K210869 · Paragon 28, Inc. · HSB · Jun 28, 2021 · Orthopedic
Device Facts
Record ID
K210869
Device Name
Phantom Hindfoot TTC/TC Nail System
Applicant
Paragon 28, Inc.
Product Code
HSB · Orthopedic
Decision Date
Jun 28, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Phantom® Hindfoot TTC/TC Nail system is intended for tibiotalocalcaneal arthrodesis (fusion) and to provide stabilization of the hindfoot and ankle including the transverse tarsal joints coupling the mid-foot to the hindfoot. Examples of specific indications include: - Post-traumatic or degenerative arthritis - Previously infected arthrosis - Revision of failed ankle arthrodesis - Revision of failed total ankle arthroplasty - Talar deficiency conditions such as avascular necrosis of the talus (requiring tibiocalcaneal arthrodesis) - Neuromuscular deformity or other neuromuscular disease with severe deformity or instability of the ankle - Rheumatoid arthritis - Osteoarthritis - Nonunions or pseudarthrosis of hindfoot and distal tibia - Trauma (severe or malunited tibial pilon fracture) - Charcot foot (neuroarthropathy) - Severe end-stage degenerative arthritis - Instability and skeletal defects after tumor resection - Pantalar arthrodesis - Severe foot/ankle deformity
Device Story
Phantom® Hindfoot TTC/TC Nail System; intramedullary fixation device; consists of nails, screws, and accessory components; used for tibiotalocalcaneal arthrodesis and hindfoot/ankle stabilization; implanted by orthopedic surgeons in clinical settings; provides structural support for bone fusion; nail shape and screw diameter variations accommodate patient anatomy; secures bone segments to facilitate healing in cases of trauma, deformity, or disease.
Clinical Evidence
No clinical data; evidence based on bench testing (static and dynamic mechanical testing per ASTM F1264), a cadaveric implantation study, and biocompatibility evaluation per ISO 10993-1.
Technological Characteristics
Intramedullary nail and bone screw system; metallic construction; mechanical performance tested per ASTM F1264; biocompatibility per ISO 10993-1; includes various sizes/lengths/configurations; non-software device.
Indications for Use
Indicated for patients requiring tibiotalocalcaneal arthrodesis or hindfoot/ankle stabilization due to conditions including post-traumatic/degenerative arthritis, failed arthrodesis/arthroplasty, talar deficiency, neuromuscular deformity, rheumatoid arthritis, nonunions, trauma, Charcot foot, tumor resection, or severe deformity.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Anatomic Ankle Arthrodesis Interlocking Nail (K130147)
Related Devices
K251850 — Phantom® Hindfoot TTC/TC Nail System · Paragon 28, Inc. · Jul 14, 2025
K191782 — TTC Phantom Intramedullary Nail System · Paragon 28, Inc. · Sep 6, 2019
K201227 — Phantom® Hindfoot TTC/TC Nail System · Paragon 28, Inc. · Jul 17, 2020
K192163 — TTC Phantom® Intramedullary Nail System · Paragon 28, Inc. · Jan 31, 2020
K102577 — FORCE MEDICAL IM NAIL SYSTEM · Force Medical, Inc. · May 3, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 28, 2021
Paragon 28, Inc. Haylie Hertz Regulatory Affairs Specialist 14445 Grasslands Dr. Englewood, Colorado 80112
Re: K210869
Trade/Device Name: Phantom® Hindfoot TTC/TC Nail System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB, HWC Dated: May 28, 2021 Received: June 2, 2021
Dear Haylie Hertz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael Owens Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K210869
Device Name Phantom® Hindfoot TTC/TC Nail System
#### Indications for Use (Describe)
The Phantom® Hindfoot TTC/TC Nail system is intended for tibiotalocalcaneal arthrodesis (fusion) and to provide stabilization of the hindfoot and ankle including the transverse tarsal joints coupling the mid-foot to the hindfoot. Examples of specific indications include:
- · Post-traumatic or degenerative arthritis
- · Previously infected arthrosis
- Revision of failed ankle arthrodesis
- · Revision of failed total ankle arthroplasty
- · Talar deficiency conditions such as avascular necrosis of the talus (requiring tibiocalcaneal arthrodesis)
