iTotal Identity Cruciate Retaining Knee Replacement System
K190562 · Conformis, Inc. · JWH · Aug 8, 2019 · Orthopedic
Device Facts
Record ID
K190562
Device Name
iTotal Identity Cruciate Retaining Knee Replacement System
Applicant
Conformis, Inc.
Product Code
JWH · Orthopedic
Decision Date
Aug 8, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The iTotal Identity Cruciate Retaining (CR) Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include: • Painful joint disease due to osteoarthritis, traumatoid arthritis, rheumatoid arthritis or osteonecrosis of the knee. · Post traumatic loss of joint function. · Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability. · Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants. · Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. This implant is intended for cemented use only.
Device Story
Patient-specific tricompartmental knee replacement system; utilizes patient imaging and proprietary/off-the-shelf software to design implants matching patient-specific anatomy; femoral component (CoCrMo); tibial tray/stem (Ti6AL4V-ELI); tibial keel cap (UHMWPE); polyethylene inserts (UHMWPE or iPoly XE); patellar component (UHMWPE or iPoly XE). Used in clinical settings by orthopedic surgeons for total knee arthroplasty. Includes patient-specific single-use ancillary instruments and reusable manual instruments to assist in component positioning and bone cutting. Output is a custom-fit implant system; aids in restoring joint function and stability; reduces pain; improves mobility for patients with severe knee disease.
Clinical Evidence
Bench testing only. Testing included cadaveric instrumentation/usability, insert interlock/modularity, tibial tray stem fatigue, tibial tray stem assembly/disassembly torque, tibial tray/stem particle analysis/fretting corrosion, biocompatibility, MRI compatibility, patella test justification, tibial insert justification (contact area, range of motion, constraint), sterilization validation, and reusable instrument cleaning/sterilization validation.
Technological Characteristics
Semi-constrained, cemented knee prosthesis. Materials: CoCrMo (femoral), Ti6AL4V-ELI (tibial tray/stem), UHMWPE/iPoly XE (inserts/patella). Patient-specific design via imaging/software. Ancillary instruments: single-use (nylon, stainless steel, R-Radel) and reusable. Sterilization: Sterile (SAL 10^-6) for implants/single-use instruments; non-sterile for reusable instruments.
Indications for Use
Indicated for patients with knee joint pain and disability requiring total knee replacement due to osteoarthritis, traumatic/rheumatoid arthritis, osteonecrosis, post-traumatic loss of function, moderate varus/valgus/flexion deformity, or failed prior partial/hemi-arthroplasty/osteotomy. Revision procedures allowed if anatomic landmarks are identifiable on imaging. Cemented use only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Total Joint Orthopedics, Inc., Klassic Knee System (K180159)
Conformis iTotal Posterior Stabilized (PS) Knee Replacement System (K161668)
Submission Summary (Full Text)
{1}------------------------------------------------
August 8, 2019
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Conformis, Inc. Paul Smolenski Sr. RA Manager 600 Technology Park Drive, 4th Floor Billerica, Massachusetts 01821
Re: K190562
Trade/Device Name: iTotal Identity Cruciate Retaining Knee Replacement System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, OOG, OIY Dated: July 9, 2019 Received: July 10, 2019
Dear Paul Smolenski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For CAPT Raquel Peat, PhD, MPH, USPHS Director OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K190562
Device Name
Conformis iTotal Identity Cruciate Retaining (CR) Knee Replacement System
Indications for Use (Describe)
The iTotal Identity Cruciate Retaining (CR) Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.
The Indications for Use include:
• Painful joint disease due to osteoarthritis, traumatoid arthritis, rheumatoid arthritis or osteonecrosis of the knee.
· Post traumatic loss of joint function.
· Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
· Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants.
· Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.
This implant is intended for cemented use only.
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
Prescription Ooc (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
| | This section applies only to requirements of the Paperwork Reduction Act of 1995. |
|--|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <b>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*</b> |
| | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: |
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Conformis. The logo consists of a stylized letter "C" in orange and black, with the word "CONFORMIS" written below it. The first four letters of the word are in black, while the last four letters are in orange. The logo is simple and modern, and the colors are eye-catching.
