Fine Osteotomy™

K253300 · Bodycad Laboratories, Inc. · HRS · Oct 28, 2025 · Orthopedic

Device Facts

Record IDK253300
Device NameFine Osteotomy™
ApplicantBodycad Laboratories, Inc.
Product CodeHRS · Orthopedic
Decision DateOct 28, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

Fine Osteotomy™ is a system intended for open- and closed-wedge osteotomies, treatment of bone and joint deformities, fixation of fractures and malalignment caused by injury or disease, such as osteoarthritis, of the distal femur and proximal tibia. Fine Osteotomy disposable instrumentation is intended to assist in pre-operative planning and/or in guiding the marking of bone and/or guiding of surgical instruments in non-acute, non-joint replacing osteotomies around the knee. Fine Osteotomy is a patient-specific device.

Device Story

Fine Osteotomy is a patient-specific orthopedic fixation system for distal femur and proximal tibia osteotomies. System comprises anatomically contoured titanium alloy (Ti-6Al-4V ELI) plates and compatible self-tapping screws; includes disposable instrumentation for pre-operative planning and surgical guidance. Device functions as load-sharing internal fixation to stabilize bone fragments and maintain alignment during healing. Used by surgeons in non-acute, non-joint replacing knee osteotomies. Closed-wedge plates feature an oblong hole for controlled axial compression. Device is purely mechanical; no electrical components or software.

Clinical Evidence

Bench testing only. Mechanical engineering analysis verified the structural integrity of the modified closed-wedge plates (oblong hole). Surface finish evaluation confirmed the glass-bead-blasted finish meets smoothness requirements. No clinical data were required.

Technological Characteristics

Materials: Titanium alloy Ti-6Al-4V ELI (ASTM F136). Principle: Mechanical load-sharing internal fixation. Form factor: Low-profile, anatomically contoured plates with self-tapping screws. Features: Oblong hole for axial compression; glass-bead-blasted surface finish. Energy source: None (mechanical). Connectivity: None. Software: None.

Indications for Use

Indicated for patients requiring open- or closed-wedge osteotomies, treatment of bone/joint deformities, or fixation of fractures and malalignment due to injury or disease (e.g., osteoarthritis) of the distal femur and proximal tibia.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION October 28, 2025 Bodycad Laboratories Inc. Nadine Adia Regulatory Affairs and Quality Director 2035 rue du Haut-Bord Quebec, G1N4R7 Canada Re: K253300 Trade/Device Name: Fine Osteotomy™ Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC, PBF Dated: September 29, 2025 Received: September 29, 2025 Dear Nadine Adia: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K253300 - Nadine Adia Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K253300 - Nadine Adia Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, CHRISTOPHER FERREIRA -S Christopher Ferreira, M.S. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253300 | ? | | Please provide the device trade name(s). | | ? | | Fine Osteotomy™ | | | | Please provide your Indications for Use below. | | ? | | Fine Osteotomy™ is a system intended for open- and closed-wedge osteotomies, treatment of bone and joint deformities, fixation of fractures and malalignment caused by injury or disease, such as osteoarthritis, of the distal femur and proximal tibia. Fine Osteotomy disposable instrumentation is intended to assist in pre-operative planning and/or in guiding the marking of bone and/or guiding of surgical instruments in non-acute, non-joint replacing osteotomies around the knee. Fine Osteotomy is a patient-specific device. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} 510(k) #: K253300 510(k) Summary Prepared on: 2025-09-29 | Contact Details | 21 CFR 807.92(a)(1) | | --- | --- | | Applicant Name | Bodycad Laboratories Inc. | | Applicant Address | 2035 rue du Haut-Bord Quebec G1N4R7 Canada | | Applicant Contact Telephone | +14186229250 | | Applicant Contact | Mrs. Nadine Adia | | Applicant Contact Email | nadia@bodycad.com | | Device Name | 21 CFR 807.92(a)(2) | | Device Trade Name | Fine Osteotomy | | Common Name | Single/multiple component metallic bone fixation appliances and accessories | | Classification Name | Plate, Fixation, Bone | | Regulation Number | 888.3030 | | Product Code(s) | HRS, HWC, PBF | | Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | K250923 | Fine Osteotomy™ | | Product Code | HRS | | Device Description Summary | 21 CFR 807.92(a)(4) | | The Fine Osteotomy™ System is an orthopedic fixation system used to stabilize bone fragments during corrective osteotomies of the distal femur or proximal tibia. The system supports proper alignment and promotes bone healing through rigid internal fixation. The system includes: Metallic bone plates (femoral and tibial) in open- and closed-wedge configurations. Bone screws compatible with the plates. Reusable surgical instruments for positioning and fixation. The plates and screws are made of titanium alloy Ti-6Al-4V ELI, a biocompatible material widely used in orthopedic implants. The closed-wedge plates now include an oblong hole to enable controlled axial compression at the osteotomy site. An additional glass-bead-blasted surface finish for the plates has been validated to meet the same smoothness requirement as the existing polished finish. The plates are low-profile, anatomically contoured, and fixed with compatible self-tapping screws. The device functions as a load-sharing internal fixation system to maintain alignment during healing. | | | Intended Use/Indications for Use | 21 CFR 807.92(a)(5) | | Fine Osteotomy™ is a system intended for open- and closed-wedge osteotomies, treatment of bone and joint deformities, fixation of fractures and malalignment caused by injury or disease, such as osteoarthritis, of the distal femur and proximal tibia. Fine Osteotomy disposable instrumentation is intended to assist in pre-operative planning and/or in guiding the marking of bone and/or guiding of surgical instruments in non-acute, non-joint replacing osteotomies around the knee. Fine Osteotomy is a patient-specific device. | | | Indications for Use Comparison | 21 CFR 807.92(a)(5) | {5} Fine Osteotomy™ is a system intended for open- and closed-wedge osteotomies, treatment of bone and joint deformities, fixation of fractures and malalignment caused by injury or disease, such as osteoarthritis, of the distal femur and proximal tibia. Fine Osteotomy disposable instrumentation is intended to assist in pre-operative planning and/or in guiding the marking of bone and/or guiding of surgical instruments in non-acute, non-joint replacing osteotomies around the knee. Fine Osteotomy is a patient-specific device. ## Technological Comparison 21 CFR 807.92(a)(6) The subject Fine Osteotomy™ System has the same fundamental technological characteristics as the predicate device (K250923). Both systems share the same principle of operation (rigid internal fixation of the osteotomy site), intended use (stabilization of bone fragments in planned osteotomies of the distal femur or proximal tibia), design concept (anatomically contoured metallic plates fixed with compatible screws), energy source (mechanical fixation only, no electrical components), and material (titanium alloy Ti-6Al-4V ELI per ASTM F136). The only changes introduced are: - The addition of an oblong hole in the closed-wedge plates to allow controlled axial compression at the osteotomy site. - The addition of a glass-bead-blasted surface finish for the plates. These modifications do not change the principle of operation, materials, or overall device function and do not raise new questions of safety or effectiveness compared to the predicate device. ## Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) Verification and Validation Activities: - Mechanical engineering analysis confirmed that the addition of the oblong hole in the closed-wedge plates does not introduce new risks or worst-case scenarios and that the device continues to meet performance requirements. - Surface finish evaluation confirmed that the new glass-bead-blasted finish meets smoothness requirements and does not introduce new risks to safety or performance. All results met predefined acceptance criteria. No new clinical data were required, as the changes do not affect the intended use or raise new questions of safety or effectiveness. The device remains safe, effective, and substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...