The Acumed Cannulated Screw System consists of screws, washers, and accessories and is generally intended for fixation of fracture, fusion and osteotomies of large and small bones appropriate for size of device, which may include the following: Minimally invasive reconstruction of fractures and joints; Adjuvant for osteosynthesis in complex joint fractures; Multifragment joint fractures: Simple metaphyseal fractures of the wrist, ankle, elbow, and shoulder; Condylar fractures; Epiphyseal and metaphyseal fractures in children; Osteochrondritis dissecans; Ostero-Chondral Fractures; Ligament avulsion injuries; Ligament fixation; Other small fragment, cancellous bone fractures; Small joint fusion; Areas where accurate screw placement is vital; Metatarsal and phalangeal osteotomies; Fractures of the tarsals, metatarsals and other fractures of the foot; Avulsion fractures and fractures of metatarsal V; Tarso-metatarsal and metatarso-phalangeal arthrodesis; Tarsal Fusions; Calcaneal and talar fractures; Subtalar arthrodesis; Ankle arthrodesis; Fractures of small joints, such as: Ankle fractures, Navicular fractures of the fibula, malleolus, and calcaneus; Distal tibia and pilon fractures; Acetabular fractures; Other fractures of the pelvic ring; Sacroiliac joint disruptions; Fractures of the femoral head and neck; Supracondylar femoral fractures; Slipped capital femoral epiphyses: An adjunct to DHS in basilar neck fractures; Pediatric femoral neck fractures; Intercondylar femur fractures; Intracapsular fractures of the hip; Fractures of the distal femur and proximal tibia; Patellar fractures; Tibial plateau fractures; Small fragments of the hand and wrist; Fractures of the carpals and metacarpals; Carpal and metacarpal arthrodesis; Scaphoid fracture and other fractures of the hand; Phalangeal and interphalangeal fractures; Fractures of the ulna and radius; Radial head fractures; Fractures of the olecranon and distal humerus; Humeral head fractures; Ligament fixation at the proximal humerus; and Glenoid fractures. Washers may be used with the screws in certain applications.
Device Story
Acumed Cannulated Screw System comprises titanium alloy screws, washers, and accessories for bone fixation. Used for fracture, fusion, and osteotomy procedures across large and small bones; facilitates minimally invasive reconstruction. Screws range 3.0mm–7.3mm diameter and 8mm–150mm length. Provided sterile or non-sterile. Operated by surgeons in clinical/OR settings to provide mechanical stabilization of bone fragments. Benefits include accurate placement via cannulation and structural support for healing.
Clinical Evidence
Bench testing only. Testing performed in accordance with ASTM F543 standards.
Technological Characteristics
Materials: Titanium alloy per ASTM F136. Design: Cannulated screw system; diameters 3.0mm–7.3mm; lengths 8mm–150mm. Energy source: None (mechanical). Connectivity: None. Sterilization: Provided sterile and non-sterile.
Indications for Use
Indicated for fixation of fractures, fusions, and osteotomies of large and small bones in pediatric and adult populations. Includes complex joint, metaphyseal, condylar, epiphyseal, and cancellous bone fractures; ligament fixation; arthrodesis; and osteochondritis dissecans.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
K123890 (pg 1/3)
Acumed Cannulated Screw System 510(k) Notification
## 510(k) Summary
# MAR 1 8 2013
# Contact Details
______________________________________________________________________________________________________________________________________________________________________________
Applicant Name:
Acumed LLC 5885 NW Cornelius Pass Road, Hillsboro, OR 97124-9432
Kara Budor, Regulatory Specialist 503-207-1412
Date Prepared: December 17, 2012
#### Device Name
Trade Name: Acumed Cannulated Screw System
Common Name: Cannulated Screw
Classification: 21 CFR 888.3040 Smooth or threaded metallic bone fixation fastener.
