The PolyLock System is indicated for use in the treatment of bone fractures, arthrodesis, osteochondritis, and interfragmentary indications in the small bones of the hand and foot. Including, fusions of the forefoot and midfoot. stabilization of 1st, 2nd, 3rd, 4th, and 5th tarsometatarsal fusions, intercuneiform, talo-navicular, calcaneo-cuboid, and medial column fusions, first metatarsal osteotomies for hallux valgus/rigidus correction, first metatarsal fracture fixation, arthrodesis of the first metatarsal cuneiform joint, and arthrodesis fusion of the first metatarsophalangeal joint.
Device Story
PolyLock Small Bone Plating System; metallic bone fixation plates and screws for small bone orthopedic reconstruction, trauma, and osteotomies. System features low-profile, limited-contact plates with threaded holes for poly-axial and locking screws (2.0mm, 2.4mm, 2.7mm diameters; 8mm-26mm lengths). Instrumentation includes drill bits, countersinks, K-wires, reamers, and drivers. Used in clinical settings by surgeons to stabilize bone segments; facilitates manual/dynamic compression. Implants and specific instruments (drill bits, reamers, wires) are single-use. Provides mechanical stabilization to bone fragments to promote healing/fusion.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by non-clinical bench testing, including validation of cleaning, packaging, and sterilization, alongside engineering rationales for performance equivalence.
Technological Characteristics
Materials: Stainless steel (ASTM F899) and Ti-6Al-4V (ASTM F136). Design: Low-profile, limited-contact plates with poly-axial and locking screw holes. Connectivity: None (mechanical device). Sterilization: Not specified (implants/instruments).
Indications for Use
Indicated for patients requiring treatment of bone fractures, osteotomies, arthrodesis, osteochondritis, and interfragmentary fixation in small bones of the hand and foot, including forefoot/midfoot fusions and hallux valgus/rigidus correction.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
K202959 — PolyLock Plating System · Fusion Orthopedics, LLC · Nov 24, 2020
K211843 — PolyLock Plating System · Fusion Orthopedics · Mar 4, 2022
K181820 — Medline Unite Mini Plates and Screws · Medline Industries, Inc. · Nov 1, 2018
K163039 — ORTHOLOC® 3Di Small Bones Plating · Wrightmedicaltechnologyinc · Mar 13, 2017
K172441 — EPIC Extremity Fracture Plate System · Epic Extremity, LLC · Nov 6, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The first logo on the left is the Department of Health & Human Services logo. The second logo on the right is the FDA logo, which stands for U.S. Food & Drug Administration. The FDA logo is in blue.
November 18, 2019
Fusion Orthopedics, LLC Eli Jacobson Official Correspondent 4135 S. Power Rd. Suite 110 Mesa, Arizona 85212
Re: K192323
Trade/Device Name: PolyLock Small Bone Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: August 9, 2019 Received: August 27, 2019
Dear Eli Jacobson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
### K192323
Device Name PolyLock Small Bone Plating System
### Indications for Use (Describe)
The PolyLock System is indicated for use in the treatment of bone fractures, arthrodesis, osteochondritis, and interfragmentary indications in the small bones of the hand and foot. Including, fusions of the forefoot and midfoot. stabilization of 1st, 2nd, 3rd, 4th, and 5th tarsometatarsal fusions, intercuneiform, talo-navicular, calcaneo-cuboid, and medial column fusions, first metatarsal osteotomies for hallux valgus/rigidus correction, first metatarsal fracture fixation, arthrodesis of the first metatarsal cuneiform joint, and arthrodesis fusion of the first metatarsophalangeal joint.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond fo, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary: PolyLock Small Bone Plating System
| Date Prepared | August 9th, 2019 |
|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted By | Fusion Orthopedics, LLC<br>4135 S. Power Rd., Suite 110<br>Mesa, AZ 85212<br>800-403-6876 |
| Primary Contact | Eli Jacobson<br>4135 S. Power Rd., Suite 110<br>Mesa, AZ 85212<br>800-403-6876 Tele<br>e-mail: eli@fusionorthopedics.com |
| Trade Name | PolyLock Small Bone Plating System |
| Common Name | Plate, Fixation, Bone |
| Class | II |
| Product Code | HRS, HWC |
| Regulation | 21 CFR 888.3030 - Single/multiple component metallic bone fixation appliances and accessories |
| Device Panel | Orthopedic |
| Predicate Devices | M2 Modular Foot-Osteosynthesis System (K072740) [Primary Predicate]<br>Medline UNITE: Mini Foot Plating System (K181820)<br>Medline UNITE: Foot Recon Plating System (K151235)<br>Wright Medical ORTHOLOC 3Di Small Bone Plating System (K163039)<br>Wright Medical MaxLock Extreme (K113048)<br>DePuy Synthes LCP ForeFoot/MidFoot (K100776)<br>DePuy Synthes Locking Foot Module (K071264) |
| Reference<br>Device | FuzeFix Screw System, Fusion Orthopedics (K170038) |
| Device<br>Description | The PolyLock System consists of various shapes and sizes of plates for the management of<br>small bone orthopedic osteotomies, reconstruction, and trauma. Features include a low<br>profile, limited contact plate capable of dynamic/manual compression, with threaded screw<br>holes accepting both poly-axial and locking screws. The system also consists of poly-axial<br>and locking screws with diameters consisting of 2.0mm, 2.4mm and 2.7mm and lengths<br>from 8mm to 26mm. System instrumentation includes: drill bits, countersinks, K-wires, olive<br>wires, depth gauges, reamers, driver shafts, guides, plate benders, clamps, and driver<br>handles to facilitate the placement of the screws. The implants are intended for single use<br>only. Instruments designed for bone removal are intended for single use only, such as: drill<br>bits, countersinks, reamers, K-wires, and olive wires. |
| Indications for<br>Use | The PolyLock System is indicated for use in the treatment of bone fractures, osteotomies,<br>arthrodesis, osteochondritis, and interfragmentary indications in the small bones of the hand<br>and foot. Including, fusions of the forefoot and midfoot, stabilization of 1st, 2nd, 3rd, 4th, and<br>5th tarsometatarsal fusions, intercuneiform, navicular-cuneiform, talo-navicular, calcaneo-<br>cuboid, and medial column fusions, first metatarsal osteotomies for hallux valgus/rigidus |
| | correction, first metatarsal fracture fixation, arthrodesis of the first metatarsal cuneiform<br>joint, and arthrodesis/fusion of the first metatarsophalangeal joint. |
| Materials | Stainless steel (ASTM F899)<br>Ti-6AI-4V (ASTM F136) |
| Substantial<br>Equivalence<br>Claimed to<br>Predicate Devices | The PolyLock Small Bone Plating System is substantially equivalent to the predicate<br>devices in terms of intended use, design, materials used, mechanical safety, and<br>performances. |
| Non-clinical Test<br>Summary | Validations were performed on the cleaning, packaging and sterilization of the implants and<br>associated surgical instruments. Engineering rational was performed to show performance<br>equivalence. The results of the testing demonstrate that the device is substantially<br>equivalent to the predicate device identified above. |
| Clinical Test<br>Summary | No clinical studies were performed |
| Conclusions:<br>Non- clinical and<br>Clinical | Fusion Orthopedics LLC considers the PolyLock Small Bone Plating System to be<br>equivalent to the predicate devices listed above. This conclusion is based upon the devices'<br>similarities in principles of operation, technology, materials and indications for use. |
In accordance with 21 CFR 807.92 of the Federal Code of Regulations
{4}------------------------------------------------
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.