K141401 · Feet For Life Podiatry Center · HWC · Feb 20, 2015 · Orthopedic
Device Facts
Record ID
K141401
Device Name
EASY OUT
Applicant
Feet For Life Podiatry Center
Product Code
HWC · Orthopedic
Decision Date
Feb 20, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Easy Out™ system is indicated for temporary internal fixation and stabilization of osteotomies and fractures during normal healing of ankle, elbow and small bones such as those in the foot and wrist.
Device Story
Easy Out™ system comprises partially threaded, self-tapping, self-drilling solid bone screws; used for temporary internal fixation and lag compression of osteotomies and fractures in small bones (ankle, elbow, foot, wrist). Screws range 2.0mm–4.0mm diameter and 12mm–60mm length; fabricated from medical grade titanium alloy. Used by orthopedic surgeons or podiatrists in clinical/surgical settings. Device provides mechanical stabilization to bone fragments to facilitate normal healing. Output is physical fixation of bone segments.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing including static and dynamic 3-point bend tests (ASTM F1264-03), torsional strength tests, and axial pull-out tests (ASTM F543-07).
Indicated for temporary internal fixation and stabilization of osteotomies and fractures in ankle, elbow, and small bones (foot/wrist) during normal healing.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Small Bone Innovations, Inc. SBI K-Wires (K051605)
Zimmer®, Inc. Zimmer® Plates and Screws System (ZPS) (K143066)
K143624 — Headless Compression Screw System · Skeletal Dynamics, LLC · Jan 13, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes and a staff. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 20, 2015
Feet for Life Podiatry % Ms. Carrie Hetrick Emergo Group 816 Congress Avenue, Suite 1400 Austin, Texas 78701
Re: K141401 Trade/Device Name: Easy Out™ Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC Dated: January 20, 2015 Received: January 21, 2015
Dear Ms. Hetrick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Ms. Carrie Hetrick
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K141401
Device Name Easy Out™
Indications for Use (Describe)
The Easy Out™ system is indicated for temporary internal fixation of osteotomies and fractures during normal healing of ankle, elbow and small bones such as those in the foot and wrist.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
for
## Easy Out™
#### 1. Submission Sponsor
Feet For Life Podiatry Centers 8637 Delmar Blvd. Saint Louis, MO 63124 United States of America Phone: (314) 983-0303 Fax: (314) 983-2777 Contact: Dr. Michael Horwitz, CEO
#### 2. Submission Correspondent
Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701 Cell Phone: 720.838.4113 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Carrie Hetrick, Senior Consultant, RA Email: project.management@emergogroup.com
#### 3. Date Prepared
February 19, 2015
### 4. Device Identification
Trade/Proprietary Name: Easy Out™ Common/Usual Name: Temporary Internal Fixation Device Classification Name: Smooth or threaded metallic bone fixation fastener Classification Regulation: 888.3040 Classification Panel: Orthopedic Product Code: HWC Device Class: II
### 5. Legally Marketed Predicate Device(s)
Small Bone Innovations, Inc. SBI K-Wires (510(k) Number K051605, cleared August 10, 2005)
Zimmer®, Inc. Zimmer® Plates and Screws System (ZPS) - Non-Sterile/ Non-Sterile ZPS Screws and Washers (510(k) Number K143066, cleared November 28, 2014)
Stryker® Corporation Asnis™ Micro Cannulated Screw (formally Howmedica Osteonics Corp.) (510(k) Number: K071092) 2.0mm and 3.0mm diameter implants
{4}------------------------------------------------
Stryker® Corporation, Asnis™ III Cannulated Screws (510(k) Number K024060, cleared December 20, 2002)
### 6. Device Description
Easy Out™ system consists of various, partially threaded solid screws designed to create lag compression used for temporary internal fixation small bones and fragments. The Easy Out™ bone screws are self-tapping, self-drilling screws with thread diameters of 2.0 mm, 2.5 mm, 3.0 mm, 3.5mm, 4.0 mm, and lengths that range from 12 mm to 60 mm. All compression screws are fabricated from medical grade titanium alloy (per ASTM F-136). Associated instrumentation is available. All compression screws are offered "non sterile".
#### 7. Indication for Use Statement
The Easy Out™ system is indicated for temporary internal fixation and stabilization of osteotomies and fractures during normal healing of ankle, elbow and small bones such as those in the foot and wrist.
#### 8. Comparison to Predicate Devices
The Feet For Life Podiatry Centers Easy Out™ system is similar in intended use, basic shape, compatible screw diameters, materials and performance characteristics to the predicate devices. The subject devices are provided non-sterile.
#### 9. Non-Clinical Performance Data
Performance testing and engineering analysis were conducted to characterize the performance of the Easy Out™ device. The device functioned as intended and the observed test results demonstrate substantial equivalence to the predicate devices. Static and dynamic 3-point bend tests, static torsional tests and axial pull-out tests were conducted to support substantial equivalence. Static and dynamic Three Point Bend testing was conducted in accordance with ASTM F1264-03 (07), Standard Specification and Test Method for Intramedullary Fixation Devices. Additionally testing was conducted to characterize the torsional strength of the screw as well as axial pullout per ASTM F543-07(e1), Standard Specification and Test Methods for Metallic Medical Bone Screws.
Performance testing and engineering analysis were conducted to characterize the performance of the Feet for Life Easy Out™ device. The results of this testing demonstrate that Easy Out™ performs as intended and demonstrates that the device is substantially equivalence to its predicate devices.
The Easy Out™ device complies with the applicable voluntary standards for biocompatibility and sterilization.
{5}------------------------------------------------
#### 10. Clinical Performance Data
Clinical data was not needed to support the safety and effectiveness of the subject device. Easy Out™ and its predicates are all indicated for temporary fixation. Easy Out™ does not raise any new issues of safety and effectiveness.
#### 11. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
The Feet for Life Easy Out™ device has the same or similar intended use, indications, principles of operation, and technological characteristics as the predicate systems. Easy Out™ device is substantially equivalent to the predicate devices in terms of intended use, design, materials, and function.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.