The MedShape DynaClip Bone Staple is intended to be used for the following: Fracture and osteotomy fixation and joint arthrodesis of the hand and foot. Fixation of proximal tibial metaphysis osteotomy. Fixation of small bone fragments (i.e. small fragments of bone which are not comminuted to the extent to preclude staple placement). These fragments may be located in long bones such as the femur, fibula and tibia in the lower extremities; the humerus, ulna or radius in the upper extremities; the clavicle and ribs; and in flat bone such as the pelvis, scapula and sternum.
Device Story
DynaClip Bone Staple is a sterile, single-use orthopedic implant; constructed from Nickel-Titanium (NiTiNOL) alloy. Device utilizes shape memory properties to provide interference fit and sustained compression across fracture sites or osteotomies to promote bone fusion. Staple is provided pre-loaded on a disposable inserter for surgical application. Used by surgeons in clinical settings for fixation of small bone fragments and joint arthrodesis. Output is mechanical stabilization of bone segments; aids in healing by maintaining compression. Benefits include secure fixation of bone fragments and promotion of fusion.
Clinical Evidence
No clinical data was collected. Evidence consists of bench testing including pull-out fixation strength, elastic static bending strength, constant amplitude bending fatigue strength (ASTM F564), corrosion resistance (ASTM F2129), bacterial endotoxin testing (ANSI/AAMI ST72:2011), and Finite Element Analysis.
Technological Characteristics
Material: NiTiNOL (ASTM F2063). Principle: Interference fit/shape memory compression. Form factor: Bone staple, pre-loaded on disposable inserter. Sterilization: Gamma irradiation. Connectivity: None (mechanical implant).
Indications for Use
Indicated for patients requiring fracture/osteotomy fixation, joint arthrodesis, or fixation of small bone fragments in the hand, foot, long bones (femur, fibula, tibia, humerus, ulna, radius), clavicle, ribs, or flat bones (pelvis, scapula, sternum).
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
MedShape, Inc. Jack Griffis Sr. VP, Advanced Research 1575 Northside Drive NW, Suite 440 Atlanta, Georgia 30318
Re: K193305
Trade/Device Name: DynaClip Bone Staple Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: JDR Dated: March 20, 2020 Received: March 25, 2020
Dear Jack Griffis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
April 23, 2020
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
FOR Vesa Vuniqi Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| Form Approved: OMB No. 0910-0120 | DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|----------------------------------|-----------------------------------------|
| Expiration Date: 06/30/2020 | Food and Drug Administration |
| See PRA Statement below. | <b>Indications for Use</b> |
| 510(k) Number ( <i>if known</i> ) | K193305 |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | DynaClip Bone Staple |
| Indications for Use (Describe) | The MedShape DynaClip Bone Staple is intended to be used for the following: <ul><li>Fracture and osteotomy fixation and joint arthrodesis of the hand and foot.</li><li>Fixation of proximal tibial metaphysis osteotomy.</li><li>Fixation of small bone fragments (i.e. small fragments of bone which are not comminuted to the extent to preclude staple placement). These fragments may be located in long bones such as the femur, fibula and tibia in the lower extremities; the humerus, ulna or radius in the upper extremities; the clavicle and ribs; and in flat bone such as the pelvis, scapula and sternum.</li></ul> |
| Type of Use (Select one or both, as applicable) | <div><span style="display:inline-block; width:15px;"><input checked="" type="checkbox"/></span>Prescription Use (Part 21 CFR 801 Subpart D)</div> <div><span style="display:inline-block; width:15px;"><input type="checkbox"/></span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
| FORM FDA 3881 (7/17) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 | EF |
|----------------------|-------------|----------------------------------------|----|
|----------------------|-------------|----------------------------------------|----|
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the MedShape logo. The logo consists of a blue abstract shape on the left and the word "MEDSHAPE" in gray on the right. The abstract shape is a combination of a square and a circle, with the square positioned above the circle.
