K972327 · Pacific Materials and Interfaces · JDQ · Aug 11, 1997 · Orthopedic
Device Facts
Record ID
K972327
Device Name
METAGEN ACTIVELOCK CERCLAGE SYSTEM
Applicant
Pacific Materials and Interfaces
Product Code
JDQ · Orthopedic
Decision Date
Aug 11, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3010
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Repair of long bone fractures due to trauma or reconstruction Reattachment of the greater trochanter in total hip arthroplarty, surface replacement arthroplasty, or other procedures involving trochanteric osteotomy
Device Story
System secures/stabilizes bone segments during orthopedic trauma/joint reconstruction. Components: 4.5mm wide metal band (Ti-6Al-4V or Nitinol), titanium sleeve, and super-elastic Nitinol clamp. Surgeon passes band around bone; threads through clamp stretched by manual cerclage applicator instrument. Tensioning device integral to applicator achieves fracture reduction. Releasing stretching force allows Nitinol clamp to retract/lock band securely. Used in OR by surgeons. Benefits: provides positive locking mechanism exceeding strength of traditional twisted wire assemblies; utilizes super-elastic properties for secure fixation.
Clinical Evidence
Bench testing only. Static tensile tests compared Metagen band/Nitinol clamp system against 16 and 18 gauge AISI Type 316L stainless steel cerclage wires. Results demonstrated that the Metagen system provides tensile strength equal to or greater than clinically accepted monofilament stainless steel cerclage wire.
Technological Characteristics
Materials: Ti-6Al-4V ELI (ASTM F-136) or Nitinol (Ni-Ti shape memory alloy). Sensing/Actuation: Super-elastic effect of Nitinol clamp provides locking force. Form factor: 4.5mm x 250mm band, titanium sleeve, Nitinol clamp. Sterilization: Gamma irradiation (25 kG minimum) per ANSI/AAMI ST32-1991. Connectivity: None (manual surgical instrument system).
Indications for Use
Indicated for patients requiring bone fixation for long bone fractures due to trauma or reconstruction, or reattachment of the greater trochanter during total hip arthroplasty, surface replacement arthroplasty, or other trochanteric osteotomy procedures.
Regulatory Classification
Identification
A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.
K983976 — METAGEN ACTIVELOCK WIRE CERCLAGE SYSTEM · Metagen, LLC · Dec 30, 1998
K151888 — Frontier Medical Devices In-line Orthopedic Cable Cerclage System · Frontier Medical Devices, Inc. · Oct 30, 2015
K153242 — Frontier Medical Devices In-Line Orthopedic Cable Cerclage System · Frontier Medical Devices, Inc. · Jan 13, 2016
K030256 — ISO-ELASTIC CERCLAGE SYSTEM · Kinamed, Inc. · Oct 21, 2003
K993903 — WHITESIDE BIOMECHANICS TITANIUM LOW-PROFILE CABLE, MODEL BIO-1-1013 · Whiteside Biomechanics, Inc. · Feb 14, 2000
Submission Summary (Full Text)
{0}------------------------------------------------
# 14972327
## AUG | | 1997
# 510(k) SUMMARY METAGEN ACTIVELOCK™ CERCLAGE SYSTEM
June 13, 1997
#### ADMINISTRATIVE INFORMATION
Manufacturer Name
Metagen, LLC 428 Technology Drive East Menomonie, Wisconsin 54751
Official Contact Person
Wesley D. Johnson President Telephone (715) 232-4880 Fax (715) 235-9570
Representative/Consultant
Floyd G. Larson Pacific Materials and Interfaces 4329 Graydon Road San Diego, CA 92130 Telephone (619) 792-1235 FAX (619) 792-1236
## DEVICE NAME
Classification Name Bone Fixation Cerclage Metagen Activelock™ Cerclage System Trade/Proprietary Name Common Name Cerclage Band System
{1}------------------------------------------------
## ESTABLISHMENT REGISTRATION NUMBER
Metagen, LLC is registered with FDA under Establishment Registration Number 55922.
#### DEVICE CLASSIFICATION
Bone fixation cerclage systems have been classified by FDA as Class II devices, as shown in 21 CFR § 888.3010. The device is reviewed by the Orthopedic and Rehabilitation Devices Panel and the Product Code for the device is JDQ. The cerclage applicator instrument is described in this Notification in order to assist in explaining the operation of the clamping system, but it is not intended to be a subject of the submission. It is an orthopedic manual surgical instrument (Product Code HXN), classified by FDA as a Class I device and exempt from Premarket Notification, according to 21 CFR § 888.4540.
