K962706 · Biodynamic Technologies, Inc. · HWC · Nov 13, 1996 · Orthopedic
Device Facts
Record ID
K962706
Device Name
EZ-FIX CANNULATED SCREW SYSTEM
Applicant
Biodynamic Technologies, Inc.
Product Code
HWC · Orthopedic
Decision Date
Nov 13, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The EZ-Fix™ Cannulated Screws are indicated for single use to stabilize intracapsular, certain cervical and subcapital fractures of the femur; fractures of the distal femur and proximal tibia; pelvis and acetabular fractures and certain fractures of the shoulder and elbow. They are contraindicated for attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, lumbar spine.
Device Story
EZ-Fix™ Cannulated Screw System consists of titanium alloy (Ti-6Al-4V ELI) bone fixation screws; used by orthopedic surgeons for fracture stabilization in femur, tibia, pelvis, acetabulum, shoulder, and elbow. Screws feature 7.3 mm thread diameter, 20 mm thread length, and 3.45 mm cannulation for 3.2 mm guide pin placement. Design includes self-reaming, self-tapping, and reverse-cutting flutes. Supplied non-sterile; requires steam autoclave sterilization by user. Device provides mechanical fixation to facilitate bone healing; clinical benefit derived from stabilization of fractures.
Clinical Evidence
No clinical data provided. Equivalence established via design comparison and mechanical testing performed by Orthopaedics Biomechanics Laboratory, with results reported as identical to referenced predicate devices.
Technological Characteristics
Material: Titanium alloy (Ti-6Al-4V ELI). Dimensions: 7.3 mm thread diameter, 20 mm thread length, 3.45 mm cannulation. Features: Self-reaming, self-tapping, reverse-cutting flutes. Standards: Tested per ASTM F117. Sterilization: Non-sterile, steam autoclavable. Energy source: None (mechanical).
Indications for Use
Indicated for single-use stabilization of intracapsular, cervical, and subcapital femur fractures; distal femur and proximal tibia fractures; pelvis and acetabular fractures; and specific shoulder and elbow fractures. Contraindicated for fixation to posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Alphatec Large Cannulated Screw System
Richards Universal and 8.0 mm Cannulated Screw
Ace Screw
Howmedica Asnis 2 Guided Screw System
Related Devices
K021233 — AAP CANNULATED SCREWS · Aap Implantate AG · Dec 9, 2002
K963937 — NORTHWEST PRECISION TECHNOLOGIES 6.5 MM CANNULATED CANNULATED CANCELLOUS SCREW · Northwest Precision Technologies · Jan 22, 1997
K102903 — PIONEER CANNULATED SCREW SYSTEM · Pioneer Surgical Technology, Inc. · Oct 20, 2010
K123890 — ACUMED CANNULATED SCREW SYSTEM · Acumed, LLC · Mar 18, 2013
K102343 — EMERGE MEDICAL CANNULATED SCREW FIXATION SYSTEM · Emerge Medical · Nov 23, 2010
Submission Summary (Full Text)
{0}
K962706
NOV 13 1996
# Section 510(k) Premarket Notification
Summary of Safety and Effectiveness Information
Regulatory Authority: Safe Medical Devices Act of 1990, 21 CFR 807.92
Device Trade Name: EZ-Fix™ Cannulated Screw System
Common Name: Large Cannulated Screws
Registration Number: 888.3040
Classification Name: Screw, Fixation, Bone
Establishment Name & Registration Number:
Name: Biodynamic Technologies, Inc.
East Newport Center Drive
Deerfield Beach, Florida 33442
(305) 421-3166 (305) 570-6368 FAX
Number: 1035157
Classification:
Device Class: Class II
Classification Panel: Orthopedic
Contact Person:
Danny Hodgeman
Biodynamic Technologies, Inc.
East Newport Center Drive
Deerfield Beach, Florida 33442
(305) 421-3166 (305) 570-6368 FAX
Special Controls:
Not applicable to this device.
Device Description:
The EZ-Fix™ Cannulated Screws are indicated for single use to stabilize intracapsular, certain cervical and subcapital fractures of the femur; fractures of the distal femur and
Page 14 of 17
{1}
10/29/1996 11:03 35457
PAGE 03
proximal tibia; pelvis and acetabular fractures and certain fractures of the shoulder and elbow. They are contraindicated for attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, lumbar spine.
The EZ-Fix™ Cannulated Screws are fabricated from titanium alloy (Ti-6Al-4V ELI). They are tested per ASTM standards and have a thread diameter and length of 7.3 mm and 20 mm respectively and a shaft diameter of 5.3 mm. They accept a 3.2 mm guide pin through the 3.45 mm cannulation. They are self reaming and tapping and have reverse cutting flutes.
The EZ-Fix™ Cannulated Screws are available in 35 mm to 125 mm lengths are supplied non-sterile.
## Substantially Equivalent Devices:
1. Alphatec Large Cannulated Screw System
See Appendix III for promotional materials for the comparison device.
2. Richards Universal and 8.0 mm Cannulated Screw
See Appendix III for promotional materials for the comparison device.
3. Ace Screw
See Appendix III for promotional materials for the comparison device.
4. Howmedica Asnis 2 Guided Screw System
See Appendix III for promotional materials for the comparison device.
