K963313 · Exactech, Inc. · JDI · Nov 14, 1996 · Orthopedic
Device Facts
Record ID
K963313
Device Name
EXACTECH ALL POLY ACETABULAR CUP
Applicant
Exactech, Inc.
Product Code
JDI · Orthopedic
Decision Date
Nov 14, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3350
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Exactech All Polyethylene Acetabular Cup is indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, osteonecrosis, congenital hip dysplasia, rheumatoid arthritis, ankylosing spondylitis and/or posttraumatic degenerative problems. It is also potentially indicated for revision of failed previous reconstructions where sufficient bone stock is present.
Device Story
Acetabular cup prosthesis for total hip arthroplasty; cemented use. Device consists of UHMWPE cup with Cobalt Chrome radiographic marker wire. Available in various diameters (42-70mm OD; 22-32mm ID) to accommodate femoral heads. Implanted by orthopedic surgeons in clinical/OR settings. Device provides articulating surface for femoral components; restores hip joint function; reduces pain/improves mobility in patients with degenerative hip conditions. Design based on established Charnley-style geometry; intended for use with Exactech femoral components.
Clinical Evidence
No clinical data provided for this specific device. Biocompatibility and safety are supported by historical clinical literature (Charnley et al., 1973, 1975; Mirra et al., 1976) and laboratory toxicity studies regarding the use of UHMWPE in orthopedic applications.
Technological Characteristics
Materials: Medical Grade UHMWPE (ASTM F648-84) and Cobalt Chrome radiographic marker wire (ASTM F90-87). Design: Cemented acetabular cup with outer interlock surface. Sterilization: Gamma irradiation (SAL 10^-6) per AAMI guidelines. Connectivity: None (mechanical device).
Indications for Use
Indicated for skeletally mature patients undergoing primary total hip replacement for osteoarthritis, osteonecrosis, congenital hip dysplasia, rheumatoid arthritis, ankylosing spondylitis, or posttraumatic degeneration; also for revision of failed reconstructions with sufficient bone stock. Contraindicated in patients with active infection, insufficient bone stock, neuromuscular disorders preventing hip control, or patients whose weight, age, or activity level suggests early system failure.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
T.A.R.A. by DePuy
Dual Lock by DePuy
New England Baptist System by DePuy
Triad System by Johnson and Johnson
Modular Hip System by Richard
Spectron Total Hip System by Richards
Exactech® All Poly Cup
Related Devices
K963671 — OSTEONICS FLANGED POLYETHYLENE ACETABULAR CUP · Osteonics Corp. · Feb 3, 1997
K992153 — PE-PLUS ACETABULAR CUP · Plus Orthopedics · Sep 23, 1999
K964003 — METAGEN ALL POLY ACETABULAR CUP SYSTEM · Metagen, LLC · Mar 21, 1997
K955137 — IM C-MENT ACETABULAR CAP · Implant Ideas, LLC · Oct 10, 1996
K111848 — EXETER X3 RIMFIT ACETABULAR CUP · Stryker Corp. · Oct 5, 2011
Submission Summary (Full Text)
{0}
EXACTECH®
005
NOV 14 1996
4613 N.W. 6th STREET
GAINESVILLE, FLORIDA 32609
PHONE (352) 377-1140
(800) 392-2832
FAX (352) 378-2617
K963313
510(k) Summary
**Sponsor:**
Exactech, Inc.
4613 N.W. 6th Street
Gainesville, FL 32609
(352) 377-1140 Phone
(352) 378-2617 Fax
**Contact:**
Martha C. Miller
V.P. Regulatory Affairs
August 20, 1996
{1}
006
# 510(K) Summary
# SUMMARY OF THE SAFETY AND EFFECTIVENESS INFORMATION IN THE PREMARKET NOTIFICATION FOR THE EXACTECH ALL POLY ACETABULAR CUP
Exactech, Inc.
Establishment Registration Number 1038671
The Exactech All Poly Acetabular Cup is made of similar materials and is of a similar design to prostheses that were on the market before May 28, 1976. Additionally, the All Poly Acetabular Cup is of similar design to other components on the market that have been determined to be equivalent to devices on the market prior to May 28, 1976. These predicates include, but are not limited to:
- T.A.R.A. by DePuy
- Dual Lock by DePuy
- New England Baptist System by DePuy
- Triad System by Johnson and Johnson
- Modular Hip System by Richard
- Spectron Total Hip System by Richards
- Exactech® All Poly Cup
In addition, Exactech provided to the FDA, design drawings, and material specifications characterizing the All Poly Acetabular Cup.