# 510K SUMMARY (page 1 of 5)
| Submitter's Name<br>and Address: | Conformis, Inc.<br>600 Technology Park Drive<br>Billerica, MA 01821 |
|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration<br>Number(s): | 3009844603 and 3004153240 |
| Date Summary was<br>Prepared: | March 4th, 2019 |
| Contact Person: | Paul Smolenski<br>Sr. Regulatory Affairs Manager<br>Telephone: 781-374-5586<br>Fax: 781-345-0147<br>E-mail: paul.smolenski@conformis.com |
| Trade/Device Name: | Conformis itotal Identity® Cruciate Retaining (CR) Knee<br>Replacement System (iTotal CR KRS) |
| Common Name: | Total Knee Replacement System |
| Device Class: | Class II |
| Regulation Number<br>and Description: | 21 CFR 888.3560<br>Knee joint patellofemorotibial polymer/metal/polymer semi-<br>constrained cemented prosthesis. |
| Classification Name: | Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented,<br>Polymer/Metal/Polymer |
| Product Code:<br>(Classification) | JWH |
| Product Code(s):<br>(Subsequent) | OOG: Knee Arthroplasty Implantation System<br>OIY: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented,<br>Polymer + Additive/Metal/Polymer + Additive |
| Legally Marketed<br>Predicate Device<br>(Primary Predicate): | Conformis iTotal® Cruciate Retaining Knee Replacement System<br>("iTotal CR KRS"), K180906, cleared May 16, 2018). |
| Legally Marketed<br>Predicate Device | Total Joint Orthopedics, Inc., Klassic® Knee System (K180159,<br>March 9, 2018), hereafter referred to as "TJO Klassic Knee." |
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Conformis. The logo consists of a stylized letter "C" in orange and black, with the word "CONFORMIS" written in black and orange below it. The "CON" is in black, and the "FORMIS" is in orange.
# 510K SUMMARY (page 2 of 5)
Conformis iTotal Posterior Stabilized (PS) Knee Replacement System (iTotal PS KRS), K161668, cleared July 15, 2016
Device Description: The iTotal Identity® Cruciate Retaining Knee Replacement System (hereafter referred to as the "iTotal Identity CR KRS") is a patient specific tricompartmental faceted posterior cruciate ligament (PCL) retaining knee replacement system. The iTotal Identity CR KRS is a semi-constrained, cemented knee implant which consists of femoral, tibial, and patellar components.
> Using patient imaging and a combination of proprietary and off the shelf software, a patient-specific implant is designed that best meets the geometric and anatomic requirements of the specific patient. The femoral component is manufactured from cobalt chromium molybdenum (CoCrMo) alloy. The tibial component includes a metal tray and tray keel stem extension manufactured from titanium (Ti6AL4V-ELI) alloy, a tibial tray keel cap manufactured from polyethylene (UHMWPE), and either one or two polyethylene inserts. These inserts may be manufactured from either UHMWPE or iPoly® XE (a highly cross-linked Vitamin E infused polyethylene). The patellar component is provided in either a round or oval dome shape and may be manufactured from either UHMWPE or iPoly® XE.
For user convenience, and similar to the primary predicate iTotal CR KRS, single-use, patient-specific ancillary orthopedic manual surgical instruments designed for use with the proposed iTotal Identity CR KRS are provided to assist in the positioning of the total knee replacement components intraoperatively and in guiding the cutting of bone. In addition, reusable orthopedic manual surgical instruments are provided separately.
Indications for Use: The iTotal Identity® Cruciate Retaining (CR) Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.
The Indications for Use include:
- · Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid arthritis or osteonecrosis of the knee.
- · Post traumatic loss of joint function.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Conformis. The logo consists of a stylized letter "C" in orange and black, with the word "CONFORMIS" written below it. The first four letters of the word are in black, while the last four letters are in orange. The logo is simple and modern, and it is likely used to represent the company's brand.
# 510K SUMMARY (page 3 of 5)
- Moderate varus, valgus or flexion deformity in which the . ligamentous structures can be returned to adequate function and stability.
- · Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants.
- Revision procedures provided that anatomic landmarks . necessary for alignment and positioning of the implant are identifiable on patient imaging scans.
