ACE HUMERUS & RADIUS PLATES
Device Facts
| Record ID | K955472 |
|---|---|
| Device Name | ACE HUMERUS & RADIUS PLATES |
| Applicant | Ace Medical Co. |
| Product Code | HRS · Orthopedic |
| Decision Date | Apr 1, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3030 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The Distal Radius Plate System is designed for internal fixation of the distal radius. The distal radius plates are available in left and right volar configurations and in two sizes, small and large. The Humerus Plate System is designed for internal fixation of the proximal humerus. The humerus plates are available in right and left configurations and in three hole and five hole length. Cortical and cancellous bone screws are required to secure the plates to the bone. A variety of screw are available for this application in order to accommodate different fracture types and locations, bone density and surgeon preference.
Device Story
Anatomically contoured titanium alloy bone plates; used for internal fixation of distal radius and proximal humerus fractures; requires cortical or cancellous bone screws for attachment; plates available in multiple hole configurations and sizes; used by orthopedic surgeons in clinical settings; provides mechanical stabilization of bone fractures to facilitate healing; performance validated via static cantilever bending tests compared to predicate devices.
Clinical Evidence
Bench testing only. Static cantilever bending tests performed on proximal humerus, volar distal radius, and dorsal distal radius plates. Results compared to Synthes and DePuy predicate devices; ACE plates demonstrated comparable bending strength.
Technological Characteristics
Biocompatible titanium alloy construction; anatomically contoured bone plates; configurations include left/right, multiple hole counts (3, 5), and sizes (small, large); mechanical fixation via cortical/cancellous bone screws; non-powered, passive orthopedic implant.
Indications for Use
Indicated for internal fixation of distal radius fractures and proximal humerus fractures in patients requiring bone stabilization.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- DePuy ORIF Small Fragment Plates
- Synthes Small Plates
- Synthes Small Titanium Plates