K061955 · Creative Medical Designs, Inc. · HRS · Aug 4, 2006 · Orthopedic
Device Facts
Record ID
K061955
Device Name
RAYHACK LOW PROFILE LOCKING PLATE AND INSTRUMENTS
Applicant
Creative Medical Designs, Inc.
Product Code
HRS · Orthopedic
Decision Date
Aug 4, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Rayhack Low Profile Locking Plate is a metallic surgical implant intended for long bone fixation utilized to assist healing but not intended to replace normal body structures. The plate and screws which attach to the bone are temporary internal fixation devices which align the bone surfaces in order to permit bone healing.
Device Story
Rayhack Low Profile Locking Plate is a metallic surgical implant for long bone fixation. Device consists of a stainless steel plate, two 2.7mm locking screws, four 3.5mm standard cortical screws, and one 2.7mm standard interfragmentary cortical screw. Used by surgeons in clinical settings to align bone surfaces to facilitate healing. Acts as a temporary internal fixation device; does not replace normal body structures. Instrumentation provided to assist with surgical procedure. Benefits include stabilization of bone fractures to permit natural healing.
Clinical Evidence
Bench testing only. Bone model testing compared holding strength and stiffness of the subject plate against the predicate plate, demonstrating substantially equivalent performance.
Technological Characteristics
Metallic bone fixation plate and screw system. Material: stainless steel. Fixation: locking and standard cortical screws. Form factor: low-profile plate. Non-powered, mechanical device.
Indications for Use
Indicated for long bone fixation to assist healing in patients requiring temporary internal bone surface alignment.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
K071563 — PERI-LOC PERIARTICULAR LOCKED PLATING SYSTEM · Smith & Nephew, Inc. · Aug 8, 2007
K211324 — 2.4mm Distal Radius System · Auxein Medical Private Limited · Mar 14, 2022
K141121 — AXSOS 3 TI LOCKING PLATE SYSTEM · Stryker Trauma AG · Jun 20, 2014
K111253 — ARTHREX DISTAL EXTREMITY PLATE SYSTEM · Arthrex, Inc. · Aug 2, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
# K061955 510(K) SUMMARY
#### 1. SUBMITTER:
AUG - 4 2006
Creative Medical Designs, Inc. 13914 Shady Shores Drive Tampa, FL 33613 813 - 875-9999 (phone) 813-961-5543 (fax)
Contact: Jane Rayhack, CEO Date Prepared: July 6, 2006
#### 2. DEVICE:
Trade Name: Rayhack Low Profile Locking Plate and Instruments Common Name: Bone Fixation Plate Class: II Classification Name: Single/multiple component bone fixation appliances and accessories Fastener, Fixation, non-degradable soft tissue
#### 3. PREDICATE DEVICE:
Rayhack Osteotomy System (K952766)
#### 4. DEVICE DESCRIPTION:
The Rayhacks Low Profile Locking Plate and Instruments is an implant intended for fixation of long bones to assist healing. The implant is fabricated from stainless steel. It is fixated with two 2.7mm locking screws, four 3.5mm standard cortical screws and one 2.7 standard interfragmentary cortical screw. In addition to the plate and screws, there are instruments intended to assist with the procedure.
#### 5. INTENDED USE:
The Rayhack" Low Profile Locking Plate is a metallic surgical implant intended for long bone fixation utilized to assist healing but not intended to replace normal body structures. The plate and screws which attach to the bone are temporary internal fixation devices which align the bone surfaces in order to permit bone healing.
#### 6. COMPARISON OF CHARACTERISTICS:
The proposed Rayhack Low Profile Locking Plate is fabricated from identical materials as the predicate device. In addition, the plate and instrumentation is
{1}------------------------------------------------
K061955 (pg 2 of 2)
essentially identical to the predicate except it allows for the use of locking screws and the plate is a lower profile. Instrumentation has been modified to work with the revised plate.
### 7. PERFORMANCE DATA:
The following performance data was proved in support of the substantial equivalence determination:
Bone Model Testing: The holding strength and stiffness characteristics of the new plate compared to the predicate plate.
The testing demonstrates substantially equivalent performance between the two devices
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird with three horizontal lines forming its wings. The bird is positioned above a circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES UNITED STATES OF AMERICA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 4 2006
Creative Medical Design, Inc. % Ms. Jane Rayhack CEO 13914 Shady Shores Drive Tampa, Florida 33613
Re: K061955
Trade/Device Name: Rayhack® Low Profile Locking Plate Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS Dated: July 10, 2006 Received: July 11, 2006
Dear Ms. Rayhack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Jane Rayhack
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Herbert Semmler
for
E
I
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): _ ¥06 | 9 5 5
Device Name: Rayhack 6 Low Profile Locking Plate
Indications for Use:
The Rayhack Low Profile Locking Plate is a metallic surgical implant intended for long bone fixation utilized to assist healing but not intended to replace normal body structures. The plate and screws which attach to the bone are temporary internal fixation devices which align the bone surfaces in order to permit bone healing.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Helut Leun
(Division Sign-Off) Page of Division of General, Restorative, and Neurological Devices
**510(k) Number** Ks61955
(Posted November 13, 2003)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.