- · Neuromuscular deformity or other neuromuscular disease with severe deformity or instability of the ankle
- Rheumatoid arthritis
- · Osteoarthritis
- Nonunions or pseudarthrosis of hindfoot and distal tibia
- · Trauma (severe or malunited tibial pilon fracture)
- · Charcot foot (neuroarthropathy)
- · Severe end-stage degenerative arthritis
- Instability and skeletal defects after tumor resection
- · Pantalar arthrodesis
- · Severe foot/ankle deformity
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
| 510(k) Number: | K210869 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name: | Phantom® Hindfoot TTC/TC Nail System |
| Manufacturer: | Paragon 28, Inc.<br>14445 Grasslands Dr.<br>Englewood, CO 80112 |
| Contact: | Ms. Haylie Hertz<br>Regulatory Affairs Specialist<br>Phone: 303.720.0017<br>hhertz@paragon28.com |
| Date Prepared: | May 28, 2021 |
| Classification: | 21 CFR §888.3020, Intramedullary Nail, Bone Screw<br>21 CFR §888.3040, Screw, fixation, bone |
| Class: | II |
| Product Code: | HSB, HWC |
| Common Name: | Intramedullary Nail |
| Predicate Device: | Phantom® Hindfoot TTC/TC Nail System (K201227) |
| Reference Device: | Anatomic Ankle Arthrodesis Interlocking Nail (K130147) |
| Indications for Use: | The Phantom® Hindfoot TTC/TC Nail System is intended for<br>tibiotalocalcaneal arthrodesis (fusion) and to provide stabilization of<br>the hindfoot and ankle including the transverse tarsal joints coupling<br>the mid-foot to the hindfoot. Examples of specific indications<br>include:<br>• Post-traumatic or degenerative arthritis<br>• Previously infected arthrosis<br>• Revision of failed ankle arthrodesis<br>• Revision of failed total ankle arthroplasty<br>• Talar deficiency conditions such as avascular necrosis of the<br>talus (requiring tibiocalcaneal arthrodesis)<br>• Neuromuscular deformity or other neuromuscular disease<br>with severe deformity or instability of the ankle<br>• Rheumatoid arthritis |
{4}------------------------------------------------
- Osteoarthritis
- Nonunions or pseudarthrosis of hindfoot and distal tibia ●
- Trauma (severe or malunited tibial pilon fracture) ●
- Charcot foot (neuroarthropathy) ●
- Severe end-stage degenerative arthritis
- Instability and skeletal defects after tumor resection
- Pantalar arthrodesis
- Severe foot/ankle deformity ●
Device Description: The Paragon 28® Phantom® Hindfoot TTC/TC Nail System is comprised of intramedullary nails, screws and accessory components. The Phantom® nails are offered in a variety of sizes lengths, and configurations to accommodate variations in patient anatomy. The Phantom® screws insert through the intramedullary nail to secure the construct. These are offered in varying lengths to accommodate the anatomical fixation required.
# Substantial Equivalence:
The intended use of the subject device is the same as the predicate device. In addition, the subject device possesses the same technological characteristics as the predicate device, including performance, basic design, material, manufacturing, and sizes. Differences between the Phantom® nails and the predicate device (i.e. nail shape and screw diameter) were shown not to raise new questions of safety and effectiveness. Therefore, the Phantom® Hindfoot TTC/TC Nail System is substantially equivalent to the predicate device with respect to indications, design, function, and performance.
# Preclinical Testing:
Static and dynamic mechanical testing per ASTM F1264 was conducted on the Phantom® Hindfoot Curved Nail. A cadaveric implantation study was also conducted. Biocompatibility was evaluated per ISO 10993-1 and CDRH's 2020 Biocompatibility Guidance Document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Additional testing includes bacterial endotoxins and sterilization.
## Conclusion:
The Phantom® Hindfoot Curved Nail has the same intended use and equivalent technological characteristics as the predicate devices. Additionally, the preclinical testing supports the substantial equivalence of the Phantom® Hindfoot Curved Nail to the predicate. Therefore, the Phantom® Hindfoot TTC/TC Nail System is substantially equivalent for its intended use.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.