TIGHTROPE SYNDESMOSIS REPAIR KIT, TITANIUM, MODEL AR-8920DS; STAINLESS STEEL. MODEL AR-8921DS

K043248 · Arthrex, Inc. · HTN · Feb 16, 2005 · Orthopedic

Device Facts

Record IDK043248
Device NameTIGHTROPE SYNDESMOSIS REPAIR KIT, TITANIUM, MODEL AR-8920DS; STAINLESS STEEL. MODEL AR-8921DS
ApplicantArthrex, Inc.
Product CodeHTN · Orthopedic
Decision DateFeb 16, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Arthrex TightRope™ Syndesmosis Device is intended as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as an adjunct in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting. Specifically, the Arthrex TightRope™ Syndesmosis Device is intended to provide fixation during the healing process following a syndesmotic trauma, such as fixation of syndesmosis (syndesmosis disruptions) in connection with Weber B and C ankle fractures.

Device Story

Device consists of two metal buttons (stainless steel or titanium) pre-threaded with FiberWire suture; includes long straight needle with pull-through suture. Used by surgeons in clinical settings for syndesmotic trauma fixation. Buttons are placed on bone surfaces; suture is tensioned to provide fixation during healing. Acts as alternative to traditional screw fixation. Benefits include stabilization of syndesmosis disruptions in Weber B/C ankle fractures; adjunct support for fracture repair/fixation systems.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Materials: Stainless steel or titanium buttons, FiberWire suture. Form factor: Two-button system with pre-threaded suture. Principle: Mechanical bone fixation via button/suture construct. Non-electronic; no software.

Indications for Use

Indicated for patients requiring fixation during healing of syndesmotic trauma, including syndesmosis disruptions associated with Weber B and C ankle fractures, and as an adjunct for metaphyseal/periarticular small bone fragment fracture repair where screws are not indicated.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K043248 pge f^{2} Pages 11-12 ### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS TightRope™ Syndesmosis Device FEB 1 6 2005 | NAME OF SPONSOR: | Arthrex, Inc.<br>1370 Creekside Boulevard<br>Naples, Florida 34108-1945 | |------------------|--------------------------------------------------------------------------------------------------------------------------| | 510(K) CONTACT: | Sally Foust, RAC<br>Sr. Regulatory Affairs Specialist<br>Telephone: (239) 643-5553 extension 1251<br>FAX: (239) 598-5539 | | TRADE NAME: | TightRope™ Syndesmosis Device | | COMMON NAME: | Button/Suture | #### DEVICE PRODUCT CODE/CLASSIFICATION: HWC: Single/multiple component metailic bone fixation appliances and accessories: 21 CFR 888.3030 MBI: Fastener, Fixation, Nondegradable, Soft Tissue Smooth or threaded metallic bone fixation fastener: 21 CFR 888.3040 GAT: Nonabsorbable poly (ethylene terephthalate) surgical suture: 21 CFR 878.5000 #### PREDICATE DEVICES K033717: TranSet™ Fracture Fixation System (Bonutti) K963172: 4.0 / 4.5 mm Screw (Synthes) K003077: SmartScrew™. Bionx Implants, Ltd. (Linvatec) K030900, OTPS™ Biodegradable Fixation System (Inion Ltd.) K984550, EndoButton (Smith & Nephew) pre-1976, Kirschner (K) Wire (Synthes) #### DEVICE DESCRIPTION AND INTENDED USE The TightRope™ Syndesmosis Device is designed as two differently sized metal buttons, both stainless steel or both titanium, and FiberWire™ suture. The buttons are pre-threaded with FiberWire suture, looped twice through the buttonholes. A long straight needle with pull-through FiberWire suture is also looped through the leading button. The Arthrex TightRope™ Syndesmosis Device is intended as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as an adjunct in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting. {1}------------------------------------------------ page 2 of 2 Specifically, the Arthrex TightRope™ Syndesmosis Device is intended to provide fixation during the healing process following a syndesmotic trauma, such as fixation of syndesmosis (syndesmosis disruptions) in connection with Weber B and C ankle fractures. ## SUBSTANTIALLY EQUIVALENCE Arthrex has determined that the TightRope™ Syndesmosis Device is substantially equivalent to the predicate devices where basic features and intended uses are the same. Any design differences between the Arthrex TightRope™ Syndesmosis Device when compared to predicate devices used in the standard medical practice for the treatment of syndesmosis are considered minor and do not raise any questions concerning safety and effectiveness. Any differences have been found to have no apparent effect on the performance, function, or intended use of the device. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES FEB I & 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Sally Foust, RAC Sr. Regulatory Affairs Specialist Arthrex, Inc. 1370 Creekside Boulevard Naples, Florida 34108-1945 Re: K043248 Trade/Device Name: Arthrex TightRope™ Syndesmosis Devices Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HTN Dated: February 7, 2005 Received: February 8, 2005 Dear Ms. Foust: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Mrs. Foust This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet addresshttp://www.fda.gov/cdrb/industry/support/index.html Sincerely yours. Sincerely yours, Mark N. Wilkinson Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page 10 510(k) Number (if known): ___________________________________________________________________________________________________________________________________________________ # Device Name: Arthrex TightRope™ Syndesmosis Device ### Indications for Use: The Arthrex TightRope™ Syndesmosis Device is intended as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as an adjunct in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting. Specifically, the Arthrex TightRope™ Syndesmosis Device is intended to provide fixation during the healing process following a syndesmotic trauma, such as fixation of syndesmosis (syndesmosis disruptions) in connection with Weber B and C ankle fractures. | Prescription Use<br>(Per 21 CFR 801 Subpart D) | <div style="text-align:center;"><i>Yes</i></div> | |-------------------------------------------------------|--------------------------------------------------| | AND/OR Over-The-Counter Use<br>(21 CFR 801 Subpart C) | <div style="text-align:center;"><i>No</i></div> | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ![Signature](signature.png) (Division Sign-Off) Division of General, Restorative, and Neurological Devices | 510(k) Number | K043248 | |---------------|---------| | Page | 1 of 1 |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...