#### CONFORMANCE WITH PERFORMANCE STANDARDS
No performance standards have been established under Section 514. Voluntary standards with which the Metagen Activelock 114 Cerclage System complies include American Society for Testing and Materials (ASTM) designation F-136 (Standard Specification for Wrought Titanium 6Al-4V ELI Alloy for Surgical Implant Applications) and American National Standards Institute (ANSI)/Association for the Advancement of Medical Instrumentation (AAMI) Process Control Guidelines for Gamma Radiation Sterilization (ANSI/AAMI ST32-1991).
#### PACKAGING/LABELING/PRODUCT INFORMATION
The Metagen Activelock™ Cerclage System bone bands and clamps will be packaged in a radiation sterilizable, disposable pouch or tray. The band/clamp assembly will be held in a support tray for easy loading onto the cerclage applicator instrument. The sterilizable pouch or tray will be packaged inside a disposable paper box. Product will be provided either non-sterile (appropriately labeled) or sterile. Sterilization will be accomplished by means of Co" gamma irradiation at a dose of 25 kG (2.5 Mrad) minimum. Sterilization will be validated by the bioburden method, using American National Standards Institute (ANSI)/Association for the Advancement of Medical Instrumentation (AAMI) Process Control Guidelines for Gamma Radiation Sterilization (ANSI/AAMI ST32-1991). The sterility assurance level (SAL) that Metagen intends to meet for the sterile version of the Metagen Activelock™ Cerclage System is 10 . The device is not represented to be "pyrogen free." The cerclage applicator instrument will be packaged separately, non-sterile, in heat-sealed pouches.
{2}------------------------------------------------
#### Metagen Activelock Ccrclage System 510(k) Summary
#### INTENDED USE
The Metagen Activelock TM Cerclage System is intended for
- repair of long bone fractures due to trauma or reconstruction .
- reattachment of the greater trochanter in wull hip arthroplasty, surface replacement arthroplasty, . or other procedures involving trochanteric ustoulumy
#### DEVICE DESCRIPTION
#### Design Characteristics
The Metagen Activelock™ Cerclage System is used in the management of orthopedic trauma and total joint reconstruction to secure and stabilize segments of bone. The fundamental components of the system are a 4.5 mm wide x 250 mm long metal band, a titanium slecve which slides onto the band, and a Nitinol clamp which secures the band. There are two metal band designs in the system; a .30 mm thick Ti 6Al-4V band, and a .Smin thick Nitinol band. Both bands in this system are secured with the same Nitinol clamp. The sleeve is a low profile titanium block with slots to accept the band and sharp tines which project into the bony surface. The Nitinol clamp component provides a positive method of securing the band. When vicwed from the top the clamp is an open. rectangular block. A side view shows the through slot that uccepts and ultimately secures the band. Because of the super-elastic properties of the Niinol material, the clamp component can be physically stretched along its with a cerclage applicator instrument . The cerclage applicator instrument device by design limits the elongation of the slot well within the clamp's elastic range. The clamp is in no danger of undergoing permanent (plastic) deformation.
When the slot is elongated, the band fits easily through the slot. Upon release of the stretching force, the clamp retracts toward its resting position and engages the hand with sufficient force to lock the band securely in place. Mechanical testing of the assembled device comparing the band in both titanium and Nitinol to 18 and 16 gage twisted monofilament wire assemblies shows that the Activelock ™ system exceeds the strength of the twisted wire assemblies.
#### General Operative Technique
The system is clinically applied to fractured bone segments in the same manner as currently marketed wires or cables. The "Icader" end of the band is passed around the bone segments with the aid of a blunt suture which is tied through a hole in the band. The optional titanium sleeve may be placed onto the band at this time or carlier. The tip of the band is then threaded
{3}------------------------------------------------
#### 510(k) Summary
through the clamp after the clamp has been stretched by the cerclage applicator instrument. Tensioning of the band to cause bony reduction is achieved by securing the leader end of the band to a tensioning device which is integral with the stretcher device. Tensioning can occur only when the clamp is in the stretched position. When adequate fracture reduction and alignment are obtained, the Nitinol clamp is released, securing the band in place. Typically, several band assemblies are used to stabilize a single fracture.
#### Material Composition
The band of the Metagen Activelock™ Cerclage System is made from either titanium alloy (Ti-6Al-4V ELI) conforming to ASTM designation F-136 (Standard Specification for Wrought Titanium 6A1-4V ELI Alloy for Surgical Implant Applications), or from Nitinol (shape memory alloy). The use of titanium alloy is widespread in commercially distributed, permanently implanted medical devices and the material is widely considered to be biocompatible, even being used as a negative control in biocompatibility testing.
