PERI-LOC PERIARTICULAR LOCKED PLATING SYSTEM VOLAR DISTAL RADIUS LOCKING PLATE FOR THE UPPER EXTREMITY-ASTM F139

K081106 · Smith & Nephew, Inc. · HRS · Jun 18, 2008 · Orthopedic

Device Facts

Record IDK081106
Device NamePERI-LOC PERIARTICULAR LOCKED PLATING SYSTEM VOLAR DISTAL RADIUS LOCKING PLATE FOR THE UPPER EXTREMITY-ASTM F139
ApplicantSmith & Nephew, Inc.
Product CodeHRS · Orthopedic
Decision DateJun 18, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The PERI-LOC Periarticular Locked Plating System can be used for adult and pediatric patients, as well as patients with osteopenic bone. PERI-LOC bones plates and screws are indicated for fixation of pelvic, small and long bone fractures, including those of the tibia, fibula, femur, pelvis, acetabulum, metacarpals, metatarsals, humerus, ulna, radius, calcaneus, and clavicle. Components in the PERI-LOC™ Periarticular Locked Plating System are for single use only.

Device Story

PERI-LOC™ Periarticular Locked Plating System consists of contoured locking bone plates and locking/non-locking bone screws; designed for surgical fixation of bone fractures. Device utilizes screw-to-plate locking mechanism to create fixed-angle construct; aids in maintaining fracture reduction. Intended for use by surgeons in clinical/OR settings. Benefits include stable fixation for various fracture types, including in osteopenic bone. Subject device represents material change to ASTM F139 stainless steel.

Clinical Evidence

No clinical data provided; substantial equivalence is based on design, material, and intended use similarities to previously cleared predicate devices.

Technological Characteristics

Metallic bone fixation system; components manufactured from ASTM F139 stainless steel. Features contoured locking plates and locking/non-locking screws. Employs screw-to-plate locking mechanism for fixed-angle stability. Single-use device.

Indications for Use

Indicated for fixation of pelvic, small, and long bone fractures (tibia, fibula, femur, pelvis, acetabulum, metacarpals, metatarsals, humerus, ulna, radius, calcaneus, clavicle) in adult and pediatric patients, including those with osteopenic bone.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary of Safety and Effectiveness ## PERI-LOC™ Periarticular Locked Plating System ## Volar Distal Radius Locking Plate for the Upper Extremity - ASTM F139 Material Change | Submitted By: | Smith & Nephew, Inc.,<br>Orthopaedic Division<br>1450 Brooks Road<br>Memphis, TN 38116 | JUN 1 8 2008 | |-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|--------------| | Date: | February 13, 2008 | | | Contact Person: | Laura Sejnowski, Regulatory Affairs Specialist<br>Tel: (901) 399-5349 Fax: (901) 399-1557 | | | Proprietary Name: | PERI-LOC™ Periarticular Locked Plating System -<br>Volar Distal Radius Locking Plate for the Upper Extremity<br>ASTM F139 Material Change | | | Common Name: | Bone Plates and Bone Screws | | | Classification Name and Reference: | 21 CFR 888.3030, single/multiple component metallic bone<br>fixation appliances and accessories - Class II | | | Device Product Code and Panel Code: | HRS, HWC / Orthopedics / 87 | | #### Device Description: Volar Distal Radius Locking Plates for the Upper Extremity manufactured from ASTM F139 material are line additions to the PERI-LOC™ Periarticular Locked Plating System cleared under K051735 and K061352. Like the predicate devices listed below, the subject components include various sizes of contoured locking bone plates and locking/non-locking bone screws made from stainless steel. PERI-LOC™ locking bone plates and locking bone screws incorporate a screw-to-plate locking feature which forms a locked, fixed angle construct to aid in holding fracture reduction. #### Intended Use: The PERI-LOC™ Periarticular Locked Plating System can be used for adult patients with osteopenic bone. PERI-LOC™ bone plates and screws are indicated for fixation of pelvic, small and long bone fractures, including those of the tibia, fibula, femur, pelvis, acetabulum, metacarpals, metatarsals, humerus, ulna, radius, and clavicle. #### Technological Characteristics: PERI-LOC™ Volar Distal Radius Plates manufactured from ASTM F139 material are similar to legally marketed devices listed below in that they share similar indications for use, are manufactured from similar materials, and incorporate similar technological characteristics. #### Substantial Equivalence Information: When compared to the predicate devices listed below, substantial equivalence is based on similarities in design features, overall indications for use, and material composition. - . PERI-LOC™ Periarticular Locked Plating System for the Upper Extremity - K051735 - . PERI-LOC™ Periarticular Locked Plating System for the Upper Extremity Device Modifications – K061352 - . Smith & Nephew Locked Plating System (PERI-LOC™ Periarticular Locked Plating System) – K033669 - . PERI-LOC™ Periarticular Locked Plating System VLP Plates/Screws – K071563 {1}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The seal is black and white and appears to be a scanned image. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Smith & Nephew, Inc. % Orthopaedic Division Ms. Laura Sejnowski Regulatory Affairs Specialist 1450 Brooks Road Memphis, TN 38116 JUN 1 8 2008 Re: K081106 Trade/Device Name: Peri-Loc™ Periarticular Locked Plating System - Volar Distal Radius Locking Plate for the Upper Extremity Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: May 15, 2008 Received: May 19, 2008 Dear Ms. Sejnowski: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {2}------------------------------------------------ Page 2 - Ms. Laura Sejnowski forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark M. Milliman Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Premarket Notification Indications for Use Statement 510(k) Number (if known): ___ K 0 8 || 0 6 PERI-LOC™ Periarticular Locked Plating System -Device Name: Volar Distal Radius Locking Plate for the Upper Extremity - ASTM F139 Material Change Indications for Use: The PERI-LOC Periarticular Locked Plating System can be used for adult and pediatric patients, as well as patients with osteopenic bone. PERI-LOC bones plates and screws are indicated for fixation of pelvic, small and long bone fractures, including those of the tibia, fibula, femur, pelvis, acetabulum, metacarpals, metatarsals, humerus, ulna, radius, calcaneus, and clavicle. Components in the PERI-LOC™ Periarticular Locked Plating System are for single use only. Prescription Use × AND/OR (Part 21 CFR 801.109) Over-the-Counter Use (Optional Format 1-2-96) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) MilfReddl fr mkm Division of General, Restorative, and Neurological Devices 510(k) Number K091106
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...