Class: II
Product Code: HWC, OUR
## Legally Marketed Predicate Device(s)
There are six predicate devices. The comparison is to the Synthes 6.5 mm Cannulated Screw, the Synthes 7.0/7.3 mm Cannulated Screws, the aap Cannulated Screw, the Osteonics Osteo 4.0mm Cannulated Screw System, the Instratek Mini Cannulated Titanium Headed and Headless Screw Set, and the Pioneer Cannulated Screw System.
| 510(k) Number | Product Code | Trade Name | Applicant |
|---------------|----------------------|-----------------------------------------------------------|--------------------------------|
| K021932 | HWC, OUR<br>888.3040 | Synthes 6.5 mm Cannulated Screw | Synthes (USA) |
| K962011 | HWC<br>888.3040 | Synthes 7.0/7.3 mm Cannulated<br>Screws | Synthes (USA) |
| K111316 | HWC<br>888.3040 | aap Cannulated Screw | aap Implantate AG |
| K983165 | HWC<br>888.3040 | Osteo 4.0mm Cannulated Screw<br>System | Osteonics Corporation |
| K120493 | HWC<br>888.3040 | Mini Cannulated Titanium Headed<br>and Headless Screw Set | Instratek, Inc. |
| K102903 | HWC, OUR<br>888.3040 | Pioneer Cannulated Screw System | Pioneer Surgical<br>Technology |
{1}------------------------------------------------
#### Device Description
The Acumed Cannulated Screw System consists of screws, washers, and accessories used for fixation of fracture, fusion and osteotomies of large and small bones appropriate for the size of the device.
Screws are available in a variety of diameters and lengths to accommodate various indications and patient anatomy. All screws and washers are made of titanium alloy per ASTM F136. Screws range in diameter from 3.0mm to 7.3mm and in lengths from 8mm to 150mm.
All implants are provided sterile and non-sterile.
#### Intended Use/Indications for use
The Acumed Cannulated Screw System consists of screws, washers, and accessories and is generally intended for fixation of fracture, fusion and osteotomies of large and small bones appropriate for size of device, which may include the following: Minimally invasive reconstruction of fractures and joints; Adjuvant for osteosynthesis in complex joint fractures; Multifragment joint fractures: Simple metaphyseal fractures of the wrist, ankle, elbow, and shoulder; Condylar fractures; Epiphyseal and metaphyseal fractures in children; Osteochrondritis dissecans; Ostero-Chondral Fractures; Ligament avulsion injuries; Ligament fixation; Other small fragment, cancellous bone fractures; Small joint fusion; Areas where accurate screw placement is vital; Metatarsal and phalangeal osteotomies; Fractures of the tarsals, metatarsals and other fractures of the foot; Avulsion fractures and fractures of metatarsal V; Tarso-metatarsal and metatarso-phalangeal arthrodesis; Tarsal Fusions; Calcaneal and talar fractures; Subtalar arthrodesis; Ankle arthrodesis; Fractures of small joints, such as: Ankle fractures, Navicular fractures of the fibula, malleolus, and calcaneus; Distal tibia and pilon fractures; Acetabular fractures; Other fractures of the pelvic ring; Sacroiliac joint disruptions; Fractures of the femoral head and neck; Supracondylar femoral fractures; Slipped capital femoral epiphyses: An adjunct to DHS in basilar neck fractures; Pediatric femoral neck fractures; Intercondylar femur fractures; Intracapsular fractures of the hip; Fractures of the distal femur and proximal tibia; Patellar fractures; Tibial plateau fractures; Small fragments of the hand and wrist; Fractures of the carpals and metacarpals; Carpal and metacarpal arthrodesis; Scaphoid fracture and other fractures of the hand; Phalangeal and interphalangeal fractures; Fractures of the ulna and radius; Radial head fractures; Fractures of the olecranon and distal humerus; Humeral head fractures; Ligament fixation at the proximal humerus; and Glenoid fractures. Washers may be used with the screws in certain applications.
#### Substantial Equivalence Comparison
The basic comparison between the Acumed Cannulated Screw System and the predicate devices is given in the table below.
