MODIFICATION TO HOFFMANN II MICRO EXTERNAL FIXATION SYSTEM
Applicant
Howmedica Osteonics Corp.
Product Code
KTT · Orthopedic
Decision Date
Jan 20, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Hoffmann® II Micro External Fixation System is an intended to be used with the Half Pins or Transfixing Pins of the Hoffmann® External Fixation System and the Components of the Hoffmann® II External Fixation System. It is intended to be used in the stabilization of open and/or unstable fractures and where soft tissue injury precludes the use of other fracture treatments such as IM rodding, casting and other mcans of internal fixation.
Device Story
External fixation system for bone fracture stabilization. Components include stainless steel hardware and carbon connecting rods. Used in clinical settings by surgeons to stabilize fractures when internal fixation or casting is contraindicated due to soft tissue injury. System functions as a mechanical frame to maintain bone alignment during healing. Provides structural support to stabilize unstable or open fractures.
Indicated for stabilization of open and/or unstable fractures in patients where soft tissue injury precludes alternative treatments like IM rodding or casting.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Hoffmann® II Micro™ External Fixation System (K033476)
Related Devices
K211112 — OIC External Fixation System · Orthopaedic Implant Company · May 5, 2021
K141912 — AOS SMALL BONE EXTERNAL FIXATION, THREADED HALF PINS, AOS SMALL BONE EXTERNAL FIXATION, CARBON FIBER RODS, AOS SMALL BON · Advanced Orthopaedic Solutions, Inc. · Oct 29, 2014
K050048 — HOFFMAN II MICRO EXTERNAL FIXATION SYSTEM · Howmedica Osteonics Corp. · Mar 4, 2005
K183682 — OIC External Fixation System · Orthopaedic Implant Company · Apr 8, 2019
K960014 — PRACTIFIX EXTERNAL FIXATION SYSTEM · Orthopaedic Innovations, Inc. · Mar 18, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
# Summary of Safety and Effectiveness Hoffmann® II Micro™ External Fixation System Line Extension
Proprietary Name: System
Common Name:
Classification Name and Reference
Device Product Code:
For Information contact:
Hoffmann® II Micro™ External Fixation
External Fixation Frame Component
Single/multiple component metallic bone fixation appliances and accessories, 21 CFR 888.3030
87 KTT, 87 LXT & 87 JEC
Vivian Kelly, Regulatory Affairs Specialist Howmedica Osteonics Corp. 59 Route 17 Allendale, NJ 07401-1677 Phone: (201) 831-5581 Fax: (201) 831-6038
Date Summary Prepared:
December 5, 2003
## Description:
This Special 510(k) submission is intended to address modifications to the predicate Hoffmann® II Micro "External Fixation System. The subject and predicate Hoffmann® II Micro" External Fixation Systems are both fabricated from stainless steel components and carbon connecting rods.
## Intended Use:
The Hoffmann® II Micro External Fixation System is an intended to be used with the Half Pins or Transfixing Pins of the Hoffmann® External Fixation System and the Components of the Hoffmann® II External Fixation System. It is intended to be used in the stabilization of open and/or unstable fractures and where soft tissue injury precludes the use of other fracture treatments such as IM rodding, casting and other mcans of internal fixation.
## Substantial Equivalence:
The subject Hoffmann® II Micro " External Fixation System shares the same intended use, and basic design concepts as that of the currently available Hoffmann® II Micro™ External Fixation System. Mechanical testing demonstrated comparable mechanical properties to the predicate components.
{1}------------------------------------------------
Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an abstract caduceus symbol, which is a symbol associated with healthcare and medicine. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 0 2004
Ms. Vivian Kelly Regulatory Affairs Specialist Stryker Howmedica Osteonics, Corp. 59 Route 17 South Allendale, NJ 07401
Re: K033476
Trade/Device Name: Hoffmann® II Micro™ External Fixation System Regulation Number: 21 CFR 888.3030, 21 CFR 888.3040 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories; Smooth or threaded metallic bone fixation fastener, Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Codc: KTT, LXT, JEC Dated: January 6, 2004 Received: January 7, 2004
Dear Ms. Kelly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendnients, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2- Ms. Vivian Kelly
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 8,07.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
l Mark n Wilkerson
elia M. Witten, Ph.D., M.D., M.D Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### 510(k) Number (if known): K033476
Device Name: Hoffmann® II Micro External Fixation System
Indications For Use:
The Hoffmann® II Micro External Fixation System is an intended to be used with the Half Pins or Transfixing Pins of the Hoffinann® External Fixation System and the Components of the Hoffmann® II External Fixation System. It is intended to be used in the stabilization of open and/or unstable fractures and where soft tissue injury precludes the use of other fracture treatments such as IM rodding, casting and other means of internal fixation.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Usc
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
for Mark A. Millman
K033476
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.