Hercules Syndesmosis Implant System

K220260 · In2bones USA, LLC · HTN · Mar 31, 2022 · Orthopedic

Device Facts

Record IDK220260
Device NameHercules Syndesmosis Implant System
ApplicantIn2bones USA, LLC
Product CodeHTN · Orthopedic
Decision DateMar 31, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Hercules® Syndesmosis Implant System is intended to provide fixation during the healing process following a syndesmotic trauma such as fixation of syndesmosis disruptions) in connection with Weber B and C ankle fractures and as an adjunct in external and intramedullary fixation systems. The Hercules® Syndesmosis Implant System is also intended to provide fixation during healing process after joint reconstruction in the midfoot and forefoot including correction of hallux valgus deformity.

Device Story

Sterile, single-use bone fixation system; stabilizes syndesmotic trauma and midfoot/forefoot reconstructions. Comprises UHMWPE suture tensioned between two low-profile titanium alloy buttons. Used by orthopedic surgeons in clinical/OR settings. Functions as stand-alone fixation or adjunct to plate/intramedullary systems. Provides mechanical stabilization during bone healing. Surgeon inserts device using provided ancillary instruments to secure bone segments; tensioning suture between buttons maintains reduction. Benefits patient by providing stable fixation during recovery.

Clinical Evidence

Bench testing only. Performed cyclic displacement and load to failure testing to demonstrate mechanical equivalence. Biocompatibility validated per ISO 10993-1. Sterilization validated per ISO 11137-2 (gamma) and ISO 11135 (ETO). Packaging validated per ISO 11607-1. Endotoxin testing performed per ANSI/AAMI ST72.

Technological Characteristics

Materials: UHMWPE suture, titanium alloy buttons. Principle: Suture-button fixation system. Form factor: Low-profile buttons with suture. Sterilization: Gamma (ISO 11137-2) or ETO (ISO 11135). Connectivity: None (mechanical device).

Indications for Use

Indicated for patients requiring fixation during healing of syndesmotic trauma (e.g., syndesmosis disruptions in Weber B/C ankle fractures), as an adjunct to external/intramedullary fixation, or for joint reconstruction in midfoot/forefoot (including hallux valgus correction).

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION". March 31, 2022 In2Bones USA, LLC Christine Scifert VP of Quality Assurance and Regulatory Affairs 6000 Poplar Ave, Suite 115 Memphis, Tennessee 38119 Re: K220260 Trade/Device Name: Hercules® Syndesmosis Implant System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HTN Dated: January 27, 2022 Received: January 31, 2022 Dear Christine Scifert: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Limin Sun, Ph.D. Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K220260 Device Name Hercules® Syndesmosis Implant System Indications for Use (Describe) The Hercules® Syndesmosis Implant System is intended to provide fixation during the healing process following a syndesmotic trauma such as fixation of syndesmosis disruptions) in connection with Weber B and C ankle fractures and as an adjunct in external and intramedullary fixation systems. The Hercules® Syndesmosis Implant System is also intended to provide fixation during healing process after joint reconstruction in the midfoot and forefoot including correction of hallux valgus deformity. Type of Use (Select one or both, as applicable) | <span style="font-family: Arial, sans-serif;"></span> <span style="font-size: 10pt;"> <span style="font-family:Wingdings;">ü</span> Prescription Use (Part 21 CFR 801 Subpart D)</span> | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <span style="font-family: Arial, sans-serif;"></span> <span style="font-size: 10pt;"> <span style="font-family:Wingdings;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) Summary # Hercules® Syndesmosis Implant System March 30, 2022 | Company: | In2Bones USA, LLC<br>6000 Poplar Ave, Suite 115<br>Memphis, TN 38119<br>901-260-7931 | |--------------------|--------------------------------------------------------------------------------------------------| | Company Contact: | Christine Scifert | | Trade Name: | Hercules® Syndesmosis Implant System | | Common Name: | Fixation button and suture | | Classification: | II | | Regulation Number: | 21 CFR 888.3030 – Single/multiple component metallic bone fixation appliances<br>and accessories | | Panel: | 87-Orthopedic | | Product Code(s): | HTN – Washer, bolt nut | The In2Bones Hercules® Syndesmosis Implant System is a sterile, single-use, Device Description: device intended to stabilize syndesmotic trauma of the ankle. The subject system consists of UHMWPE suture tensioned between two low profile titanium alloy buttons, designed to complement plate fixation or to allow use as a stand-alone device. This device is also packaged with various ancillary instruments to aid in insertion. ### Indications for Use: The Hercules® Syndesmosis Implant System is intended to provide fixation during the healing process following a syndesmotic trauma such as fixation of syndesmosis disruptions) in connection with Weber B and C ankle fractures and as an adjunct in external and intramedullary fixation systems. The Hercules® Syndesmosis Implant System is also intended to provide fixation during healing process after joint reconstruction in the midfoot and forefoot including correction of hallux valgus deformity. Substantial Equivalence: The subject components were demonstrated to be substantially equivalent to the following systems previously cleared by the FDA: {4}------------------------------------------------ Primary Predicate: - . K043248- Arthrex Tightrope Additional Predicates: - K170249 Wright Medical Gravity Syndesmosis LP ● - K170440 – Arthrosurface KISSLoc Suture System The subject Hercules® Syndesmosis Implant System is made of UHMWPE sutures and titanium alloy and has been demonstrated to be substantially equivalent to the previously cleared devices identified above as the products are similar in indications, materials and geometry. Reference Devices: - K063778 and K181774 Teleflex Force Fiber ● - K083070 and K130033 BioMet ZipTight ToggleLoc ● ### Performance Testing: Cyclic displacement and load to failure bench testing (mechanical testing) was performed to demonstrate substantial equivalence to the predicate devices. The Hercules® Syndesmosis Implant System was validated per ISO 10993-1 for biocompatibility, ISO 11137-2 for gamma sterilization, ISO 11135 for ETO sterilization and ISO 11607-1 for packaging. Endotoxin testing was completed per ANSI/AAMI ST72. ### Conclusion Based on the test results and the comparison to the predicate devices, the subject device is determined to be substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...