{2}------------------------------------------------
#### Acumed Cannulated Screw System 510(k) Notification
#### K123890 (pg 3/3)
| Predicate | Material | Diameter | Length | Cannulated | Provided<br>sterile /<br>non-sterile |
|------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|-------------------|------------------|------------|--------------------------------------|
| Acumed<br>Cannulated Screw<br>System | Titanium alloy<br>per ASTM F136 | 3.0mm to<br>7.3mm | 8mm to<br>150mm | Yes | Sterile and<br>non-sterile |
| Synthes 6.5mm<br>Cannulated Screw | SS and Ti alloy | 6.5mm | 20mm to<br>200mm | Yes | Unknown |
| Synthes 7.0/7.3mm<br>Cannulated Screws | SS and Ti alloy | 7.0mm to<br>7.3mm | Unknown | Yes | Sterile |
| aap Cannulated<br>Screw | Ti alloy per<br>ASTM F136 or.<br>ISO 5832-2<br>SS per ASTM<br>F138 or ISO<br>5832-1 | 2.7mm to<br>7.5mm | Unknown | Yes | Unknown |
| Osteonics<br>Corporation<br>Osteo 4.0mm<br>Cannulated Screw<br>System | Titanium 6A1-4V<br>ELI alloy | 4.0mm | Unknown | Yes | Sterile and<br>non-sterile |
| Instratek Mini<br>Cannulated<br>Titanium Headed<br>and Headless<br>Screw Set | Titanium alloy<br>(Ti6AL4V) | 2.5mm to<br>4.0mm | Unknown | Yes | Unknown |
| Pioneer Cannulated<br>Screw System | Biodur 108 per<br>ASTM F2229 | 3.5mm to<br>7.5mm | 8mm to<br>200mm | Yes | Sterile and<br>non-sterile |
The Acumed Cannulated Screw System, the Synthes 6.5 mm Cannulated Screw, the Synthes 7.0/7.3 mm Cannulated Screws, the aap Cannulated Screw, the Osteonics Osteo 4.0mm Cannulated Screw System, the Instratek Mini Cannulated Titanium Headed and Headless Screw Set, and the Pioneer Cannulated Screw System are all sets of cannulated screws which are used to achieve fixation. There are some differences, but none of them raise new issues of safety or effectiveness. The Acumed Cannulated Screw System is substantially equivalent to the predicate devices.
### Non-clinical Testing
The non-clinical testing included in this submission includes testing to ASTM F543.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold font. The text is centered in the image. The text is likely the name of a government agency.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 18, 2013
Acumed LLC % Ms. Kara Budor Regulatory Specialist 5885 Northwest Cornelius Pass Road Hillsboro, Oregon 97124
Re: K123890
Trade/Device Name: Acumed Cannulated Screw System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC, OUR Dated: December 17, 2012 Received: December 18, 2012
Dear Ms. Budor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
#### Page 2 – Ms. Kara Budor
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Image /page/4/Picture/7 description: The image shows the name "Erin I.D. Keith" in a stylized font. The first name "Erin" is in a bold, sans-serif font. The middle initial "I.D." is in a more decorative font with concentric circles and lines. The last name "Keith" is in a similar style to the middle initial, with concentric circles and lines.
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Acumed Cannulated Screw System 510(k) Notification
# Indications for Use
510(k) Number (if known); K123890 (pg 1/1)
Device Name: Acumed Cannulated Screw System
Indications for Use:
The Acumed Cannulated Screw System consists of screws, washers, and accessories and is generally intended for fixation of fracture, fusion and osteotomies of large and small bones appropriate for size of device, which may include the following: Minimally invasive reconstruction of fractures and joints: Adjuvant for osteosynthesis in complex joint fractures; Multifragment joint fractures: Simple metaphyseal fractures of the wrist, ankle, elbow, and shoulder; Condylar fractures; Epiphyseal fractures in children; Osteochrondritis dissecans; Ostero-Chondral Fractures; Ligament avulsion injuries; Ligament fixation; Other small fragment, cancellous bone fractures; Small joint fusion; Areas where accurate screw placement is vital; Metatarsal and phalangeal osteotomies; Fractures of the tarsals, metatarsals and other fractures of the foot; Avulsion fractures of metatarsal V: Tarso-metatarsal and metatarso-phalangeal arthrodesis: Tarsal Fusions: Calcaneal and talar fractures; Subtalar arthrodesis; Ankle arthrodesis; Fractures of small joints, such as: Ankle fractures, Navicular fractures of the fibula, malleolus, and calcaneus: Distal tibia and pilon fractures: Acetabular fractures: Other fractures of the pelvic ring; Sacroiliac joint disruptions; Fractures of the femoral head and neck: Supracondylar femoral fractures; Slipped capital femoral epiphyses; An adjunct to DHS in basilar neck fractures; Pediatric femoral neck fractures; Intercondylar femur fractures; Intracapsular fractures of the hip; Fractures of the distal femur and proximal tibia; Patellar fractures; Tibial plateau fractures; Small fragments of the hand and wrist; Fractures of the carpals and metacarpals; Carpal and metacarpal arthrodesis; Scaphoid fracture and other fractures of the hand; Phalangeal and interphalangeal fractures; Fractures of the ulna and radius; Radial head fractures: Fractures of the olecranon and distal humerus: Humeral head fractures: Ligament fixation at the proximal humerus; and Glenoid fractures. Washers may be used with the screws in certain applications.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D).
AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth Frank -S
Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.