iTotal Identity Cruciate Retaining (CR) Total Knee Replacement, iTotal Identity Posterior Stabilizing (PS) Total Knee Replacement

K203447 · Conformis, Inc. · JWH · Dec 22, 2020 · Orthopedic

Device Facts

Record IDK203447
Device NameiTotal Identity Cruciate Retaining (CR) Total Knee Replacement, iTotal Identity Posterior Stabilizing (PS) Total Knee Replacement
ApplicantConformis, Inc.
Product CodeJWH · Orthopedic
Decision DateDec 22, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Intended Use

The iTotal Identity® Cruciate Retaining (CR) Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The iTotal PS Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartments.

Device Story

Patient-specific, tri-compartmental knee replacement system (CR and PS versions). Inputs: patient imaging scans. Transformation: proprietary and off-the-shelf software designs implants to match patient-specific geometry and anatomy. Output: custom femoral (CoCrMo), tibial (Ti6AL4V-ELI tray, UHMWPE/iPoly® XE insert), and patellar (UHMWPE/iPoly® XE) components. Used in surgical settings by orthopedic surgeons. Ancillary patient-specific manual instruments guide bone cuts and component positioning. Benefits: improved anatomical fit for total knee arthroplasty.

Clinical Evidence

Bench testing only. Physical and chemical characterization compared the subject iPoly® XE material to the predicate material to confirm equivalency. No clinical data presented.

Technological Characteristics

Cemented, semi-constrained knee prosthesis. Materials: CoCrMo (femoral), Ti6AL4V-ELI (tibial tray), UHMWPE or iPoly® XE (patellar/inserts). Patient-specific design via imaging-based software. Includes single-use patient-specific manual instruments and reusable manual instruments. Sterilization methods unchanged from predicates.