## 510(k) Summary
510(k) Number: K193305
March 20th, 2020 Date Submitted:
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
- A. Submitter: MedShape, Inc. 1575 Northside Drive, Suite 440 Atlanta, Georgia 30318 (877) 343-7016
- B. Company Contact: Jack Griffis SVP, Advanced Research (678) 235-3311 (direct) (404) 249-9158 (fax) jack.griffis@medshape.com
- C. Device Information: Trade Name: DynaClip Bone Staple Common Name: Staple, Fixation, Bone
- D. Classification: Orthopedic Panel JDR, 21 CFR 888.3030 Class II
- ட். Predicate Device: BioMedical Enterprises (BME), Inc. (now DePuySynthes), Elite™ Implant, K150125 BioMedical Enterprises (BME), Inc. (now DePuySynthes), Speed™, Speed Arc™, Speed Titan™ Implant, K142292
- F. Reference Predicate Device(s): MedShape, Inc., DynaClip Implant, K181781
- G. Physical Description:
The proposed MedShape DynaClip Bone Staple is a sterile, single use orthopedic implant designed to use the principles of interference fit to hold the implant body into a predrilled hole and across the target fracture site. The DynaClip Bone Staple is intended to be used for fracture and osteotomy fixation, including joint arthrodesis, and fixation of small bone fragments, and is comprised of Nickel Titanium alloy commonly referred to as NiTiNOL (reference ASTM F2063, Standard Specification for Wrought Nickel-Titanium Shape Memory Alloys for Medical Devices and Surgical Implants). The staple is provided pre-loaded on a disposable inserter.
{4}------------------------------------------------
## H. Indications for Use:
The MedShape DynaClip Bone Staple is intended to be used for the following:
- . Fracture and osteotomy fixation and joint arthrodesis of the hand and foot.
- Fixation of proximal tibial metaphysis osteotomy. .
- . Fixation of small bone fragments (i.e. small fragments of bone which are not comminuted to the extent to preclude staple placement). These fragments may be located in long bones such as the femur, fibula and tibia in the lower extremities; the humerus, ulna or radius in the upper extremities; the clavicle and ribs; and in flat bone such as the pelvis, scapula and sternum.
- l. Comparison of Characteristics:
The MedShape DynaClip Bone Staple was designed to have equivalent technological characteristics as the BioMedical Enterprises (BME), Inc. (now DePuy-Synthes), Elite™ Implant cleared under K150125 and the Speed™, Speed Arc™, Speed Titan™ Implant, cleared under K142292. They all have the same materials (NiTiNOL construction as per ASTM 2063, Standard Specification for Wrought Nickel-Titanium Shape Memory Alloys for Medical Devices and Surgical Implants), method of operation (pre-loaded on an inserter), method of sterilization (gamma irradiation) and operate under the same principles of interference fixation and sustained compression to promote fusion of bone. The devices differ only by cross-section and physical sizes offered.
NOTE: The proposed DynaClip device is identical to the previously cleared device (under K181781), except for the addition of some configurations with 4-legs, and this traditional Premarket Notification is primarily for expanded indications for use.
## J. Substantial Equivalence – Non-Clinical Testing:
The proposed device was subjected to the following performance tests to support the assertion of substantial equivalence:
- Bench testing per ASTM F564, Standard Specification and Test Methods for . Metallic Bone Staples;
- o Pull-out fixation strength
- Elastic static bending strength o
- Constant amplitude bending fatigue strength o
- . Corrosion resistance testing per ASTM F2129, Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements to Determine the Corrosion Susceptibility of Small Implant Devices
- . Bacterial endotoxin testing based on an endotoxin limit of 20EU/device per ANSI/AAMI ST72:2011
- Mechanical comparison via Finite Element Analysis
- No clinical data was collected, as it was deemed unnecessary to support the 0 assertion of safety
Image /page/4/Picture/19 description: The image contains the logo for MedShape. The logo consists of a blue abstract shape resembling interconnected circles and a square, followed by the word "MEDSHAPE" in a gray, sans-serif font. The word is slightly stylized, with a registered trademark symbol next to it.
{5}------------------------------------------------
## Substantial Equivalence – Conclusions: K.
No new questions of safety or effectiveness were identified during device testing; therefore, the DynaClip Bone Staple is considered substantially equivalent to the predicate device. In addition, MedShape asserts that any differences from the predicate devices do not affect safety or effectiveness.
Jack Griffis
SVP, Advanced Research
(404) 678-3311(direct)
(404) 249-9158 (fax) jack.griffis@medshape.com
Image /page/5/Picture/5 description: The image shows the logo for MedShape. The logo consists of a blue abstract shape on the left and the word "MEDSHAPE" in gray on the right. The abstract shape is a combination of a square and a circle. The word "MEDSHAPE" is written in a sans-serif font.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.