Nitinol is an alloy consisting of nearly equal atomic percentages of nickel and titanium. The name is derived from its major elements (Ni.Ti) and the fact that it was originally developed at the US Naval Ordnance Laboratory (NOL). This is one of several types of alloy commonly referred to as a shape memory alloy (SMA). SMAs are unique in that they can exhibit dramatically different properties than do conventional alloys or metals.
One unique property is the shape memory effect in which an SMA can be plastically deformed into a desired shape then restored to its original shape with the application of heat. Another unique property is the superelastic effect in which an SMA can be mechanically deformed to a large strain (i.e. 8%). returning to its original shape upon removal of the deforming force. Although the composition of the SMA is similar for both effects previously mentioned, the temperature ranges in which the effects occur vary. The Metagen Activelock™ Cerclage System uses the superelastic effect to provide the clamping force needed to securely lock the band in place. The temperature sensitive shape memory property of the material is not utilized.
The TiO, oxide layer found on Nitinol leads to its excellent biocompatibility and corrosion resistance. Polarization testing of Nitinol, stainless steel, and titanium in Hank's solution has demonstrated that the corrosion resistance of Nitinol is between that of stainless steel and titanium alloy. Also, in vivo testing has demonstrated the biocompatiblity of Nitinol in animals, indicating that Nitinol is more biocompatible than stainless steel.235
<sup>1</sup> Speck K, and Fraker A, "Anodic Polarization Behavior of Ti-Ni and Ti-6Al-4V in Simulated Physiological Solutions," J Dent Res, 59(19):1590, 1980.
<sup>4</sup> Castleman, IS and Motzkin, SM, "The Biocompatibility of Nitinol," Biocompatibility of
Clinical Implant Materials, vol. 1, Williams DF (ed), CRC Press, pp129-154, 1981.
<sup>3</sup> Duerig, TW, et al., "Superelastic Nitinol for Medical Devices," Medical Plastics and Biomaterials, March/April, pp 30-43, 1997.
{4}------------------------------------------------
#### Metagen Activelock™ Cerclage System Page S 510(k) Summary
Nitinol has been used since the early 1980s for permanent implantable devices in Japan, Germany, China and Russia * The Nirinol material used in this device is the same in composition as Nitinol shape memory alloys used in numerous orthopedic, orthodontic and intravascular medical devices. Its use in commercially distributed, permanently implanted thedical devices has been established in the United States since 1989. with the clearance of the Mitck Bone Anchon (K892126). 3 Nitinol is also used in the Simon Nitinol Filter (K894703), a vena cava filier designed to arrest significant thrombi on their way to the right side of the heart and lungs, while permitting the passage of normal blood. The CRI Cynosar Catheler (K904785/A) is also a Nitinol device cleared by FDA. In addition, Nitinol is used in a wide array of staples, and pins used for bone fixation. Copies of references on Nitinol and its use in biological applications are shown in Exhihit V
#### EQUIVALENCE TO MARKETED PRODUCT
Metagen submits the following information to demonstrate that the Metagen Activelock™ Cerclage System is substantially equivalent in indications and design principles to the following predicate devices, each of which has been determined by FDA to be substantially equivalent to pre-amendment devices or is a preamendments device: Encore Orthopedius CCG Cerclage System, DePuy Control Cable System, Howmedica Dall-Miles Stainless Stecl Cable, and monofilament cerclage wire from manufacturers such as Ethicon and 7.immer.
#### Intended Uses
The indications for use for the Metagen Activelock™ Cerclage System and the predicate devices are substantially the same. All are intended for repair of long bone fractures on reconstruction, reather hent of the greater trochanter in total hip arthroplasty, surface replacement arthroplasty, or other procedures involving trochanteric osteotomy.
#### Design and Materials
The design and functional characteristics of the Metagen Activelock™ Cerclage System and the Encore Orthopedics ( ( ) Cerclage System (K932024) are fundamentally the same. Each is a cerclage band system that includes a slide fastencr at one end that allows the band to encircle the bone and attach to itself in a helt-like manner. Each uses a manual tightening instrument w achieve reduction of a fracture and consolidation of bony fragments. The CCG Cerclaye System uses a band made of commercially purc (CP) titanium, while the band of the Metagen Activelock™ Cerclage System is made from either Ti-6Al-4V alloy or Nitinol.
<sup>4</sup> Fukuyo S, et al., "Shape Mcmory Implants," Engineering Aspects of Shape Memory Alloys, Duerig TW, ct. al. (cds), Boston, Butterworth-Heinemann, p 170. 1990.
