ULTRASLIDE ACROMIOCLAVICULAR AND SYNDESMOTIC REPAIR DEVICE

K082095 · Smith & Nephew Endoscopy, Inc. · HTN · Oct 21, 2008 · Orthopedic

Device Facts

Record IDK082095
Device NameULTRASLIDE ACROMIOCLAVICULAR AND SYNDESMOTIC REPAIR DEVICE
ApplicantSmith & Nephew Endoscopy, Inc.
Product CodeHTN · Orthopedic
Decision DateOct 21, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Smith & Nephew Ultraslide Acromioclavicular and Syndesmotic Repair Device is intended as an adjunct in fracture repair providing fixation during the healing process following a syndesmotic trauma, such as fixation of acromioclavicular separations due to coracoclavicular ligament disruptions or fixation of ankle syndesmosis due to anterior inferior tibiofibular ligament and/or posterior inferior tibiofibular ligament disruptions.

Device Story

Device provides endoscopic fixation for syndesmotic trauma; used as adjunct in fracture repair. Components include two machined titanium ENDOBUTTON fixation devices connected by continuous polyester suture loop; prethreaded with traction suture for button positioning. Device remains in situ post-healing; no removal required. Used by surgeons in clinical/OR settings to stabilize ligament disruptions. Output is mechanical fixation; aids healing by maintaining bone/ligament alignment during recovery.

Clinical Evidence

Bench testing only. Statistical comparison of cyclical displacement values and tensile strength demonstrated no statistically significant differences between the subject device and predicate devices.

Technological Characteristics

Two machined titanium fixation buttons; continuous polyester suture loop; prethreaded traction suture. Mechanical fixation device. No software or electronic components.

Indications for Use

Indicated for patients requiring fixation during healing following syndesmotic trauma, specifically acromioclavicular separations (coracoclavicular ligament disruptions) or ankle syndesmosis disruptions (anterior/posterior inferior tibiofibular ligament).

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Endoscopy Sinth ? Hernew by 150 Minon nom Band Arthoven, MA. (1) 810 SECTION IV 2000 000 078 140 140 1118 www.tmth.nephose.com K082095-R We are ### OCT 2 1 2002 # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION as required by the Safe Medical Devices Act of 1990 and codified in 21 CFR 807.92 upon which the substantial equivalence is based. Traditional 510(k) Submission of Smith & Nephew Ultraslide Acromioclavicular and Syndesmotic Repair Device Date Prepared: July 22, 2008 A. Submitter's Name: Smith & Nephew, Inc., Endoscopy Division 150 Minuteman Road Andover, MA 01810 ### B. Company Contact Kathy Reddig Regulatory Affairs Specialist II 978-749-1321 (Phone) 978-749-1443 (Fax) kathy.reddig@smith-nephew.com ### C. Device Name | Trade Name: | Smith & Nephew Ultraslide Acromioclavicular and Syndesmotic Repair Device | |----------------------|-------------------------------------------------------------------------------| | Common Name: | Non-Absorbable Suture /Button Retention Device | | Classification Name: | Single / multiple component metallic bone fixation appliances and accessories | ### D. Predicate Devices The Smith & Nephew Ultraslide Acromioclavicular and Syndesmotic Repair Device is substantially equivalent in design, indications and fundamental scientific technology to the following legally marketed devices in commercial distribution: Arthrex Tightrope™ Acromioclavicular (AC) Device (cleared via K052776) and the Arthrex Tightrope™ Syndesmosis Device (cleared via K043248). {1}------------------------------------------------ ### E. Description of Device The Smith & Nephew Ultraslide Device allows for endoscopic fixation following a syndesmotic trauma as Coracoclavicular Separations or Ankle Syndesmotic Disruptions. The device consists of three components; two machined titanium ENDOBUTTON fixation devices on a continuous Polyester suture loop. It is prethreaded with a traction suture that is used for positioning of the fixation button. Although non-absorbable, the Ultraslide device does not require removal, ### F. Intended Use The Smith & Nephew Ultraslide Device is intended as an adjunct in fracture repair providing fixation during the healing process following a syndesmotic trauma, such as fixation of acromioclavicular separations due to coracoclavicular ligament disruptions or fixation of ankle syndesmosis due to anterior inferior tibiofibular ligament and/or posterior inferior tibiofibular ligament disruptions. ## G. Comparison of Technological Characteristics The Smith & Nephew Ultraslide Device is substantially equivalent in design and function to the currently marketed Arthrex Tightrope™. The proposed and the predicate devices both have the same indications and the same fundamental scientific technology. # H. Summary Performance Data Statistical comparison of the test data of the cyclical displacement values and tensile strength demonstrates that there are no statistically significant differences between the Smith & Nephew Ultraslide Device and the predicate devices. {2}------------------------------------------------ Image /page/2/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three legs, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is simple and recognizable, conveying the department's role in public health and welfare. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Smith & Nephew, Inc. Endoscopy Division % Ms. Kathy Reddig Regulatory Affairs Specialist 150 Minuteman Road Andover, MA 01810 OCT 2 1 2008 Re: K082095 > Trade/Device Name: Smith & Nephew Ultraslide Acromioclavicular and Syndesmotic Repair Device Regulation Number: 21 CFR 888.3030 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HTN Dated: July 22, 2008 Received: July 24, 2008 Dear Ms. Reddig: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {3}------------------------------------------------ Page 2 - Ms. Kathy Reddig CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Mark N. Miller Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Indications for Use 510(k) Number (if known): K08 2095 Device Name: Smith & Nephew Ultraslide Acromioclavicular and Syndesmotic Repair Device Indications For Use: The Smith & Nephew Ultraslide Acromioclavicular and Syndesmotic Repair Device is intended as an adjunct in fracture repair providing fixation during the healing process following a syndesmotic trauma, such as fixation of acromioclavicular separations due to coracoclavicular ligament disruptions or fixation of ankle syndesmosis due to anterior inferior tibiofibular ligament and/or posterior inferior tibiofibular ligament disruptions. Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use No er 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Mark A. Millerson (Division Sign-Off) (Division Sign-On) Division of General, Restorative, Division of Neurological Devices 510(k) **Number** Page 18 of 58
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...