MAXLOCK EXTREME EXTREMITY PLATING SYSTEM WITH VARIABLE ANGLE TECHNOLOGY
Applicant
Orthohelix Surgical Designs, Inc.
Product Code
HRS · Orthopedic
Decision Date
Apr 20, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Intended Use
Variable Angle technology in MTP, TMT, MFT, MXL, DFX, and EDL plates are indicated for fractures, fusions and osteotomies for small bones in the hand, wrist, foot and ankle in both pediatric and adult patients. Variable Angle Technology in MXL and CLA plates are indicated for the fractures, fusions, and osteotomies of the clavicle and small bones in the hand, wrist, foot and ankle.
Device Story
Bone fixation system modification adding variable angle locking capability to existing plating systems. Consists of polymer ring mating with locking plate and specialized locking screw. Allows screw insertion at angles up to 15° in any direction while maintaining angular stability. Used by surgeons in clinical settings for orthopedic procedures. Provides stable fixation for small bone fractures, fusions, and osteotomies.
Clinical Evidence
Bench testing only. Dynamic and static mechanical testing performed to confirm substantial equivalence to predicate devices.
Technological Characteristics
Metallic bone fixation plates with integrated polymer ring for variable angle locking. Allows 15° angular adjustment. Compatible with existing OrthoHelix plate systems.
Indications for Use
Indicated for fractures, fusions, and osteotomies of small bones in hand, wrist, foot, and ankle, and clavicle. Applicable to both pediatric and adult patients.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
K111041 — MINI VARIABLE SYSTEM · Orthohelix Surgical Designs, Inc. · Jun 6, 2011
K242814 — Tandry Q-Locking Plate System · Microware Precision Co., Ltd. · Nov 14, 2024
K131186 — SYNTHES 2.7MM AND 3.5MM VARIABLE ANGLE LCP MIDFOOT/HINDFOOT SYSTEM · Synthes (USA) Products, LLC · Oct 8, 2013
K150099 — Depuy Synthes Variable Angle Locking Hand System (1.3 mm and 2.0 mm Plates and Screws) · Synthes (USA) Products, LLC · Mar 23, 2015
K152145 — OsteoMed ExtremiLOCK Wrist Plating System · Osteomed · Dec 21, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
# APR 2 0 2010
KI00618 (1)2)
### 510(K) Summary of Safety and Effectiveness MaxLock Extreme® Extremity Plating System with Variable AngleTechnology
Class II
Proprietary Name:
MaxLock Extreme Extremity Plating System with Variable Angle Technology
888.3030 - Single/multiple component metallic
HRS - Single/multiple component metallic bone
bone fixation appliances and accessories
fixation appliances and accessories
Common Name:
Bone Fixation Screws
Classification Name and Reference:
Regulatory Class:
Device Product Code:
Contact Information:
Derek Lewis Vice-President of Research and Development OrthoHelix Surgical Designs Inc. 1065 Medina Rd., Suite 500 Medina. OH 44256 Phone: 330 869-9562 Fax: 330 247-1598 E-Mail: dlewis@orthohelix.com
Summary Date:
April 15, 2010
#### Device Description
This special 510(K) submission is a modification to the previously cleared Modular Foot and Clavicle systems to add a variable angle locking screw construct. No changes have been made to the current locking plates; this addition will be compatible with all plates currently in the system. The OrthoHelix variable angle construct consists of a polymer ring which mates with the locking plate and allows for a specially designed locking screw to be inserted at angles up to 15° in any direction while maintaining angular stability.
{1}------------------------------------------------
#### Intended Use
Variable Angle technology in MTP, TMT, MFT, MXL, DFX, and EDL plates are indicated for fractures, fusions and osteotomies for small bones in the hand, wrist, foot and ankle in both pediatric and adult patients.
Variable Angle Technology in MXL and CLA plates are indicated for the fractures, fusions, and osteotomies of the clavicle and small bones in the hand, wrist, foot and ankle.
#### Substantial Equivalence
The MaxLock Extreme® Extremity Plating System with Variable Angle Technology is substantially equivalent to currently marketed devices, the OrthoHelix Modular Foot System (K073624) and Clavicle Plating System (K090289) and the Stryker VariAx System (K060613). The new technology is a modification to the OrthoHelix Modular Foot System, K073624 and Clavicle Plating System, K090289. The new technology only differs from the predicate devices with inclusion of a polymer locking ring and a specially designed locking screw to facilitate a variable angle locking mechanism.
Dynamic and static mechanical testing confirm that the MaxLock Extreme® implants with Variable Angle Technology are substantially equivalent to their predicate and meet the specified requirements for their intended use. No new issues of safety and efficacy have been raised during the development of this system.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
APR 2 0 2010
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OrthoHelix Surgical Designs, Inc. % Mr. Derek Lewis Vice President of Research and Development 1065 Medina Road, Suite 500 Medina, Ohio 44256
Re: K100618
Trade/Device Name: MaxLock Extreme® Extremity Plating System with Variable Angle Technology Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories
Regulatory Class: II Product Code: HRS Dated: April 5, 2010 Received: April 6, 2010
Dear Mr. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general-controls-provisions-of-the-Act-include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
#### Page 2 - Mr. Derek Lewis
CFR Part 807): labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Barbara buchner
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): ID KIOO6 18
Device Name: MaxLock Extreme Extremity Plating System with Variable Angle Technology
Indications for Use:
: , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , ,
Variable Angle technology in MTP, TMT, MFT, MXL, DFX, and EDL plates are indicated for fractures, fusions and osteotomies for small bones in the hand, wrist, foot and ankle in both pediatric and adult patients.
Variable Angle Technology in MXL and CLA plates are indicated for the fractures, fusions, and osteotomies of the clavicle and small bones in the hand, wrist, foot and ankle.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page of
Sonita fr mxm
(Division Sign-off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number _
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.