{5}------------------------------------------------
#### 510(k) Summary
The Metagen Activelock™ Cerclage System is also comparable to preamendments stainless steel suture wire (16 gage and 18 gage) used for cerclage applications in that fracture reduction and bony fragment consolidation are achieved by applying tension to the device after it has been wrapped or threaded around the bone and fixing it in place while tension is maintained.
Substantial equivalence to cerclare cable systems. including the DePuy Control Cable System (K934557) and Howmedica Dall-Miles Stainless Steel Cable (K844068, K900926, K961569, K934058, K961283) is based on their use for fracture reduction and bony fragment consolidation by applying tension to the device after it has been wrapped or threaded around the bone and fixing it in place while tension is maintained, plus the similarity of clamping methods. The predicate devices use a crimp, swage or sleeve in order to maintain tension on the cable after placement and tensioning.
#### Mechanical Testing
In order to show equivalence of the Metagen Band/NiTi Clamp System to clinically accepted monofilament stainless steel cerclage wire under tensile toad, static tensile tests were performed. Both 16 gage and 18 gage AISI Type 316L stainless steel cerclage wires were tested and were compared with both maximum material condition (MMC) and least material condition (LMC) clamps and bands of the Metagen system. Based on the test results, the data support the hypothesis that the Metagen band/NiTi clamp design provides tensile strength equal to or greater than that of clinically accepted 16 and 18 gage stainless steel cerclage wire.
{6}------------------------------------------------
# SUMMARY: TABLE OF SUBSTANTIAL EQUIVALENCE
The Metagen Activelock™ Cerclage System is substantially equivalent to the Encore Orthopedics CCG Cerclage System, monofilament suture wire, DePuy Control Cable System and Howmedica Dall-Miles Stainless Steel Cable in the following respects:
| | Subject<br>Device | Predicate Devices | | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|-----------------------------------------------------------------|------------------------------------------------------------------------|---------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| | Metagen<br>Activelock™<br>Cerclage<br>System | Encore<br>Orthopedics<br>CCG<br>Cerclage<br>System<br>(K932024) | Mono-<br>filament<br>suture<br>wire<br>(Ethicon,<br>Zimmer,<br>others) | DePuy<br>Control™<br>Cable<br>System<br>(K934557) | Howmedica<br>Dall-Miles<br>Stainless<br>Steel Cable<br>(K844068,<br>K900926,<br>K961569,<br>K934058,<br>K961283) |
| INTENDED USE | | | | | |
| Repair of long bone<br>fractures,<br>reattachment of the<br>greater trochanter, or<br>other orthopedic<br>repairs where<br>cerclage wiring is<br>indicated | YES | YES | YES | YES | YES |
| DESIGN | | | | | |
| Cerclage type | Band | Band | Mono-<br>filament<br>wire | Multi-<br>filament<br>cable | Multi-<br>filament<br>canle |
| Clamping method | Release of<br>superelastic<br>clamp | Slide<br>fastener | Wire<br>twist | Crimp<br>sleeve | Crimp<br>sleeve |
| MATERIALS | | | | | |
| Cerclage element | Ti alloy or<br>Nitinol | CP Ti | Stainless<br>steel | Co-Cr-W-<br>Ni alloy | Stainless<br>steel |
| Clamp | Nitinol | CP Ti | NA | Co-Cr-W-<br>Ni alloy | Stainless<br>steel |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its wing and tail. The eagle is positioned to the right of the text, which is arranged in a circular pattern around the logo. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Floyd G. Larson President Pacific Materials and Interfaces . . . . . . 4329 Graydon Road San Diego, California 92130
AUG | | | |
Re : K972327 Trade Name: Metagen Activelock™ Cerclage System Regulatory Class: II Product Code: JDQ Dated: June 13, 1997 Received: June 23, 1997
Dear Mr. Larson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{8}------------------------------------------------
#### Page 2 - Mr. Floyd G. Larson
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radioloqical Health
Enclosure
{9}------------------------------------------------
#### 08:44 AM FROM Pacific Materials. . 08-05-97 00 301 827 4348 K97232
Indications for Use Statement
Device Name: Metagen Activeluck™ Cerclage System
Indications for Usc:
Repair of long bone fractures due to trauma or reconstruction
Reattachment of the greater trochanter in total hip arthroplarty, surface replacement arthroplasty, or other procedures involving trochanteric osteotomy
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY ) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR Over-The-Counter Use
Division of General Restorative Devices
510(k) Number K972327
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.