K211612 · Medline Industries, Inc. · HTN · Dec 10, 2021 · Orthopedic
Device Facts
Record ID
K211612
Device Name
Medline UNITE REFLEX Dynamic Discs
Applicant
Medline Industries, Inc.
Product Code
HTN · Orthopedic
Decision Date
Dec 10, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Medline UNITE® REFLEX™ Dynamic Discs are indicated for use in bone reconstruction, osteotomies, arthrodesis, joint fusion, fracture repair, and fracture fixation of bones appropriate for the size of the device. Discs are intended for single use only.
Device Story
Medline UNITE® REFLEX™ Dynamic Discs are orthopedic washers manufactured from nickel titanium alloy (nitinol). Used in conjunction with bone screws to increase the surface area of the screw head during bone reconstruction, osteotomies, arthrodesis, joint fusion, and fracture fixation. Operated by surgeons in clinical settings. Discs utilize chemical etching and passivation to form a protective oxidation layer. Device provides mechanical support to bone fixation constructs; benefits include improved load distribution across the screw head. No software or algorithm involved.
Clinical Evidence
No clinical data. Bench testing only. Testing included transformation temperature (ASTM F2004), galvanic corrosion (ASTM F3044), and corrosion susceptibility (ASTM F2129). Results showed no pitting/corrosion, and material release was below the 35µg/day nickel limit.
Technological Characteristics
Material: Nickel titanium alloy (nitinol). Surface treatment: Chemical etching and passivation. Dimensions: Various diameters (Ø3.0mm to Ø7.0mm). Principle: Mechanical washer to increase screw head surface area. Standards: ASTM F2004 (transformation temperature), ASTM F2129 (corrosion susceptibility), ASTM F3044 (galvanic corrosion). Non-sterile. Single use.
Indications for Use
Indicated for bone reconstruction, osteotomies, arthrodesis, joint fusion, fracture repair, and fracture fixation in patients requiring bone fixation appropriate for the device size. Intended for single use.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 10, 2021
Medline Industries, Inc. Jennifer Mason Senior Regulatory Affairs Specialist Three Lakes Drive Northfield, Illinois 60093
Re: K211612
Trade/Device Name: Medline UNITE® REFLEX™ Dynamic Discs Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HTN Dated: September 1, 2021 Received: September 3, 2021
Dear Jennifer Mason:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K211612
Device Name Medline UNITE® REFLEX™ Dynamic Discs
Indications for Use (Describe)
The Medline UNITE® REFLEX™ Dynamic Discs are indicated for use in bone reconstruction, osteotomies, arthrodesis, joint fusion, fracture repair, and fracture fixation of bones appropriate for the size of the device. Discs are intended for single use only.
Type of Use (Select one or both, as applicable)
| <span style="font-size:100%;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
4-001
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for Medline. The logo features the word "MEDLINE" in white, bold sans-serif font. Above and below the word is a white star-like shape with four points. The background of the logo is a solid dark blue color.
# 510(k) SUMMARY [AS REQUIRED BY 21CFR807.92(c)]
# Submitter / 510(k) Sponsor
Medline Industries, Inc. Three Lakes Drive Northfield, IL 60093
Registration Number: 1417592
# Contact Person
Jennifer Mason Senior Regulatory Affairs Specialist Phone: 847-643-3652 Email: jamason(@medline.com
Summary Preparation Date November 22, 2021
Type of 510(k) Submission
Traditional
# Device Name / Classification
Trade Name: Medline UNITE® REFLEX™ Dynamic Discs Common Name: Washer, Bolt Nut Classification Name: Washer, Bolt Nut Product Code: HTN Classification Panel: Orthopedic Regulatory Class: Class II Regulation Number: 21 CFR 888.3030
## Primary Predicate Device
Medline Cannulated Screws K130319
Reference Device Medline UNITE® REFLEX™ Nitinol Staple System K210482
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is the logo for Medline. The logo is white text on a blue background. The text "MEDLINE" is in bold, sans-serif font. To the right of the text is a white starburst symbol.
