K031181 · Smith & Nephew, Inc. · KTT · Jul 14, 2003 · Orthopedic
Device Facts
Record ID
K031181
Device Name
EXTERNAL FIXATION SYSTEMS
Applicant
Smith & Nephew, Inc.
Product Code
KTT · Orthopedic
Decision Date
Jul 14, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
External Fixation System components are intended to be used on adults or pediatric patients as required and are intended to be used for fracture fixation (open and closed); post-traumatic joint contracture which has resulted in loss of range of motion; fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction: pseudoarthrosis or non-union of long bones; limb lengthening by epiphyseal or metaphyseal distraction of bony or soft tissue deformity; correction of segmental bony or soft tissue defects; joint arthrodesis; and management of comminuted intra-articular fractures of the distal radius.
Device Story
External fixation system components, including bar clamps, used for orthopedic fracture management, reconstruction, and corrective surgery. Devices provide mechanical rigidity and stability to bone segments. Operated by surgeons in clinical/OR settings. Components are applied to stabilize fractures, facilitate distraction for limb lengthening, or correct deformities. Output is physical stabilization of bone/joint structures, aiding healing and restoring range of motion. Benefits include improved fracture alignment and management of complex orthopedic conditions.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Metallic bone fixation fasteners and accessories. Materials are metal alloys consistent with predicate devices. Mechanical design provides rigidity and stability for orthopedic fixation. No software or electronic components.
Indications for Use
Indicated for adult and pediatric patients requiring fracture fixation (open/closed), management of post-traumatic joint contracture, fractures/disease causing loss of range of motion, fractures requiring distraction, pseudoarthrosis or non-union of long bones, limb lengthening, correction of bony/soft tissue deformity or segmental defects, joint arthrodesis, and comminuted intra-articular distal radius fractures.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness External Fixation Systems
| Submitted by: | Smith & Nephew, Inc. Orthopaedic Division 1450 Brooks Road Memphis, TN 38116 |
|-------------------------------------|---------------------------------------------------------------------------------|
| Date: | April 14, 2003 |
| Contact Person: | David Henley Senior Clinical / Regulatory Affairs Specialist |
| Proprietary Name: | External Fixation Systems |
| Common Name: | External Fixation Accessories |
| Classification Name and Reference: | Smooth or threaded metallic bone fixation fastener, 21 CFR 888.3040, Class II |
| Device Product Code and Panel Code: | KTT / Orthopedics / 87 |
#### Device Description:
External fixation devices, such as the external fixation bar clamps described herein, are specially designed components used in the management of bone fractures and reconstructive, as well as corrective, orthopedic surgery. The metal alloys used in their manufacture are chosen to address a wide range of applications. These devices have been designed to allow for the appropriate amount of rigidity and stability.
## Intended Use:
External Fixation System components are intended to be used on adults or pediatric patients as required and are intended to be used for fracture fixation (open and closed); post-traumatic joint contracture which has resulted in loss of range of motion; fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction: pseudoarthrosis or non-union of long bones; limb lengthening by epiphyseal or metaphyseal distraction of bony or soft tissue deformity; correction of segmental bony or soft tissue defects; joint arthrodesis; and management of comminuted intra-articular fractures of the distal radius.
## Technological Characteristics:
The principle of operation of these reprocessed devices is very similar to that of the predicates. There are no changes in intended use, performance specifications or method of operation. The reprocessed devices utilize similar designs, the same materials and technological characteristics when compared to the predicate devices.
#### Substantial Equivalence Information:
The intended use of the reprocessed external fixation bar clamps is identical to that of the Smith & Nephew Unilateral Fixator (K994143) and the Smith & Nephew Unilateral Wrist Fixator for Distal Radius Fractures (K994143). The design and the function of the subject devices are very similar to the predicate devices and they are manufactured from the same materials.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Public Health Service
JUL 1 4 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. David Henley Senior Clinical and Regulatory Affairs Specialist Smith & Nephew, Inc. 1450 Brooks Road Memphis, Tennessee 38116
Re: K031181
Trade/Device Name: External Fixation Systems Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: KTT Dated: April 14, 2003 Received: April 15, 2003
Dear Mr. Henley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{2}------------------------------------------------
Page 2 - Mr. David Henley
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Nil R.P. Ogden
Colin M. Witte, Ph.D., M.D., FACS
Celia M. Witten, Ph.D., N Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use Statement
510(k) number (if known): K03118
Device Name: External Fixation Systems
Indications for Use:
External Fixation System components are intended to be used on adults or pediatric patients as required and are intended to be used for fracture fixation (open and closed); post-traumatic joint contracture which has resulted in loss of range of motion; fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction; pseudoarthrosis or non-union of long bones; limb lengthening by epiphyseal or metaphyseal distraction; correction of bony or soft tissue deformity; correction of segmental bony or soft tissue defects; joint arthrodesis; and management of comminuted intra-articular fractures of the distal radius.
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) NRo for cmw
Division of General, Restorative and Neurological Devices
510(k) Number K031181
Prescription Use (Per 21 CFR 801.109) Over-the-Counter Use (Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.