The Piccolo Composite foot and ankle plates are indicated for fixation of osteotomies, fusions, fractures, nonunions, malunions and replantations of small bone fragments in adult and adolescent (12 - 21 years) patients, including the foot and ankle, and including in osteopenic bone. The MTP plates are indicated for treatment of deformations, fractures, nonunions and replantations of the 1st metarsophalangeal joint and 1st metatarsal bone.
Device Story
Piccolo Composite Plate System consists of carbon fiber reinforced PEEK (CFR-PEEK) plates and titanium alloy screws; includes instruments for implantation. Plates feature tantalum thread for fluoroscopic visualization. Used by surgeons for bone fixation in foot and ankle procedures. System provides mechanical stabilization of bone fragments; supports healing of fractures, osteotomies, and fusions. Benefits include radiolucency of CFR-PEEK material compared to traditional metal plates, allowing improved imaging of bone healing.
Clinical Evidence
No clinical data. Bench testing only. Performance characteristics evaluated via static and dynamic bending tests per ASTM F 382 and bacterial endotoxin testing.
Technological Characteristics
Plates: Carbon fiber reinforced polyetheretherketone (CFR-PEEK) with tantalum thread for radiopacity. Screws: Titanium alloy (locking and non-locking 2.7mm, and cannulated lag screws). Mechanical testing performed per ASTM F 382.
Indications for Use
Indicated for fixation of osteotomies, fusions, fractures, nonunions, malunions, and replantations of small bones/fragments in adult and adolescent (12-21 years) patients, including foot/ankle and osteopenic bone. MTP plates indicated for 1st metatarsophalangeal joint and 1st metatarsal bone deformations, fractures, nonunions, and replantations.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
K102597 — PICCOLO COMPOSITE PLATE SYSTEM · N.M.B. Medical Applications, Ltd. · Feb 24, 2011
K182015 — Piccolo Composite® Plate System · CarboFix Orthopedics , Ltd. · Oct 11, 2018
K130061 — PICCOLO COMPOSITE PLATE SYSTEM · CarboFix Orthopedics , Ltd. · Feb 27, 2013
K160002 — Piccolo Composite(R) Distal Volar Radius Plate System · CarboFix Orthopedics , Ltd. · Mar 9, 2016
K143496 — Piccolo Composite Plate System · CarboFix Orthopedics , Ltd. · May 29, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, resembling a bird in flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 12, 2017
CarboFix Orthopedics, Ltd. Yael Rubin Director of Regulatory Affairs 11 Ha'hoshlim St. Herzeliya, 4672411 ISRAEL
Re: K170401
Trade/Device Name: Piccolo Composite® Plate System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: April 19, 2017 Received: April 21, 2017
Dear Yael Rubin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Vincent J. Devlin -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170401
Device Name Piccolo Composite® Plate System
Indications for Use (Describe)
The Piccolo Composite foot and ankle plates are indicated for fixation of osteotomies, fusions, fractures, nonunions, malunions and replantations of small bone fragments in adult and adolescent (12 - 21 years) patients, including the foot and ankle, and including in osteopenic bone.
The MTP plates are indicated for treatment of deformations, fractures, nonunions and replantations of the 1st metarsophalangeal joint and 1st metatarsal bone.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
# CarboFix Orthopedics Ltd. Piccolo Composite® MTP Plate System
### Applicant Name
CarboFix Orthopedics, Ltd. 11 Ha'hoshlim St., Herzeliya 4672411, Israel
#### Contact Person
Yael Rubin CarboFix Orthopedics, Ltd. 11 Ha'hoshlim St., Herzeliya 4672411, Israel Tel: +972 9 9511511, Fax: +972 9 9548939
# Date Prepared
May 2017
#### Trade/Proprietary Name
Piccolo Composite® Plate System
# Common Name
Bone Plating System
#### Classification Name
Single/multiple component metallic bone fixation appliances and accessories; (21 CFR §888.3030; Class II; Product Code HRS).
{4}------------------------------------------------
#### Predicate Devices
#### Primary Predicate:
- Piccolo Composite® Plate System (CarboFix Orthopedics Ltd.; K102597, K120409. K130061, 143496, 160002)
Additional Predicates:
- 트 2.4mm/2.7mm Variable Angle LCP Forefoot/Midfoot System – First MTP Fusion (Synthes; K100776)
- MaxLock Extreme MTP Plate Systems (Tornier (OrthoHelix Surgical Designs Inc.); K101962, K100618, and more)
#### Indications for Use
The Piccolo Composite foot and ankle plates are indicated for fixation of osteotomies, fusions, fractures, nonunions, malunions and replantations of small bones and small bone fragments in adult and adolescent (12 - 21 years) patients, including the foot and ankle, and including in osteopenic bone.
The MTP plates are indicated for treatment of deformations, fractures, nonunions and replantations of the 1st metatarso-phalangeal joint and 1st metatarsal bone.
#### System Description
The Piccolo Composite MTP Plate System comprises implants (plates and screws), and a set of instruments.
The plates are made of carbon fiber reinforced polyetheretherketone (CFR-PEEK), and are marked with a tantalum thread, to provide for their visualization under fluoroscopy. The screws are made of titanium alloy. Both non-locking and locking 2.7mm screws are available, in various lengths. Cannulated Lag Screws are also provided.
{5}------------------------------------------------
#### Substantial Equivalence
The Piccolo Composite MTP Plate System intended use, design, materials, technological characteristics, and principles of operation are substantially equivalent to those of the predicate devices, as applicable.
Performance characteristics included, among others, static and dynamic bending (evaluated per ASTM F 382), and are comparable to those of predicate devices (as applicable), thus demonstrating that the device is safe and effective for its intended use. In addition, bacterial endotoxin testing was conducted.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.