SYNTHES POSTERIOR UNIVERSAL SPINAL SYSTEM
Device Facts
| Record ID | K982987 |
|---|---|
| Device Name | SYNTHES POSTERIOR UNIVERSAL SPINAL SYSTEM |
| Applicant | Synthes Spine |
| Product Code | MNI · Orthopedic |
| Decision Date | Oct 21, 1998 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3070 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
When used as a posterior pedicle screw fixation system, the Synthes Posterior Universal Spinal System when used in a posicior pealers seler indication and stabilitzation of spinal segments in skeletally mature pations as an adjunct to firstin in the treatment of the following acute and chronic instabilities or deformities of the thoracic, luuluar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, distocation, scoliosis, spinal tumor, and fuiled previous firsion (pseudarthrosis). In addition, the Synthes Posterior Universal Spinal System is intended for treatment of severe spondylolisthesis (Grade 3 and 4) of the L, - S, vertebra in skeletally mature patients receiving fusion by sponeyronsates (Orders ) implants attached to the lumbar and sucral spine with removal of the implans after the attainment of a solid fusion. The levels of pedicle server fixation for these patients are L3-52. When weed as a posterior non-pedicle screw fixation system, Syutles Posterior Universal Spinal System devices are intended for scolivic, lordotic, or kyphotic deformities (such as scolliosis, Schuermann's discesse), degenerative disedisesse (back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), and fractures of the posterior thoracolumbar spine. In addition, when used with the 3.5mm / 6.0mm parallel connector, the Synthes Posterior Universal Spinal System can be linked to the Postcrior Cervical/Thoracic Hook/Rod System.
Device Story
Posterior spinal fixation system; components include stainless steel or titanium rods, hooks, side-opening screws, variable axis screws, connectors, and clamps. Used by surgeons in clinical settings to provide immobilization and stabilization of thoracic, lumbar, and sacral spinal segments as an adjunct to fusion. System components are manually assembled and implanted to correct deformities or stabilize instabilities. Provides mechanical support to the spine to facilitate bone graft fusion; implants may be removed after solid fusion is achieved in severe spondylolisthesis cases. Can be linked to other cervical/thoracic systems via parallel connectors.
Clinical Evidence
Bench testing only.
Technological Characteristics
Materials: stainless steel or titanium. Components: rods, hooks, side-opening screws, variable axis screws, connectors, clamps, and manual surgical instruments. Form factor: posterior spinal fixation system. Connectivity: none. Sterilization: not specified.
Indications for Use
Indicated for skeletally mature patients requiring spinal immobilization and stabilization as an adjunct to fusion. Conditions include degenerative spondylolisthesis with neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, pseudarthrosis, and severe spondylolisthesis (Grade 3-4, L3-S2). Also indicated for non-pedicle screw fixation in scoliotic, lordotic, or kyphotic deformities, discogenic back pain, and thoracolumbar fractures.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Reference Devices
- Posterior Cervical/Thoracic Hook/Rod System
Related Devices
- K130279 — REFORM PEDICLE SCREW SYSTEM · Spinal USA · Jun 20, 2013
- K022949 — SYNTHES USS, CLICK'X, USS VAS, DUAL-OPENING USS AND SMALL STATURE USS · Synthes (Usa) · Mar 24, 2003
- K151695 — Romeo posterior osteosynthesis system · Spineart · Sep 10, 2015
- K964254 — TITANIUM LIBERTY POSTERIOR SPINAL SYSTEM · Sofamor Danek USA,Inc. · Apr 21, 1997