The MacroPoreOS Protective Sheet is intended to maintain the relative position of weak bony tissue such as bone grafts, bone graft substitutes, or bone fragments from comminuted fractures. The MacroPoreOS Protective sheet is also indicated for cement restriction in total joint arthroplasty procedures. Only when used in conjunction with traditional rigid fixation, the MacroPoreOS Protective Sheet is intended to maintain the relative position weak bony tissue in trauma and reconstructive orthopedic procedures involving: - Long bones - Flat bones - Short bones - Irregular bones - Appendicular skeleton - Thorax. When used alone (without traditional rigid fixation), the MacroPoreOS Protective Sheet is intended to maintain the relative position of bone graft substitutes in reconstructive orthopedic procedures involving: - Tumor resections where bone strength has not been compromised - Iliac crest harvests. This device is not intended for use in the spine. The device is not intended for load bearing indications unless used in conjunction with traditional rigid fixation.
Device Story
Resorbable, macroporous implant sheet made of polylactic acid (PLA); malleable when heated to ~65°C; cut to size by surgeon. Used in orthopedic procedures to stabilize bone grafts, substitutes, or fragments; acts as cement restrictor in joint arthroplasty. Applied alone or with internal fixation (plates/screws). Provides temporary structural support; degrades in vivo over 18 months. Used in clinical settings by surgeons. Benefits include containment of graft materials and stabilization of bone fragments without requiring permanent hardware removal.
Clinical Evidence
Bench testing only. Accelerated aging studies demonstrated strength retention for 9 months with gradual decrease to zero at 18 months. In vitro testing confirmed mechanical properties remain stable after heating to 60°C for 120 minutes. Crystallinity testing via DSC confirmed amorphous structure.
Technological Characteristics
Material: Poly (L-lactide-co-D,L-lactide) 70:30 (amorphous PLA). Form: Macroporous sheet, 0.50-2.0 mm thickness, 20x20 mm to 120x120 mm. Pore size: 500-2500 microns. Malleable when heated to 65°C. Resorbable. Non-load bearing unless used with rigid fixation.
Indications for Use
Indicated for maintaining position of bone grafts, substitutes, or fragments in orthopedic trauma and reconstructive procedures. Used with rigid fixation for long, flat, short, irregular bones, appendicular skeleton, and thorax. Used alone for tumor resections (where bone strength is intact) and iliac crest harvests. Also indicated for cement restriction in total joint arthroplasty. Contraindicated for spinal use and load-bearing applications without rigid fixation.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{1}------------------------------------------------
K994158
# K994158 Summary
### MacroPoreOS Protective Sheet
Page 1 of 4
# ADMINISTRATIVE INFORMATION
Manufacturer Name:
Official Contact:
MacroPore, Inc. 6740 Top Gun Street San Diego, CA 92121
Kenneth K. Kleinhenz Director of Regulatory Affairs Telephone (858) 458-0900 Fax (858) 458-0994
DEVICE NAME
Classification Name:
Trade/Proprietary Name:
MacroPoreOS Protective Sheet
Plate, Bone
# ESTABLISHMENT REGISTRATION NUMBER 2031733
## DEVICE CLASSIFICATION AND PRODUCT CODE
As shown in 21CFR 888.3030 Bone Fixation Appliances are intended for use in orthopedic procedures and are classified as Class II. Bone Plates have been assigned Product Code HRS.
# INTENDED USE
The MacroPoreOS Protective Sheet is intended to maintain the relative position of weak bony tissue such as bone grafts, bone graft substitutes, or bone fragments from comminuted fractures. The MacroPoreOS Protective sheet is also indicated for cement restriction in total joint arthroplasty procedures.
Only when used in conjunction with traditional rigid fixation, the MacroPoreOS Protective Sheet is intended to maintain the relative position weak bony tissue in trauma and reconstructive orthopedic procedures involving:
- · Long bones
- Flat bones
- · Short bones
- · Irregular bones
- · Appendicular skeleton
- · Thorax
When used alone (without traditional rigid fixation), the MacroPoreOS Protective Sheet is intended to maintain the relative position of bone graft substitutes in reconstructive orthopedic procedures involving:
- · Tumor resections where bone strength has not been compromised
- · Iliac crest harvests
This device is not intended for use in the spine. The device is not intended for load bearing indications unless used in conjunction with traditional rigid fixation.
{2}------------------------------------------------
## DEVICE DESCRIPTION
#### Design Characteristics
Design Charactive Sheet is a resorbable, macroporous implant in sheet form manufactured from polylactic acid (PLA). MacroPoreOS Protective Sheet can be cut with scissors to the desired shape and size. The MacroPore Power Pen can also be used to cut or shape the MacroPoreOS Protective Sheet to the desired shape or size. MacroPoreOS Protective Sheet is fully malleable when heated to approximately 65℃ (for example, by the use of sterile hot water), and thus can be conformed three dimensionally to most any anatomical orientation. The MacroPoreOS Protective Sheet can be rolled into a tube or used as a flat sheet. MacroPoreOS Protective Sheet can be used either alone or in conjunction with internal bone fixation devices such as plates and screws, which also can serve to further stabilize the anatomical region. The MacroPoreOS System includes MacroPoreOS Protective Sheets, a selection of resorbable MacroSorb Screws, MacroSorb Tacks and associated manual instruments.
