The RibFix Advantage System is indicated for the fixation, stabilization, and fusion of rib fractures and osteotomies of normal and osteoporotic bone.
Device Story
The RibFix Advantage System is a thoracic fixation device consisting of titanium bridges with threaded locking posts, locking caps, and optional washers. It is used for the thoracoscopic fixation and stabilization of rib fractures and osteotomies. The device is implanted by a surgeon; the bridge is placed on the underside of the rib, with locking posts extending through pre-drilled holes to be secured by locking caps on the anterior side. The system provides mechanical stabilization to facilitate bone fusion. It is supplied non-sterile and requires steam sterilization by the user. The device is intended to benefit patients by providing rigid fixation for rib fractures, potentially improving healing outcomes.
Clinical Evidence
No clinical data was required or provided for this submission. Substantial equivalence was established through non-clinical performance testing, including MR safety evaluations (displacement force, torque, RF heating, and image artifacts) conducted in accordance with ASTM F2052-15, ASTM F2213-17, ASTM F2182-11a, and ASTM F2119-07.
Technological Characteristics
Implants manufactured from commercially pure titanium (ASTM F67) and titanium alloys (Ti-6Al-7Nb per ASTM F1295; Ti-6Al-4V per ASTM F136/F1472). System includes bridges with threaded locking posts, locking caps, and washers. Mechanical fixation principle. Provided non-sterile; steam sterilization required. No software or electronic components.
Indications for Use
Indicated for fixation, stabilization, and fusion of rib fractures and osteotomies in patients with normal or osteoporotic bone. Contraindicated in patients with active/latent infection, sepsis, device material sensitivity, or those unable to follow postoperative instructions. Not for use on clavicle, spine, or highly comminuted fractures.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Biomet Microfixation RibFix Blu Thoracic Fixation System (K162974)
Biomet Microfixation SternaLock 360 Sternal Closure Fixation System (K163007)
Related Devices
K183317 — AdvantageRib System · Sig Medical Corporation · Feb 14, 2019
K182179 — AdvantageRib Anterior System · Sig Medical Corp. · Nov 16, 2018
K241282 — RibFix Titan Fixation System · Riverpoint Medical · Aug 19, 2024
K141241 — MATRIXRIB ENDO THORACOSCOPIC RIB PLATING SYSTEM · Synthes (USA) Products, LLC · Aug 7, 2014
K252166 — RIB LINK Fixation System · Globus Medical, Inc. · Feb 26, 2026
Submission Summary (Full Text)
{0}------------------------------------------------
January 28, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Biomet Micofixation Lauren Jasper Regulatory Affairs Manager 1520 Tradeport Drive Jacksonville, Florida 32218
Re: K203474
Trade/Device Name: Biomet Microfixation RibFix Advantage System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS. HWC. HTN Dated: November 24, 2020 Received: November 25, 2020
Dear Lauren Jasper:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K203474
Device Name Biomet MicrofixationR ibFix Advantage System
Indications for Use (Describe)
TheR ibFix Advantage System is indicated for the fixation, and fusion of rib fractures and osteotomies of normal and osteoporotic bone.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for BIOMET Microfixation. The word "BIOMET" is in large, bold, dark gray letters on the top line. Below that, in smaller, lighter gray letters, is the word "MICROFIXATION". A thin, light green line is below the word "MICROFIXATION". The words "One Surgeon. One Patient" are below the green line in a smaller font.
## 510(k) Summary
Prepared January 28, 2021
| Submitter: | Biomet Microfixation<br>1520 Tradeport Drive<br>Jacksonville, FL 32218 |
|------------|----------------------------------------------------------------------------------------------------------|
| Contact: | Lauren Jasper, Regulatory Affairs Manager<br>lauren.jasper@zimmerbiomet.com<br>Telephone: (904) 741-9259 |
Device Name: Biomet Microfixation RibFix Advantage System
| | Device Classification: |
|--|------------------------|
|--|------------------------|
| Product Code | Classification Name | Device<br>Classification | Regulation<br>Number | Regulation Description |
|---------------|-----------------------------|--------------------------|----------------------|------------------------------------------------------------------------------------|
| HRS (Primary) | Plate,<br>Fixation,<br>Bone | 2 | 888.3030 | Single/multiple component<br>metallic bone fixation<br>appliances and accessories |
| HWC | Screw,<br>Fixation,<br>Bone | 2 | 888.3040 | Smooth or threaded metallic<br>bone fixation fastener |
| HTN | Washer, Bolt<br>Nut | 2 | 888.3030 | Single/multiple component<br>metallic bone fixation<br>appliances and accessories. |
Indications for Use: The RibFix Advantage System is indicated for the fixation, stabilization, and fusion of rib fractures and osteotomies of normal and osteoporotic bone.