Indications for Use

Indicated for patients with knee joint pain and disability due to osteoarthritis, traumatic arthritis, rheumatoid arthritis, polyarthritis, or osteonecrosis; post-traumatic loss of joint function; moderate varus, valgus, or flexion deformity; or failed prior osteotomies, hemiarthroplasties, or partial implants. Suitable for revision procedures if anatomic landmarks are identifiable on imaging. Contraindicated for conditions treatable by single-compartment prostheses. Intended for cemented use only.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ December 22, 2020 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Conformis Inc. Mary Kruitwagen Sr. Regulatory Affairs Specialist 600 Technology Park Drive, 4th Floor Billerica, Massachusetts 01821 # Re: K203447 Trade/Device Name: iTotal Identity® Cruciate Retaining (CR) Knee Replacement System, iTotal Identity® Posterior Stabilizing (PS) Knee Replacement System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH, OOG, OIY Dated: November 19, 2020 Received: November 23, 2020 Dear Mary Kruitwagen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Ting Song, Ph.D., R.A.C. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) #### K203447 Device Name iTotal Identity® Cruciate Retaining Knee Replacement System #### Indications for Use (Describe) The iTotal Identity® Cruciate Retaining (CR) Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include: - · Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid arthritis or osteonecrosis of the knee. - · Post traumatic loss of joint function. - · Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability. - · Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants. - · Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. This implant is intended for cemented use only. Type of Use (Select one or both, as applicable) | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | |-------------------------------------------------------------------------------------------------| | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) #### K203447 Device Name iTotal Identity® Posterior Stabilized Knee Replacement System #### Indications for Use (Describe) The iTotal PS Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartments. The Indications for Use include: · Painful joint disease due to osteoarthritis, traumatoid arthritis, polyarthritis, polyarthritis or osteonecrosis of the knee. · Post traumatic loss of joint function. - · Moderate varus, valgus or flexion deformity. · Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants. · Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. This implant is intended for cemented use only. | Type of Use (Select one or both, as applicable) | <div></div> | |-------------------------------------------------|-------------| |-------------------------------------------------|-------------| X Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # Section 5: 510(k) Summary The following 510(k) Summary is provided as per 21 CFR 807.92 requirements | Subject Device: (Proprietary/Trade name): | iTotal Identity® Cruciate Retaining (CR) Knee Replacement System<br>iTotal Identity® Posterior Stabilized (PS) Knee Replacement System | |----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Usual Name: | Knee Replacement System | | Type of Submission: | Special 510(k) | | Device Class: | II | | Regulation Number: | 888.3560 | | Regulation Description: | Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis | | Primary Product Classification<br>(Product Code) and Description: | JWH<br>Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis | | Secondary Product Classifications<br>(Product Code) and Descriptions: | OOG, OIY<br>Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis | | Reviewing Agency: | Orthopedics | | Date of Summary Preparation: | November 19 2020 | | Submitter's Name and Address: | Conformis Inc.<br>600 Technology Park Drive<br>Fourth Floor<br>Billerica, MA 01821<br>USA | | Telephone Number: | (781) 345-9001 | | Establishment Registration<br>Number(s): | 3009844603<br>3004153240 | | Contact Name: | Mary Kruitwagen | | Contact Telephone: | 781-345-9038 | | Contact email: | Mary.Kruitwagen@conformis.com | | Alternate Contact: | Kara Johnson | | Alternate Contact Telephone: | 781-832-5402 | | Alternate Contact email: | Kara.Johnson@conformis.com | | Primary Predicate Device: | iTotal Identity® Cruciate Retaining (CR) Knee Replacement System | | Primary Predicate Device 510(k): | K190562; August 8, 2019 | | Primary Predicate Device Product<br>Classification (ProCode) and<br>Description: | JWH, OOG, OIY<br>Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis | | Secondary Predicate Device: | iTotal Identity® Posterior Stabilized (PS) Knee Replacement System | | Secondary Predicate Device 510(k): | K201023; June 16, 2020 | | Secondary Predicate Device Classification (ProCode) and Description: | Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis | | Modification | This submission is to seek clearance for use of an alternate chemically and physically equivalent Highly Cross-Linked Vitamin E Stabilized Ultra-High Molecular Weight polyethylene raw material (iPoly® XE) for the manufacture of the patellar component and the tibial inserts. | | Device Description | iTotal Identity® Cruciate Retaining (CR) Knee Replacement System (KRS) and iTotal Identity® Posterior Stabilized (PS) Knee Replacement System (KRS) are patient specific tri-compartmental faceted knee replacements systems. The iTotal Identity CR KRS is a faceted posterior cruciate ligament retaining knee replacement system. It is a semi-constrained, cemented knee implant which consists of femoral, tibial, and patellar components. iTotal Identity® PS KRS is a semi-constrained, cemented knee implant consisting of femoral, tibial, patellar and articular tibial insert components.<br><br>Using patient imaging and a combination of proprietary and off-the-shelf software, a patient-specific implant is designed that best meets the geometric and anatomic requirements of the specific patient. The femoral component is manufactured from cobalt chromium molybdenum (CoCrMo) alloy. The tibial component includes a metal tray and tray keel stem extension manufactured from titanium (Ti6AL4V-ELI) alloy, a tibial tray keel cap manufactured from polyethylene (UHMWPE) and either one or two polyethylene inserts (Two piece is CR only). The polyethylene inserts may be manufactured from either UHMWPE or iPoly® XE (a highly cross-linked vitamin E stabilized UHMWPE). The patellar component is provided in either a round or oval dome shape and may be manufactured from either UHMWPE or iPoly® XE.<br><br>For user convenience, single-use, patient-specific ancillary orthopedic manual surgical instruments designed for use with the proposed iTotal Identity® CR KRS or iTotal Identity® PS KRS are provided to assist in the positioning of the total knee replacement components intraoperatively and in guiding the cutting of bone. In addition, reusable orthopedic manual surgical instruments are provided separately. | | Indications for Use | iTotal Identity® Cruciate Retaining (CR) Knee Replacement System<br>The iTotal Identity® Cruciate Retaining (CR) Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. | {5}------------------------------------------------ {6}------------------------------------------------ The Indications for Use include: - Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid arthritis or osteonecrosis of the knee. - Post traumatic loss of joint function. - Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability. - . Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants. - Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. This implant is intended for cemented use only. ## iTotal Identity® Posterior Stabilized (PS) Knee Replacement System The iTotal PS Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include: - Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid ● arthritis, polyarthritis or osteonecrosis of the knee. - Post traumatic loss of joint function. - Moderate varus, valgus or flexion deformity. - Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or ● bicompartmental implants. - . Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. This implant is intended for cemented use only. ## Technological Characteristics The proposed and predicate devices are cemented knee implants that consist of three primary components; femoral, tibial, and patellar implants. Single-use, patient-specific ancillary surgical instruments are provided for use to assist with surgical implantation. Reusable ancillary surgical instruments, provided in a reusable instrument tray, assist with surgical implantation. The subject devices iTotal Identity Cruciate Retaining (CR) Knee Replacement System and iTotal Identity® Posterior Stabilized (PS) Knee Replacement System are the same as the predicate devices with regards to: - Indications for Use ● - Device design - Device usage - Operating principles - Packaging - Sterilization methods {7}------------------------------------------------ - Biocompatibility - . Shelf-life - . Femoral implant material - Tibial tray Material The modification is to add an alternate iPoly® XE raw material which is a physically and chemically equivalent Highly Cross-linked Vitamin E Stabilized UHMWPE XE to manufacture the - Patellar component - Tibial inserts The manufacture of the subject device is otherwise unchanged from the predicate devices. Performance Data Testing to compare the subject raw material to the predicate material was conducted. The difference between the subject raw material and the predicate raw material is that the subject raw material will not be mechanically annealed. The absence of an annealing step in the processing of the raw material does not have a statistically significant impact on the physical or chemical properties when compared to the predicate material and supports that the materials are equivalent. > In conclusion, the subject raw material raises no new issues regarding safety or effectiveness. The performance of the modified devices is expected to be the same as the currently marketed predicate devices. Substantial The modification to add an additional iPoly® XE raw material applies to the Equivalence patellar component and the tibial inserts. The Assessment of physical and chemical characterization of the predicate and subject materials shows equivalency of the subject material to the predicate material. Results support that the performance of the subject devices would be the same as the predicate devices. > The subject devices, iTotal Identity® Cruciate Retaining (CR) Knee Replacement System and iTotal Identity® Posterior Stabilized (PS) Knee Replacement System, are substantially equivalent to the predicate devices iTotal Identity® Cruciate Retaining (CR) Knee Replacement System and iTotal Identity® Posterior Stabilized (PS) Knee Replacement System. Conclusion Based on the intended use, technological characteristics, and outcome of the performance data the subject devices, iTotal Identity® Cruciate Retaining (CR) Knee Replacement System and iTotal Identity® Posterior Stabilized (PS) Knee Replacement System, are substantially equivalent to the predicate devices iTotal Identity® Cruciate Retaining (CR) Knee Replacement System and iTotal Identity® Posterior Stabilized (PS) Knee Replacement System.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...