## This implant is intended for cemented use only.
Summary of Technological Characteristics: The rationale for substantial equivalence is based on consideration of the following device use and characteristics
#### Intended Use:
The proposed iTotal Identity CR KRS and predicate devices are intended to be used as total knee replacements. The implantable components are intended for single-use only in a single patient.
#### Indications for Use:
The proposed iTotal Identity CR KRS and the primary predicate iTotal CR KRS have identical indications for use.
#### Operating Principal/Fundamental Technology:
The proposed and predicate devices are semi-constrained, cemented knee implants that consist of three primary components; femoral, tibial, and patellar.
Single-use, patient-specific ancillary surgical instruments are provided for use with both the primary predicate iTotal CR KRS and proposed iTotal Identity CR KRS to assist with surgical implantation.
Reusable ancillary surgical instruments, provided in a reusable instrument tray, are used with the predicate and proposed devices. These instruments assist with surgical implantation.
#### Design:
The proposed and predicate devices have the same basic design characteristics as they are comprised of three primary components: femoral, tibial, and patellar. The proposed iTotal ldentity CR KRS offers a tibial tray stem extension and tibial tray keel cap, which are features that are also offered with the predicate TJO Klassic Knee.
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Conformis. The logo consists of a stylized letter "C" in orange and black, with the word "CONFORMIS" written below it. The letters "O", "R", and "M" in the word "CONFORMIS" are orange, while the rest of the letters are black.
#### 510K SUMMARY (page 4 of 5)
#### Materials:
The same biocompatible materials (CoCrMo, UHMWPE, and iPoly®XE) are used to manufacture the femoral, tibial insert, and patellar components of both the proposed iTotal Identity CR KRS and primary predicate iTotal CR KRS.
Titanium alloy (Ti6AL4V-ELI), is used to manufacture the tibial tray and tibial stem extension of the proposed iTotal Identity CR KRS. The tibial tray and stem extension of the predicate TJO Klassic Knee are also manufactured from titanium alloy (Ti6AL4V).
The tibial tray keel cap of the proposed iTotal Identity CR KRS is made from the same biocompatible material (UHMWPE) that is currently used to manufacture tibial inserts and patellar implants of the primary predicate device.
The same biocompatible materials (nylon, stainless steel and R-Radel) are used to manufacture the single-use and reusable surgical instrumentation of both the primary predicate and proposed iTotal Identity CR KRS.
#### Sterilization:
The implant components of both the proposed iTotal Identity CR KRS and predicate devices are provided sterile in double pouches (SAL 1.0 x 10-6).
The single-use surgical instruments used with the proposed and primary predicate device are provided sterile in double pouches (SAL 1.0 x 10-6).
The reusable surgical instruments used with the proposed and predicate devices are provided non-sterile.
Substantial The iTotal Identity CR KRS, subject of this premarket notification, is substantially equivalent to the following Equivalence predicate devices:
> Conformis iTotal CR KRS (K180906. cleared May 16, 2018).
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Conformis. The logo consists of a stylized letter "C" in orange and black, with the word "CONFORMIS" written in black and orange below it. The first part of the word, "CON," is in black, while the second part, "FORMIS," is in orange.
### 510K SUMMARY (page 5 of 5)
Conformis iTotal PS KRS (K161668, cleared July 15, 2016)
TJO Klassic Knee Svstem (K180159, cleared March 9, 2018)
Substantial equivalence is based on intended use and the technological characteristics of the proposed iTotal Identity CR KRS, as well as non-clinical testing conducted to confirm that proposed iTotal Identity CR KRS is substantially equivalent to the predicate devices. Specifically, the following testing was performed:
- Cadaver (instrumentation testing & usability) ●
- Insert interlock/Modularity
- Tibial tray stem fatigue
- Tibial tray stem assembly/disassembly ● torque
- Tibial tray and stem extension particle ● analysis/fretting corrosion
- Biocompatibility
- MRI Compatibility ●
- Patella test justification
- Tibial Insert justification (contact area, range of motion, constraint testing).
- . Sterilization validation
- . Reusable instrument cleaning & sterilization (new tray and new instruments)
All testing has demonstrated that the proposed iTotal Identity CR KRS is substantially equivalent to the predicate devices.
Conclusion Based on intended use, technological characteristics, and testing conducted, it is concluded that the proposed iTotal Identity CR KRS is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.