## Device Description
The Medline UNITE® REFLEX™ Dynamic Discs are manufactured from nickel titanium alloy (nitinol). The discs utilize chemical etching and passivation to form a protective oxidation layer on the outer surface. The discs are offered in various diameters and thicknesses to be used in conjunction with screws of various diameters.
## Indications for Use
The Medline UNITE® REFLEX™ Dynamic Discs are indicated for use in bone reconstruction, osteotomies, arthrodesis, joint fusion and fracture repair and fracture fixation of bones appropriate for the size of the device. Discs are intended for single use only.
## Summary of Technological Characteristics
| Device<br>Characteristic | Proposed Device | Predicate (Primary)<br>Device | Reference Device | Comparison<br>Analysis |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| Product Name | Medline UNITE®<br>REFLEX™ Dynamic<br>Discs | Medline Cannulated<br>Screws | Medline UNITE®<br>REFLEX™ Nitinol<br>Staple System | Different |
| Photos of the<br>Devices | Image: Proposed Device | Image: Predicate Device | Image: Reference Device | N/A |
| 510(k)<br>Reference | K211612 | K130319 | K210482 | Different |
| Product Owner | Medline Industries, Inc. | Medline Industries, Inc. | Medline Industries, Inc. | Same |
| Product Code | HTN | HWC | JDR | Different |
| Intended Use | The Medline UNITE®<br>REFLEX™ Dynamic<br>Discs are indicated for use<br>in bone reconstruction,<br>osteotomies, arthrodesis,<br>joint fusion and fracture<br>repair and fracture fixation<br>of bones appropriate for<br>the size of the device.<br>Discs are intended for<br>single use only. | The Medline Cannulated<br>Screws are indicated for<br>use in bone reconstruction,<br>osteotomies, arthrodesis,<br>joint fusion, fracture<br>repair, and fracture<br>fixation of bones<br>appropriate for the size of<br>the device. Screws are<br>intended for single use<br>only. | The Medline UNITE<br>REFLEX Nitinol<br>Staples are intended to<br>provide fixation for<br>fractures, fusions or<br>osteotomies of the bones<br>of the hand and foot<br>such as: LisFranc<br>arthrodesis, Akin<br>Osteotomy, Scarf and<br>Chevron osteotomies | Primary Predicate<br>- Same<br>Secondary<br>Predicate –<br>Similar |
## TABLE 1: COMPARISON OF PROPOSED, PREDICATE AND REFERENCE DEVICES
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is the logo for Medline. The logo features the word "MEDLINE" in a bold, sans-serif font, stacked above a white star-like symbol. The background of the logo is a solid dark blue color. The star symbol has four points, with each point resembling a stylized wing or blade.
| | | | Staples are single use<br>only. | |
|-----------------------------|------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Regulation<br>Number | 21 CFR 888.3030 | 21 CFR 888.3030 | 21 CFR 888.3030 | Same |
| Design Features | Increases surface area of<br>screw head | Increases surface area of<br>screw head | N/A | Same |
| Design<br>Configurations | Ø3.0 or 5.75mm<br>Ø3.5 or 6.61mm<br>Ø4.0 or 7.65mm<br>Ø4.5 or 8.50mm<br>Ø5.5 or 10.20mm<br>Ø7.0 or 12.75mm | Ø2.0/Ø2.5 or 6.00mm<br>Ø3.0/Ø3.5 or 8.00mm<br>Ø4.0/Ø4.5 or 10.00mm<br>Ø6.5 or 14.00mm<br>Ø7.5 or 16.00mm | 8 x 8mm<br>10 x 10mm<br>12 x 12mm<br>15 x 15mm<br>15 x 18mm<br>18 x 20mm<br>20 x 20mm<br>25 x 20mm<br>25 x 25mm<br>25 x 27mm | Different |
| Materials | Nickel Titanium Alloy | Titanium Alloy or<br>Stainless Steel | Nickel Titanium Alloy | Primary Predicate<br>– Different<br>Secondary<br>Predicate – Same |
| Prescription vs.<br>OTC | Prescription | Prescription | Prescription | Same |
| Sterile vs. Non-<br>Sterile | Non-Sterile | Non-Sterile | Non-Sterile | Same |
| Single Use vs.<br>Reusable | Single Use Only | Single Use Only | Single Use Only | Same |
The proposed modified device. Medline UNITE® REFLEX™ Dynamic Discs is substantially equivalent to the primary predicate, the Medline Cannulated Screws. A discussion of similarities and differences is listed below.
