Frontier Medical Devices Posterior Cable Screw System
K151346 · Frontier Medical Devices, Inc. · JDQ · Aug 20, 2015 · Orthopedic
Device Facts
Record ID
K151346
Device Name
Frontier Medical Devices Posterior Cable Screw System
Applicant
Frontier Medical Devices, Inc.
Product Code
JDQ · Orthopedic
Decision Date
Aug 20, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3010
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Posterior Cable Screw System is intended to reduce pars defect and to stabilize the spinal operative site during fusion procedures as an adjunct to a rigid posterior fixation system cleared/approved for trauma and spondylolisthesis. The system is indicated for the following: Defect of the pars interarticularis Spondylolisthesis The Posterior Cable Screw System is indicated for pedicle screw attachment for these indications between T1 and the sacrum. Cables may be used for interspinous wiring. The grommet may be used with the cable. The system is not intended to be used in conjunction with a stainless steel implant.
Device Story
Posterior Cable Screw System; spinal stabilization device for pars defects and spondylolisthesis. Components: titanium alloy/cobalt chromium screws, cables, grommets. Surgical procedure: screws inserted into pedicles; cable passed around spinous process; grommet placed at spinous process; cable tensioned; set screws locked. Used in clinical/OR settings by surgeons. Acts as adjunct to rigid posterior fixation. Benefits: provides stabilization during fusion. Manual instruments used for implantation.
Clinical Evidence
Bench testing only. Dynamic mechanical testing provided to demonstrate performance equivalence to predicate systems. No clinical data included.
Technological Characteristics
Materials: medical grade titanium alloy (ASTM F136) and cobalt chromium alloy (ASTM F90). Components: screws, cables, grommets. Sterile. Manual instrumentation. No software or electronic components.
Indications for Use
Indicated for patients with pars interarticularis defects or spondylolisthesis requiring spinal stabilization. Used as an adjunct to rigid posterior fixation systems for fusion procedures between T1 and the sacrum. Contraindicated for use with stainless steel implants.
Regulatory Classification
Identification
A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.
K982987 — SYNTHES POSTERIOR UNIVERSAL SPINAL SYSTEM · Synthes Spine · Oct 21, 1998
K964533 — EBI SPINELINK SYSTEM · Electro-Biology, Inc. · Apr 9, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Frontier Medical Devices, Incorporated Ms. Bethany D. Byman Director of Regulatory Affairs 512 Fourth Street Gwinn, Michigan 49841
August 20, 2015
Re: K151346
Trade/Device Name: Frontier Medical Devices Posterior Cable Screw System Regulation Number: 21 CFR 888.3010 Regulation Name: Bone fixation cerclage Regulatory Class: Class II Product Code: JDQ Dated: May 15, 2015 Received: May 20, 2015
Dear Ms. Byman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Page 2 - Ms. Bethany D. Byman
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K151346
#### Device Name
Frontier Medical Devices Posterior Cable Screw System
Indications for Use (Describe)
The Posterior Cable Screw System is intended to reduce pars defect and to stabilize the spinal operative site during fusion procedures as an adjunct to a rigid posterior fixation system cleared/approved for trauma and spondylolisthesis. The system is indicated for the following:
· Defect of the pars interarticularis
· Spondylolisthesis
The Posterior Cable Screw System is indicated for pedicle screw attachment for these indications between 71 and the sacrum. Cables may be used for interspinous wiring. The grommet may be used with the cable. The system is not intended to be used in conjunction with a stainless steel implant.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary Pursuant to 21 CFR 807.92
| Sponsor: | Frontier Medical Devices, Inc.<br>512 Fourth Street<br>Gwinn, MI 49841 USA |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact: Bethany Byman<br>Ph: 906-232-1200<br>Fx: 906-232-1222 |
| | Prepared: July 17, 2015 |
| Name: | Posterior Cable Screw System |
| Trade name: | Frontier Medical Devices Posterior Cable Screw System |
| Common name: | Pedicle Screw and Spinal Cable System |
| Classifications: | §888.3010 Cerclage, Fixation, Metallic |
| Product Codes: | JDQ |
| Panel/ Branch: | Orthopaedic and Rehabilitation Devices Panel; Panel Code 87 |
| Primary Predicate: | Pioneer Posterior Cable Screw System (K022908) |
| Reference<br>Predicates: | ParsFix (K003351)<br>Stainless Steel Wire (pre-1976) |
| Description: | The Posterior Cable Screw System consists of screws, cables, and<br>grommets utilized for the correction of pars defects. A construct consists<br>of two cable screws with locking set screws and a cable connecting the<br>two screws. The screws are inserted into the pedicles and the cable is<br>passed around the spinous process between the two screws. The grommet<br>surrounds the cable at the location of the spinous process. The cable is<br>tensioned and the set screws are tightened and fully locked per the surgical<br>technique. The implants are manufactured from medical grade titanium<br>alloy per ASTM F136 and cobalt chromium alloy per ASTM F90. The<br>implants are provided sterile. The Posterior Cable Screw System utilizes<br>Class I manual instruments to facilitate implantation of the device<br>components. |
{4}------------------------------------------------
| Intended Use: | The Posterior Cable Screw System is intended to reduce pars defect and<br>to stabilize the spinal operative site during fusion procedures as an adjunct<br>to a rigid posterior fixation system cleared/approved for trauma and<br>spondylolisthesis. The system is indicated for the following:<br>Defect of the pars interarticularis Spondylolisthesis The Posterior Cable Screw System is indicated for pedicle screw<br>attachment for these indications between T1 and the sacrum. Cables may<br>be used for interspinous wiring. The grommet may be used with the cable.<br>The system is not intended to be used in conjunction with a stainless steel<br>implant. |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pre-Clinical<br>Performance Data: | Dynamic tests were provided to support that the Posterior Cable Screw<br>System performs in a manner substantially equivalent to that of predicate<br>systems. No new issues of safety or effectiveness were raised. |
| Technological<br>Characteristics: | The Posterior Cable Screw System and predicate devices are similar in<br>terms of indications for use, material composition, technological<br>characteristics, design characteristics, and mechanical strength. The<br>minor differences in technological characteristics that do exist do not raise<br>any new types of safety or efficacy issues. |
| Substantial<br>Equivalence: | This submission supports the position that the subject Posterior Cable<br>Screw System is substantially equivalent to previously cleared systems.<br>There are no significant differences between the subject system and the<br>predicates which would adversely affect the use of the product. Any<br>differences were not considered significant based on mechanical bench<br>testing. |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.