone reconstructions, as a quide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{1}------------------------------------------------
| Indications<br>statement: | Reprocessed external fixation devices are indicated for use in<br>patients requiring external skeletal fixation and treatment of<br>fractures, osteotomy, arthrodesis, correction of deformities,<br>fracture revision, bone reconstruction procedures, limb<br>lengthening, correction of bony or soft tissue deformities and<br>segmental bony or soft tissue defects. |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>characteristics: | The design, materials, and intended use of the Reprocessed<br>External Fixation Devices are identical to the predicate devices.<br>The mechanism of action of the Reprocessed External Fixation<br>Device is identical to the predicate devices in that the same<br>standard mechanical design, materials, shapes and sizes are<br>utilized. There are no changes to the claims, intended use,<br>clinical applications, patient population, performance<br>specifications, or method of operation. The only change is the<br>modified external fixation devices will be provided non-sterile.<br>Sterilization of the devices will occur in the hospital prior to use. |
| Performance data: | Bench and laboratory testing was conducted to demonstrate<br>performance (safety and effectiveness) of the Reprocessed<br>External Fixation Devices.<br>Validation of reprocessingFunction Testing |
| | Performance testing demonstrates that Reprocessed External<br>Fixation Devices perform as originally intended. |
| Conclusion: | Alliance Medical Corporation concludes that the modified device<br>(the Reprocessed External Fixation Device) is safe, effective<br>and substantially equivalent to the predicate devices, as<br>described herein. |
:
:
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized representation of a human face in profile, composed of three curved lines. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the emblem.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 5 2005
Ms. Moira Barton Regulatory Affairs Manager Alliance Medical Corporation 10232 South 51st Street Phoenix, Arizona 85044
Re: K052064
Trade/Device Name: Alliance Medical Corporation Reprocessed External Fixation Devices Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: KTT Dated: July 19, 2005 Received: July 29, 2005
Dear Ms. Barton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Moira Barton
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter with and hy Jurve FDA finding of substantial equivalence of your device to a legally promance hodicated on "ceresults in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you a tone of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
M. N. Mello
\$\sigma\$
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use Statement 2.
Device Name: Alliance Medical Corporation Reprocessed External Fixation Devices
Indications for Use: Reprocessed external fixation devices are indicated for Indications for OSC: Reprocessed external skeletal fixation and treatment of fractures, use in pationto roquinis, correction of deformities, fracture revision, bone osteolony, antifrodesis, obrootion of hending, correction of bony or soft tissue deformities and segmental bony or soft tissue defects.
Concurrence of CDRH, Office of Device Evaluation (ODE)
2
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
**510(k) Number** K052064
Prescription Use _ (per 21 CFR 801.109) or
Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.