K131782 · Aap Implantate AG · HRS · Nov 18, 2013 · Orthopedic
Device Facts
Record ID
K131782
Device Name
LOQTEQ HIGH TIBIA OSTEOTOMY PLATE 4.5
Applicant
Aap Implantate AG
Product Code
HRS · Orthopedic
Decision Date
Nov 18, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Intended Use
LOQTEQ® High Tibial Osteotomy Plate 4.5: - Open-wedge osteotomies at the proximal medial tibia . - Treatment of bone and joint deformities . - Treatment of malpositions caused by injuries or disorders such as . osteoarthritis LOQTEQ® Distal Femur Osteotomy Plate 4.5: - Closed-wedge osteotomies of the medial distal femur . - Treatment of bone and joint deformities . - Treatment of malpositions caused by injuries or disorders such as . osteoarthritis
Device Story
System comprises metallic bone plates and screws for internal fixation of bone fragments following osteotomy of medial proximal tibia or medial distal femur. Surgeon implants device to stabilize bone; locking screws create internal fixator. Used in clinical/surgical settings. Provides mechanical stability to bone fragments to facilitate healing of deformities or malpositions.
Clinical Evidence
Bench testing only. Implant fatigue tests with progressive loadings performed to simulate worst-case clinical scenarios. Results demonstrate mechanical performance equivalent to predicate devices.
Technological Characteristics
Implants manufactured from Ti6Al4V (ASTM F136 or ISO 5832-3). System includes bone plates (High Tibial and Distal Femur) and cortical screws (self-tapping, T25). Mechanical fixation via locking screw technology. Non-active, metallic orthopedic implant system.
Indications for Use
Indicated for patients requiring open-wedge osteotomies of the proximal medial tibia or closed-wedge osteotomies of the medial distal femur to treat bone/joint deformities and malpositions resulting from injuries or disorders like osteoarthritis.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
K113601 — AAP LOQTEQ DISTAL MEDIAL TIBIA PLATE 3.5 SYSTEM · Aap Implantate AG · Nov 8, 2012
K133440 — AXSOS 3 TI LOCKING PLATE SYSTEM LINE EXTENSION · Stryker Trauma AG · Dec 18, 2013
K132554 — AAP LOQTEQ PROXIMAL MEDIAL TIBIA PLATE 3.5 · Aap Implantate AG · Oct 3, 2013
K230507 — Zimed Distal Medial Tibial Plate and Screw System · Zimed Medikal · Jan 4, 2024
K061098 — EBI OPTILOCK PERIARTICULAR PLATING SYSTEM · Ebi, L.P. · May 24, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
.
NOV 1 8 2013
.
# Summary of Safety and Effectiveness
| Sponsor: | aap Implantate AG<br>Lorenzweg 5<br>D-12099 Berlin Germany |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact: | Dr. Christian Zietsch<br>Phone:+49-30-750-19 -193<br>Fax: +49-30-750-19 - 111 |
| Date | November 08 / 2013 |
| Trade Name: | aap LOQTEQ® Osteotomy System |
| Common Name: | Osteotomy System |
| Classification Name and<br>Reference: | 21 CFR 888.3030 Single/multiple component metallic<br>bone fixation appliances and accessories - Class II |
| Product Code: | HRS |
| Predicate device: | Synthes TomoFix™ Osteotomy System under the pre-<br>market notification K023941 (Jan 23, 2003) and Synthes<br>(USA) TomoFiXTM Medial Distal Femur Plates under the<br>premarket notification K081353 (Jul 23, 2008). |
| Device Description: | The aap LOQTEQ® Osteotomy System consists of bone<br>plates and bone screws, to be implanted by a surgeon in<br>order to achieve an internal fixation of bone fragments typi-<br>cally after osteotomy of medial proximal tibia and medial<br>distal femur. If the plates are used in conjunction with lock-<br>ing screws, a so called internal fixator will be realized (in-<br>ternal fixation).<br><br>The aap LOQTEQ® Osteotomy System consists of:<br>LOQTEQ® LOQTEQ® High Tibial Osteotomy Plate 4.5LOQTEQ® LOQTEQ® Distal Femur Osteotomy Plate 4.5 (left and right)LOQTEQ® Cortical Screw 4.5, small head, T25, self-tappingCortical Screw 4.5, small head, self-tappingLOQTEQ® Cortical Screw 4.5, T25, self-tappingSet of Instruments aap LOQTEQ® Osteotomy System Set of Instruments |
| Material: | Implants are made of Ti6AI4V (ASTM F136 or ISO 5832-3) |
{1}------------------------------------------------
#### oop Implantate AG aap LOQTEQ® Osteotomy System
LOQTEQ® High Tibial Osteotomy Plate 4.5: Indications: Open-wedge osteotomies at the proximal medial 1 tibia Treatment of bone and joint deformities -Treatment of malpositions caused by injuries or disorders such as osteoarthritis LOQTEQ® Distal Femur Osteotomy Plate 4.5: Closed-wedge osteotomies of the medial distal femur Treatment of bone and joint deformities -Treatment of malpositions caused by injuries or disorders such as osteoarthritis Substantial Equivalence The Substantial Equivalence of the new device and the predicate device is based on similar intended use, design, functionality, components and materials in use. Documentation including mechanical testing to show the substantial equivalence and safety and effectiveness has been provided with this submission. Performance Data Non-clinical tests have been performed and show (Non-Clinical and / the effectiveness and safety of the device. or Clinical): Summary of Non-clinical tests: Type of test: Implant Fatigue tests with progressive loadings, representing worst case scenario with respect to clinical use. Assessment of test results: Substantial equivalence with respect to the mechanical performance of the aap system could be stated due to the test results gained. The subject device is safe and effective, and whose performance meets the requirements of its
> pre-defined acceptance criteria and intended uses. Documentation regarding the mechanical testing to show the substantial equivalence and safety and effectiveness
has been provided with this submission.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is in a bold, sans-serif font. The words are arranged on a single line and centered horizontally. The background is plain and white.
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the symbol. The text is in all caps and is smaller than the caduceus symbol. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 18, 2013
aap Implantate AG Dr. Christian Zietsch Manager, Regulatory Affairs Lorenzweg 5 12099 Berlin GERMANY
Re: K131782
Trade/Device Name: aap LOTEQ® Osteotomy System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS Dated: October 21, 2013 Received: October 23, 2013
Dear Dr. Zietsch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
#### Page 2 - Dr. Christian Zietsch
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Mark N.Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
aap Implantate AG aap LOQTEQ® Osteotomy System
# Indications for Use Statement
510(k) Number (if known): K131782
## Device Name: aap LOQTEQ® Osteotomy System
## Indications for Use:
The indication for use of the aap LOQTEQ® Osteotomy System are:
LOQTEQ® High Tibial Osteotomy Plate 4.5:
- Open-wedge osteotomies at the proximal medial tibia .
- Treatment of bone and joint deformities .
- Treatment of malpositions caused by injuries or disorders such as . osteoarthritis
LOQTEQ® Distal Femur Osteotomy Plate 4.5:
- Closed-wedge osteotomies of the medial distal femur .
- Treatment of bone and joint deformities .
- Treatment of malpositions caused by injuries or disorders such as . osteoarthritis
Prescription Use × (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page _1_ of _1_
# Elizabeth L. Frank -S
Division of Orthopedic Devices
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.