T-PIN

K053513 · Union Surgical, LLC · JDW · Jan 13, 2006 · Orthopedic

Device Facts

Record IDK053513
Device NameT-PIN
ApplicantUnion Surgical, LLC
Product CodeJDW · Orthopedic
Decision DateJan 13, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Union Surqical, LLC implantable T-Pin® is intended for the repair of distal radius fractures, proximal ulna fractures, and comminuted wrist fractures without intraarticular fracture gapping.

Device Story

T-Pin® is a threaded, cannulated stainless steel Steinmann pin designed for surgical repair of wrist fractures; features break-off shank. Used by surgeons in clinical settings for bone fixation. Device provides mechanical stabilization of fractures. Optionally supplied with extraction tools and sterilization tray. Benefits patient by facilitating fracture repair.

Clinical Evidence

Bench testing only. Compliance with ASTM F366-82, ASTM F138-97, and ASTM F899-95 standards demonstrated through mechanical performance comparisons.

Technological Characteristics

Cannulated Steinmann pin; stainless steel (UNS S31673 per ASTM F138-97); includes break-off shank. Complies with ASTM F366-82 and ASTM F899-95. Mechanical fixation device.

Indications for Use

Indicated for patients requiring repair of distal radius fractures, proximal ulna fractures, and comminuted wrist fractures without intraarticular fracture gapping.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JAN 1 3 2006 K0535/3 p1/2 Page 17 of 18 # Summary of Safety and Effectiveness for the T-Pin® submitted by Union Surgical, LLC 834 Chestnut Street Philadelphia, PA 19107 Phone: (215)-521-3009 | Contact Person: | Al Weisenborn | |----------------------|--------------------------------------------------------------------------| | Device Trade Name: | T-Pin® | | Common Name: | Cannulated Fixation Screw | | Classification Name: | Smooth or threaded metallic bone fixation fastener per 21 CFR § 888.3040 | ### Identification of a Legally Marketed Predicate Device The Union Surgical, LLC T-Pin® is substantially equivalent to Taras Threaded Fixation Pin that is legally marketed and distributed by Union Surgical, LLC pursuant to K032532. ### Device Description The Union Surgical. LLC T-Pin® is a threaded and cannulated stainless steel Steinmann pin used for the repair of wrist fractures. The device incorporates a break-off shank. The device is ontionally supplied with removal tools that permit pin extraction and a sterilization tray. ### Intended Use The Union Surgical, LLC implantable T-Pin® is intended for the repair of distal radius fractures, proximal ulna fractures, and comminuted wrist fractures without intraarticular fracture gapping. ### Summary of Technological Characteristics A 13-point comparison of technological characteristics of the Union Surgical, LLC T-Pin® and the Taras Threaded Fixation Pin was performed. The devices were found to be substantially equivalent. # Summary of Performance Data The Union Surgical. LLC T-Pin® complies with the following standards, practices, and guidances: - . ASTM F366 82 (Reapproved 2000), Standard Specification for Fixation Pins and Wires {1}------------------------------------------------ - . ASTM F138 - 97, Standard Specification for Wrought 18 Chromium-14 Nickel-2.5 Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673) - . ASTM F899 - 95, Standard Specification for Stainless Steel Billet, Bar and Wire for Surgical The Union Surgical, LLC T-Pin® is substantially equivalent to Taras Threaded Fixation Pins that are legally marketed and distributed by Union Surgical, LLC. This has been demonstrated through a 13-point technological comparison of features and a 3-parameter comparison of mechanical performance. The Implantable and tissue contact materials used to fabricate the T-Pin® and Instruments have a long history of safe usage in medical devices. Since the Union Surgical, LLC T-Pin® meets the requirements of the stated standards and embody technological characteristics essentially identical to the predicate device, we believe the device is safe and effective and performs as well as, or better than, the predicate device. The T-Pin® will be manufactured per specifications using good manufacturing practices that ensure the device is safe and effective for its intended use. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or a stylized human figure. દ્રિક Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 1 3 2006 John Taras, M.D. President Union Surgical, LLC 834 Chestnut Street, Suite G- 114 Philadelphia, PA 19107 Re: K053513 Trade/Device Name: T-Pin® Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: JDW Dated: December 12, 2005 Received: December 16, 2005 Dear Dr. Taras: We have reviewed your Section 510(k) premarket notification of intent to market the device we neve reviewed your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for assured in to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ ### Page 2 - John Taras, M.D. forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, signature Mark N. Melkerson for Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indications for Use - Indications for Users K053513 510(k) Number (if known): _ Device Name: T-Pin® Indications for Use: The Union Surqical, LLC implantable T-Pin® is intended for the repair of distal radius fractures, proximal ulna fractures, and comminuted wrist fractures without intraarticular fracture gapping. Prescription Use_ X (Per 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices Page 1 of 1 510(k) Number_ Los 353 Union Surgical, LLC 834 Chestnut Street, Philadelphia, PA 19107 Revised: January 5, 2006
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...