## Comparison to Predicate Device:
The EZ-Fix™ Cannulated Screws are substantially equivalent to the Alphatec, Ace, Richards and Howmedica large cannulated screws in that they are bone fixation screw devices indicated for single use to stabilize intracapsular, certain cervical and subcapital fractures of the femur; fractures of the distal femur and proximal tibia; pelvis and acetabular fractures and certain fractures of the shoulder and elbow. They are contraindicated for attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine. They are fabricated and tested per ASTM standards, they accept a 3.2 mm guide pin, they are self tapping, and they are supplied non-sterile. Like Alphatec, Richards and Howmedica, they have reverse cutting flutes.
The EZ-Fix™ Cannulated Screws are substantially equivalent to the Alphatec and Ace screws in that they are made from titanium alloy (Ti-6Al-4V ELI).
The EZ-Fix™ Cannulated Screws are substantially equivalent to the Alphatec, Ace and Howmedica screws in that they are self reaming and have thread lengths of 20 mm.
The EZ-Fix™ Cannulated Screws are substantially equivalent to the Alphatec screws in they are made available in lengths from 35 mm to 125 mm.
The EZ-Fix™ Cannulated Screws are substantially equivalent to the Howmedica and Richards screws in that they have a thread diameter of 7.3 mm (Howmedica and Richards screws have thread diameters of 7.0 mm and 8.0 mm respectively).
Page 15 of 17
{2}
Page 16 of 17
# Packaging:
The EZ-Fix™ Cannulated Screws and all instruments are supplied non-sterile.
# Sterilization / Re-sterilization:
The EZ-Fix™ Cannulated Screws and all instruments are supplied non-sterile.
Steam autoclavable sterilization trays will be designed to contain the EZ-Fix™ Cannulated Screw System and maintain adequate separation of the implants and instruments.
Sterilization cycles should be followed appropriately to achieve a 10⁻⁶ sterility assurance level (SAL).
See Appendix I for Sterilization Procedure
The EZ-Fix™ is non-pyrogenic. Pyrogenicity testing of the EZ-Fix™ to determine level of endotoxin performed using LAL (Limulus Amebocyte Lysate) method.
# Testing:
Awaiting from Orthopaedics Biomechanics Laboratory.
# Equivalence :
These test values are identical to those obtained on the referenced equivalent.
# Conclusion:
Based on the materials, intended uses, design, testing, and manufacturing, the EZ Fix™ Cannulated Screw System is equivalent to the referenced legally marketed comparison devices. The feature comparison chart below graphically demonstrates equivalence.
# Comparison Table:
| EZ-Fix™
Cann. Screw | Alphatec Lg.
Cann. Screw | Ace®
Screw | Richards
Universal and 8 MM | Howmedica
Asnls 2™ | SE |
| --- | --- | --- | --- | --- | --- |
| Materials | | | | | |
| Titanium Alloy | Titanium Alloy | Titanium Alloy | Stainless Steel
Stainless Steel | Stainless Steel | Yes/No |
| MRI Compatible | | | | | |
| Yes | Yes | Yes | No
No | No | Yes/No |
| Cannulated | | | | | |
| Yes | Yes | Yes | Yes
Yes | Yes | Yes |
{3}
Page 17 of 17
| Thread Diameter | | | | | |
| --- | --- | --- | --- | --- | --- |
| 7.3 MM | 6.5 MM | 6.5 MM | 6.5 MM
8.0 MM | 7.0 MM | Yes/No |
| Guide Pin Diameter | | | | | |
| 3.2 MM | 3.2 MM | 3.2 MM | 2.4 MM
3.2 MM | 3.2 MM | Yes/No |
| Screw Lengths | | | | | |
| 35 - 125 MM | 35 - 125 MM | 25-120 MM | 25 - 120 MM
50 - 125 MM | 35 -130 MM | Yes/No |
| Thread Lengths | | | | | |
| 20 MM | 20 MM | 20 MM | 16 MM
16 MM | 20 MM | Yes/No |
| Intended Use | | | | | |
| Single Use | Single Use | Single Use | Single Use
Single Use | Single Use | Yes |
| Indications | | | | | |
| Intracapsular,
certain cervical
and subcapital
fractures of the
femur. Distal femur
and proximal tibia
fractures. Pelvis
and acetabular
fractures. Certain
fractures of the
shoulder and elbow. | Same | Same | Same
Same | Same | Yes |
| Contraindications | | | | | |
| Attachment or
fixation to the
posterior elements
(pedicles) of the
cervical, thoracic,
or lumbar spine | Same | Same | Same
Same | Same | Yes |
| Performance Standards | | | | | |
| ASTM F117 | Same | Same | Same
Same | Same | Yes |
| Self Reaming | | | | | |
| Yes | Yes | Yes | No
No | Yes | Yes/No |
| Self Tapping | | | | | |
| Yes | Yes | Yes | Yes
Yes | Yes | Yes |
| Reverse Cutting | | | | | |
| Yes | Yes | No | Yes
Yes | Yes | Yes |
| Sterile | | | | | |
| Non-Sterile | Non-Sterile | Non-Sterile | Non-Sterile
Non-Sterile | Non-Sterile | Yes |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.