MacroPoreOS Protective Sheet is provided in sheets of 20 x 20 mm to 120 x 120 mm and will be provided in other sizes as needed for particular surgical procedures. The pore size ranges from 500 microns to 2500 microns in diameter, with pores distributed uniformly throughout the sheet in an offset or aligned pattern. The thickness of the MacroPoreOS Protective Sheet ranges from 0.50 mm to 2.0 mm according to the orthopedic region to be treated, however, not to exceed a total mass of 18 grams.
#### Material Composition
The MacroPoreOS Protective Sheet is fabricated from polylactic acid (PLA).
#### In Vitro Testing
Because the MacroPoreOS Protective Sheet is intended to be heated in the surgical suite to temperatures above the material's glass transition temperature to facilitate shaping to anatomic structures, testing was performed to determine the effect of prolonged heating in saline at 60°C on inherent viscosity. The testing demonstrates that viscosity stayed within an appropriate range over 120 minutes. Therefore, the relatively brief exposure anticipated during the surgical preparation of MacroPoreOS Protective Sheet is not expected to have a significant effect on its mechanical properties.
Accelerated aging testing was performed on MacroPoreOS Protective Sheet. Testing demonstrated that the MacroPoreOS Protective Sheet is as rigid and as strong as the predicate after a simulated 6 month in vivo exposure. Furthermore, simulated in vivo accelerated testing indicates that the MacroPoreOS Protective Sheet retains all of its strength for the first 9 months and a steadily decrease in strength to zero after approximately 18 months.
Mechanical testing was performed on the MacroPoreOS Protective Sheet which determined the MacroPoreOS Protective Sheet to be substantially equivalent to the mechanical strengths of the predicate devices under indication for use conditions.
{3}------------------------------------------------
#### K994158 Summary
Crystallinity was tested for by DSC (differential scanning calorimetery). This test measures the amount Of heat energy that is absorbed by a material. A crystalline material will require more energy once it of near it's melting point. This release of heat energy can be seen on a graph as a sharp spike and is referred to as a "melting endotherm". The tests ran on the sterile and non-sterile samples revealed no endothermic spikes. From this we verify the implants are amorphous or non-crystalline.
# EQUIVALENCE TO MARKETED PRODUCT
MacroSorbOS Protective Sheet shares indications and design principles with the following predicate device, which has been determined by FDA to be substantially equivalent to the following preamendment devices: Sofamar Danek Timesh (K974017).
#### Indications For Use
The MacroPoreOS Protective Sheet shares indications for use with the predicate device as both the I he Macrol oroon Trotective Sheet and the Timesh predicated for reinforcing weak bony tissue in orthopedic procedures. The MacroPoreOS Protective Sheet and the Timesh predicate also share cement restriction indication for use.
#### Design and Materials
Design and MacroSorbOS Protective Sheet and the predicate device (Sofamar Danek Timesh) are similar, consisting of a thin semi-rigid sheet with macroporosity. Both the predicate device I milesh) are similar, consistent ave a semi-rigid construction with pores of similar diameter and spacing. The pore size and spacing of the predicate device is within the pore size and spacing specifications of the MacroSorbOS Protective Sheet. The dimensions of the predicate device are also operiodions or the MacroSorbOS Protective sheet as both devices are provided in rectangular sheets that oompanders to the mechanical characteristics of the MacroSorbOS Protective Sheet are substantially equivalent to the predicate device with respect to initial and in vivo strengths and rigidity. In addition to physical characteristics, both the predicate device and the MacroSorbOS Protective Sheet can be cut to specific shapes and sizes by the end user. The titanium device differs from MacroSorbOS Protective Sheet device in that it may be left in place permanently or must be removed surgically, whereas the polymer devices are intended to be metabolized by the body and do not require removal.
{4}------------------------------------------------
| . .<br>- THE STORESSONATION<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| . | |
:
ત્તા
# UMMARY : TABLE OF SUBSTANTIAL EQUIVALENCE
.