The Food and Drug Administration, in rules listed in the Federal Register, Friday, September 4, 1987, as Hip Joint Metal/Polymer Semi-Constrained Prosthesis, Section 888.3350 and Docket No. 78N3075, as a class II Device.
## Design Considerations
The All Poly Cup has a geometry typical of currently used prostheses that incorporate an outer interlock surface intended for cemented use. The poly surface finish at the head/cup articulation is equivalent to all approved components currently sold by Exactech. It is manufactured to the same standards as detailed in Exactech's components drawings and specifications for all polyethylene acetabular components. In addition, its overall design is similar to those used by Charnley and others since the early introduction of Total Hip Arthroplasty and assures adequate component thickness as put forth by Bartel for such components.
{2}
007
# Design Parameters
The Exactech All Poly Acetabular Cup consists of various sizes and is made from Medical Grade Ultra High Molecular Weight polyethylene (UHMWPE) with a Cobalt Chrome F 90-87 radiographic marker wire. The component is produced in various inside and outside diameters and accommodates numerous femoral head sizes. The device is designed for use with all Exactech femoral components and femoral heads. A complete trial set and instrumentation is available to assist in accurate implantation of the prosthetic components. Design drawings are typical for such components that have been used in the industry since their introduction by Charnley in the late '60's.
Under this premarket notification, the device is available in outside diameters ranging from 42mm through 70mm with internal diameters accommodating femoral head sizes in 22mm, 26mm, 28mm and 32mm.
# Material Specifications
The Exactech All Poly Acetabular Cup is manufactured from Medical Grade Ultra High Molecular Weight Polyethylene corresponding with ASTM F648-84. It also has a Radiographic marker wire made from Cobalt Chrome corresponding to ASTM F90-87.
# Biocompatibility
Ultra High Molecular Weight Polyethylene has a long history of use in orthopaedic applications. Its biological response has been well characterized by a history of clinical studies (Charnley, J., Cupiz, A., "The Nine and Ten Year Results of the Low Friction Arthroplasty of the Hip", Clinical Orthopaedics, Vol 95, No. 9, 1973.; Halley, D., Charnley, J. "Results of Low Friction Arthroplasty in Patients 30 Years of Age and Younger", Clinical Orthopaedics, No. 112, October, 1975 and Mirra, J., Amstutz, H., Matos, M., and Gold, R., "The Pathology of Joint Tissues and Its Clinical Relevance in Prosthesis Failure". Clinical Orthopaedics, No. 117, June, 1976) and by laboratory studies (Turner, J., Lawrence, W., and Autian, J., "Subacute Toxicity Testing of Biomaterials Using Histopathologic Evaluation of Rabbit Muscle Tissue," Journal of Biomedical Materials Research, Vol 7, 1973. Compatibility of Biocompatibility of Materials for Total Joint Replacement". Journal of Biomed. Mater. Research, Vol 10, No.2, 1976.). These tests include data on human and animal performance and show that the tissue exhibits excellent biocompatibility.
# Sterilization
The Exactech All Polyethylene Cups will be sterilized by gamma irradiation. The Sterility Assurance Level (SAL) is $10^{-6}$. Exactech utilizes Method 3, Protocol B from the "AAMI Guideline for gamma radiation sterilization" for the sterility dose setting and validation procedure.
{3}
008
## Utilization and Implantation
Selection of the Exactech All Polyethylene Acetabular Cup depends on the judgement of the surgeon in relationship to the requirements of the patient. The surgeon should become thoroughly familiar with the technique of implantation by appropriate reading of the literature, and training in the operative skills and techniques required for total hip arthroplasty surgery.
## Indications
The Exactech All Polyethylene Acetabular Cup is indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, osteonecrosis, congenital hip dysplasia, rheumatoid arthritis, ankylosing spondylitis and/or posttraumatic degenerative problems. It is also potentially indicated for revision of failed previous reconstructions where sufficient bone stock is present.
## Contraindications
Use of the Exactech® All Poly Acetabular Cup is contraindicated in patients with active infection, patients without sufficient bone stock to allow appropriate insertion and fixation of the prosthesis, in neuromuscular disorders that do not allow control of the hip joint, and in patients whose weight, age, or activity level would cause the surgeon to expect early failure to the system.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.