- Indications for use same. The subject device, the Medline UNITE® REFLEX™ Dynamic Discs . have the exact same indications for use as the Medline Cannulated Screws.
- Design Features same. Both the subject device and the predicate device are intended to increase the . surface area of the screw head.
- Design Configurations – similar. The Medline UNITE® REFLEX™ Dynamic Discs will be offered in sizes ranging from Ø3.0 to Ø7.0 whereas, the predicate device is offered in sizes Ø2.0 to Ø7.5. The subject device falls within the size range of the current Medline Cannulated Screws.
- Materials different. The Medline UNITE® REFLEX™ Dynamic Discs are made from nitinol. The . predicate device, is made from either stainless steel or titanium. To address this difference, a reference device has been provided, the Medline UNITE® RELEX™ Nitinol Staples, which are
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Medline. The logo features the word "MEDLINE" in a bold, sans-serif font, with the letters arranged horizontally. To the right of the word, there is a stylized star-like symbol with four points. The background of the logo is a solid dark blue color, which provides a strong contrast to the white text and symbol.
made from the exact same nitinol material as the proposed Dynamic Discs. Testing was performed on the Dynamic Discs to ensure the nitinol meets the FDA guidance Technical Considerations for Non-Clinical Assessment of Medical Devices Containing Nitinol. The testing included determining the austenite finish (Af) temperature on final finished devices per ASTM F2004 and performing corrosion susceptibility testing per ASTM F2129.
# Summary of Non-Clinical Testing
## Transformation Temperature
Transformation testing was performed on the subject Medline UNITE® REFLEX™ Dynamic Discs using the DSC method per ASTM F2004 "Standard Test Method for Transformation Temperature of Nickel-Titanium Allovs by Thermal Analysis".
#### Galvanic Corrosion
The Medline UNITE® REFLEX™ Dynamic Discs were tested for galvanic corrosion per ASTM F3044 "Standard Test Method for Evaluating the Potential for Galvanic Corrosion for Medical Implants". The discs and screws representing the largest surface area ratio of the expected cathode material (titanium allov) to the expected anode material (nitinol) were chosen for testing. Microscopic examination of all 3 test samples revealed no pitting or indications of corrosion. The average calculated material release from mass loss, which includes both titanium alloy and nitinol materials, was significantly less than the CDRH recommended parenteral nickel limit value of 35µg/day for a 70kg adult.
#### Corrosion Susceptibility
Corrosion susceptibility testing was conducted per ASTM F2129 "Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements to Determine the Corrosion Susceptibility of Small Implant Devices". The discs, representing the worst-case in surface area, were tested as a construct with the screws implanted into Sawbones® to simulate in-vivo anatomic conditions. All samples achieved electrostatic breakdown potentials in excess of the minimum acceptance criteria, demonstrating acceptable corrosion susceptibility.
## Summary of Clinical Testing
Not applicable.
## Conclusion
In accordance with 21 CFR Part 807, and based on the information provided in this premarket notification, Medline Industries, Inc. concludes that the Medline UNITE® REFLEX™ Dynamic Discs are substantially equivalent to the predicate device, the Medline Cannulated Screws.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.