| Intended Use | Subject Device | Predicate Device | Related Device | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The MacroPoreOS Protective Sheet is<br>intended to maintain the relative<br>position of weak bony tissue such as<br>bone grafts, bone graft substitutes, or<br>bone fragments from comminuted<br>fractures. The MacroPoreOS<br>Protective sheet is also indicated for<br>cement restriction in total joint<br>arthroplasty procedures. Only when<br>used in conjunction with traditional<br>rigid fixation, the MacroPoreOS<br>Protective Sheet is intended to<br>maintain the relative position weak<br>bony tissue in trauma and<br>reconstructive orthopedic procedures<br>involving: Long bones, Flat bones,<br>Short bones, Irregular bones,<br>Appendicular skeleton, Thorax. When<br>used alone (without traditional rigid<br>fixation), the MacroPoreOS Protective<br>Sheet is intended to maintain the<br>relative position of bone grafts or<br>bone graft substitutes in reconstructive<br>orthopedic procedures involving:<br>Tumor resections where bone strength<br>has not been compromised, Iliac crest<br>harvests. This device is not intended<br>for use in the spine. The device is not<br>intended for load bearing indications<br>unless used in conjunction with<br>traditional rigid fixation. | MacroSorboS Protective<br>Sheet | Timesh (K974017)<br>Sofamar Danek<br>For use in any oral-maxillo-cranio-<br>facial surgical reconstructive<br>procedure, either orthognathic or<br>trauma, wherein rigid or semi-rigid<br>internal fixation is utilized as a<br>means of holding bone fragments<br>together. Alternatively, the Timesh<br>system is also indicated for use in<br>reinforcing weak bony tissues in<br>orthopaedic surgical procedures<br>such as pelvic reconstructions,<br>acetabular reconstruction, and<br>cement restriction. This product is<br>not intended for spinal use. | MacroPore Protective Sheet<br>(Protego System)<br>(K972913)<br>MacroPore Protective Sheet is<br>intended for use in trauma and<br>reconstructive procedures in the<br>midface and craniofacial skeleton:<br>1. Comminuted fractures of the<br>naso-ethmoidal and infraorbital<br>areas 2. Comminuted fractures of<br>the frontal sinus wall 3. Trauma<br>of the midface or craniofacial<br>skeleton 4. Reconstructive<br>procedures of the midface or<br>craniofacial skeleton. The system<br>is not intended for use in the<br>mandible and/or for full load<br>bearing procedures. | MacroPore Protective Sheet<br>(K983360)<br>MacroPore Protective Sheet is<br>intended to facilitate healing in<br>trauma, reconstruction and bone<br>augmentation procedures of the<br>mandible. The following specific<br>indications are included: to<br>maintain the relative position of<br>bony fragments in trauma and<br>bone graft porcedures, and to<br>contain and prevent migration and<br>shifting of autograft, allograft<br>and/or bone graft substitutes that<br>may be necessary in<br>reconstructive procedures. |
| Design | Sheets of 0.50 - 2.0 mm<br>thickness, sizes 20 x 20 mm to<br>120 x 120 mm or as required. | Sheets of 0.2mm thickness,<br>size 76 x 45 mm. | Plates, screws and tacks of<br>various sizes. | Sheets of 0.50-2.0 mm<br>thickness, sizes 20 x 20 mm to<br>120 x 120 mm or as required. |
| Material | Poly (L-lactide-co-D,L-<br>lactide) 70:30, amorphous | Titanium | Poly (L-lactide-co-D,L-lactide)<br>70:30, amorphous | Poly (L-lactide-co-D,L-lactide)<br>70:30,<br>amorphous |
| Product Code | HRS | HRS | HRS and HWC | JEY |
.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around it. Inside the circle is an abstract symbol resembling a stylized human figure or bird in flight, composed of three curved lines.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 2 4 2000
Mr. Kenneth K. Kleinhenz Director of Regulatory Affairs MacroPore, Inc. 6740 Top Gun Street San Diego, California 92121
Re: K994158
Trade Name: MacroSorbOS Protective Sheet Regulatory Class: II Product Codes: HRS, MAI, HWC Dated: April 24, 2000 Received: April 25, 2000
Dear Mr. Kleinhenz:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
# Page 2 - Mr. Kenneth K. Kleinhenz
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
ume R. bochner -
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Device Name: MacroPoreOS Protective Sheet
# Indications for Use:
Indications for Use:
The MacroPoreOS Protective Sheet is intended to maintain the relative fromers, from was The MacrororeOS Protective Sheet is meanse substitutes, or bone fragments from a bony tissue such as bone graits, bone gran-sacething, and indicated for cement restriction in total joint arthroplasty procedures.
Only when used in conjunction with traditional rigid fration, the MacroPoreoSings in traves Only when is intended to maintain the relative position weak bony tissue in trauma and reconstructive orthopedic procedures involving:
- · Long bones
- · Flat bones
- Short bones
- · Irregular bones
- · Appendicular skeleton
- Thorax
When used alone (without traditional rigid fixation), the MacroPoreOS Protective Sheet When used alone (without traditional right and in the grafts or bone grafts substitutes in reconstructive orthopedic procedures involving:
- · Tumor resections where bone strength has not been compromised
- · Iliac crest harvests
This device is not intended for use in the spine. The device is not intended for load I mis device is nor meets used in conjunction with traditional rigid fixation.
# (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NECESSARY) 」 2 w = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =
Concurrence of CDRH, Office of Device Evaluation (ODE)
X Prescription Use _
OR
Over-The-Counter Use
Mina R. lochner'
Division Sign Off
(Division Sign-Off) Division of General Restorative Devices K994158 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.