Contraindications: The RibFix Advantage system is contraindicated for:
- Use in patients with latent or active infection, sepsis, and/or device material sensitivity. 1.
- 2. Use in patients who are unwilling or incapable of following postoperative care instructions.
- 3. This device is not intended for locking post attachment or fixation to the clavicle or spine.
- 4. This device is not intended for highly comminuted fractures.
Device Description: The RibFix Advantage System consists of bridges (with locking posts) and locking caps for the thoracoscopic fixation and stabilization of ribs. These implants are manufactured from commercially pure titanium (per ASTM F67) and titanium alloys (Ti-6A1-7Nb
{4}------------------------------------------------
per ASTM F1295 and Ti-6Al-4V per ASTM F136 or ASTM F1472). The system is comprised of the following implants:
- . Titanium alloy RibFix Advantage Bridge with threaded locking posts
- . Commercially pure titanium locking caps
- . Commercially pure titanium washers (optional)
When fully assembled, the bridge is placed on the underside of the rib, the threaded locking posts extends through pre-drilled holes in the rib, and the locking caps are fixed to the locking post on the anterior side of the rib. Washers may be placed on the posts prior to assembling with locking caps. The combined threaded locking post/locking cap provides for fixation of the bridge. The devices are sold non-sterile and intended to be sterilized by the user prior to implantation. The system also includes reusable instrumentation to aid in implantation of the devices.
The design, materials, sterilization method, and indications for use are identical between the subject device and the predicate device.
## Substantial Equivalence:
Primary Predicate Device: K183317, AdvantageRib System
The similarities of the subject devices to the predicate devices are as follows:
- . The indications for use are identical to that of the predicate device.
- The design of the subject devices is equivalent to the predicate devices. ●
- The sterilization method (steam sterilization) of the subject devices is identical to the predicate devices.
- The materials of the subject devices are identical to the predicate devices. ●
The differences of the subject devices as compared to the predicate devices are as follows:
- . Update package insert to include MR conditional parameters and symbol
- Update outer package label to include MR conditional symbol
- Update cleaning instructions for the processing of reusable instruments and . new/uncompromised implants.
- . Add additional sterilization parameters applicable for other geographic locations.
- Revise the contraindications statement.
# Additional Reference Devices:
K162974. Biomet Microfixation RibFix Blu Thoracic Fixation System K163007, Biomet Microfixation SternaLock 360 Sternal Closure Fixation System
The similarities of the subject devices to these predicate devices are as follows:
- . The cleaning and sterilization instructions provided in the package insert are identical.
{5}------------------------------------------------
Non-Clinical Performance Data: The following nonclinical tests were submitted and relied on in this premarket notification submission for a determination of substantial equivalence. Testing identified in this summary has all passed acceptance criteria established by the predicate device where applicable.
MR Safety Evaluation following standards listed below:
- ASTM F2052-15, "Standard Test Method for Measurement of Magnetically Induced ● Displacement Force on Medical Devices in the Magnetic Resonance Environment"
- ASTM F2213-17, "Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment"
- ASTM F2182-11a, "Standard Test Method for Measurement of Radio Frequency Induced Heating Near Passive Implants During Magnetic Resonance Imaging"
- ASTM F2119-07 (Reapproved 2013), "Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants"
Clinical Performance Data: Clinical testing was not necessary for the determination of substantial equivalence.
Sterilization Information: There have been no changes to the sterilization method. The implants and instruments are provided non-sterile to be sterilized by steam at the end user facility.
Conclusion: The proposed device has identical indications for use as the predicate devices. The submission demonstrates that (1) any differences in technological characteristics of the predicates do not raise any new questions of safety and efficacy and (2) the proposed device is at least as safe and effective as the predicates. It is concluded that the information included in this summary supports substantial equivalence.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.