ANATOMIC ANKLE ARTHRODESIS INTERLOCKING NAIL (A3 INTERLOCKING NAIL)
K130147 · Small Bone Innovations, Inc. · HSB · May 3, 2013 · Orthopedic
Device Facts
Record ID
K130147
Device Name
ANATOMIC ANKLE ARTHRODESIS INTERLOCKING NAIL (A3 INTERLOCKING NAIL)
Applicant
Small Bone Innovations, Inc.
Product Code
HSB · Orthopedic
Decision Date
May 3, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SBi Anatomic Ankle Arthrodesis Interlocking Nail System intended for tibiotalocalcaneal arthrodesis of the ankle following: Post traumatic and degenerative arthritis involving both ankle and subtalar joints, osteoarthritis, Rheumatoid arthritis, Pseudoarthrosis, Severe foot/ankle deformity, or Instability and skeletal defects after tumor resection. These include neuro-osteoarthroplasty (Charcot's foot), avascular necrosis of the talus, failed joint replacement, failed ankle fusion, and for distal tibia fracture non-unions when used concomitantly with tibiotalocalcaneal pathology.
Device Story
System consists of intramedullary fixation implants (nails, locking screws), instruments, and trays. Used for tibiotalocalcaneal arthrodesis. Implants inserted into medullary canal of long bones to provide fixation. Shorter length screws added to existing system. Implants made of titanium; instruments/trays made of aluminum, stainless steel, PEEK, POM, and composites. Single-use only. Provided sterile or non-sterile. Used by orthopedic surgeons in clinical/surgical settings. Output is mechanical fixation of bone segments to facilitate fusion. Benefits patient by stabilizing joints and correcting deformities.
Clinical Evidence
Bench testing only. Engineering rationale provided to demonstrate that shorter screw lengths maintain structural integrity and that existing sterilization parameters remain valid for the modified configuration.
Indicated for patients requiring tibiotalocalcaneal arthrodesis due to post-traumatic/degenerative arthritis (ankle/subtalar), rheumatoid arthritis, osteoarthritis, pseudoarthrosis, severe foot/ankle deformity, instability/skeletal defects post-tumor resection, Charcot's foot, avascular necrosis of the talus, failed joint replacement, failed ankle fusion, or distal tibia fracture non-unions with tibiotalocalcaneal pathology.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Predicate Devices
SBi Anatomic Ankle Arthrodesis Interlocking Nail System (K112982)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Premarket Notification Summary of Safety and Effectiveness
MAY 0 3 2013
### Submission Information
| Manufacturer: | Small Bone Innovations, Inc. |
|------------------|--------------------------------|
| | 1380 South Pennsylvania Avenue |
| | Morrisville, PA 19067 |
| Ph: 215-428-1791 | Fax: 215-428-1795 |
Submitted By:
Small Bone Innovations, Inc. Joseph Eble 1380 South Pennsylvania Avenue Morrisville, PA 19067
Date: January 24, 2013, Revised March 27, 2013
Proprietary Name: SBi Anatomic Ankle Arthrodesis Interlocking Nail System
Classification name/Identification: Class II, An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures
Product Code: HSB
Common/Usual Name and Reference Number: Intramedullary fixation rod - 21 CJFR 888.3020
Substantial Equivalence: Documentation is provided which demonstrates the modified devices are substantially equivalent to the devices of the SBi Anatomic Ankle Arthrodesis Interlocking Nail System.
Device Description: The SBi Anatomic Ankle Arthrodesis Interlocking Nail System consists of implants, associated instruments, and trays. The implants are made of implantable grade titanium while the instrument components and trays are made of several materials: Aluminum, stainless steel, PEEK, POM, and composite materials.
This Premarket Notification is to add 5.0mm Tibial Locking Screws in 20 to 24mm lengths, 5.0 mm Talar Screws with Threaded Head in 40 to 44mm lengths, and 6.0mm Calcaneal Locking Screws with Threaded Head in 50 to 68mm lengths to the system.
Intended Use: The SBi Anatomic Ankle Arthrodesis Interlocking Nail System intended for tibiotalocalcaneal arthrodesis of the ankle following: Post traumatic and degenerative arthritis involving both ankle and subtalar joints, osteoarthritis, Rheumatoid arthritis, Pseudoarthrosis, Severe foot/ankle deformity, or Instability and skeletal defects after tumor resection. These include neuro-osteoarthroplasty (Charcot's foot), avascular necrosis of the talus, failed joint
{1}------------------------------------------------
replacement, failed ankle fusion, and for distal tibia fracture non-unions when used concomitantly with tibiotalocalcaneal pathology.
The implants are intended for single use only and will be offered sterile and non-sterile.
The shorter length screws being added by this Special 510(k) are made from implantable grade TiAl6V4 titanium (ASTM F136) which is the same material as the predicate screws.
The design of the shorter screws being added by this Special 510(k) are the same as the predicate screws in that the head designs of the outer and inner diameters and thread are the same. The only difference is that the modified screws are shorter in length.
Current Devices: The modified devices are equivalent to the current devices of the SBi Anatomic Ankle Arthrodesis Interlocking Nail System (K112982).
An Engineering Rationale is provided in the Bench Testing section which concludes that the bending loads will be lower if the screw is shorter while keeping the minor diameter the same.
Additionally the rationale concludes that the total mass of the new set configuration is decreasing (longer screws are being replaced by shorter screws); thus the existing sterilization parameters will still be valid.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle with three legs, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the eagle. The eagle is facing right, and its wings are represented by curved lines.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 3, 2013
Small Bone Innovations, Incorporated % Mr. John Minier 1380 South Pennsylvania Avenue Morrisville, Pennsylvania 19067
Re: K130147
Trade/Device Name: SBi Anatomic Ankle Arthrodesis Interlocking Nail System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: April 1, 2013 Received: April 2, 2013
Dear Mr. Minier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
## Page 2 - Mr. John Minier
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
# Erin DKeith
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 4 Statement of Indications for Use
510(k) Number: K130147
Device Name: SBi Anatomic Ankle Athrodesis Interlocking Nail System
Indications For Use:
The SBi Anatomic Ankle Athrodesis Interlocking Nail System is indicated for tibiotalocalcaneal arthrodesis of the ankle following:
- Post traumatic and degenerative arthritis involving both ankle and subtalar joints,
- Rheumatoid arthritis,
- Osteoarthritis,
- Pseudoarthrosis,
- Severe foot/ankle deformity, or
- Instability and skeletal defects after tumor resection.
These include neuro-osteoarthroplasty (Charcot's foot), avascular necrosis of the talus, failed joint replacement, failed ankle fusion, and for distal tibia fracture non-unions when used concomitantly with tibiotalocalcaneal pathology.
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Casey L. Hanley, Ph.D.
Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.