K060473 · New Deal, S.A. · HRS · May 17, 2006 · Orthopedic
Device Facts
Record ID
K060473
Device Name
TTC PLATE
Applicant
New Deal, S.A.
Product Code
HRS · Orthopedic
Decision Date
May 17, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The NEWDEAL® TTC Plates are intended for use in arthrodesis of the ankle joint and distal tibia, fractures, osteotomies, fusions and replantations of small bones including the foot and ankle.
Device Story
TTC Plates are metallic bone fixation devices used for orthopedic procedures in the foot and ankle. The system consists of a plate, fixation screws, and washers. The device utilizes a locking mechanism where washers are drilled into threaded lipped sockets at the top of each plate hole to block the screw head, creating a single integrated implant/screw unit fixed into the bone. The device is intended for use by surgeons in clinical settings to stabilize bone segments during healing. It provides mechanical fixation to support arthrodesis, fracture repair, and osteotomies, potentially improving patient outcomes by ensuring stable bone alignment.
Clinical Evidence
Bench testing only. Performance tests demonstrated mechanical properties compatible with predicate devices and intended use.
Technological Characteristics
Metallic bone fixation plate system. Includes plates, fixation screws, and locking washers. Features threaded lipped sockets for screw head locking. Operates as a mechanical fixation device. No software or electronic components.
Indications for Use
Indicated for patients requiring arthrodesis of the ankle joint and distal tibia, or treatment of fractures, osteotomies, fusions, and replantations of small bones in the foot and ankle.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Synthes Modular Foot System - 2.7 mm Module (K010321)
Related Devices
K070160 — MODIFICATION TO TTC PLATES · New Deal, S.A. · Feb 8, 2007
K060476 — LAPIDUS PLATE · New Deal, S.A. · May 17, 2006
K172973 — Life Spine Foot and Ankle Plating System · Life Spine, Inc. · Dec 21, 2017
K090139 — ANKLE ARTHRODESIS PLATE SYSTEM · Tornier, Inc. · Oct 27, 2009
K230591 — TriLEAP Plating System · Crossroads Extremity Systems · Sep 29, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
MAY 17 2008
### 510(K) SUMMARY
# TTC Plates
Submitter's name and address: Newdeal SAS 10, place d'Helvétie
69006 Lyon, France Tel: +33 4 37 47 51 51 Fax: +33 4 37 47 51 52
# Contact person and telephone number
Morgane Grenier Regulatory and Clinical Affairs Director Newdeal SAS 10, place d'Helvétie 69006 Lyon, France Tel: +33 4 37 47 51 51 Fax: +33 4 37 47 51 52
#### Alternate Contacts
#### Authorized Agent in the United States
Judith E. O'Grady, RN, MSN Sr. Vice President, Regulatory Affairs, Quality Assurance and Clinical Affairs Integra LifeSciences Corporation 311 Enterprise Drive Plainsboro, NJ 08536, USA Tel: (609) 936-2311 Fax: (609) 275-9445 E-mail: jogrady@integra-ls.com
#### Date Summary was prepared:
May 11, 2006
#### Name of the device:
| Proprietary Name: | TTC Plates |
|-----------------------|-------------------------------------------------------------------------------------------------|
| Common Name: | Plate, Fixation, Bone |
| Classification Name: | Single/multiple component metallic bone fixation appliances and<br>accessories (21CFR 888.3030) |
| Device Product Code: | HRS |
| Classification Panel: | Orthopedic |
#### Substantial Equivalence:
The TTC Plate is substantially equivalent to the Synthes Ankle Arthrodesis Plates, K022255 and the Synthes Modular Foot System - 2.7 mm Module, K010321.
#### Device Description:
The NEWDEAL® TTC Plates consists of a plate, available in different sizes, and implanted using NEWDEAL® locking system fixation screws and washers.
{1}------------------------------------------------
The NEWDEAL® locking system includes as many fixation screws as there are threaded lipped sockets on the plate and as many washers as implanted screws.
The NEWDEAL® locking system creates a single implant/screw unit fixed into the bone. The osteosynthesis screws must be driven into the bone through the holes in the plate. The system is locked by means of washers drilled into the threaded lipped socket at the top of each hole, thus blocking each screw head.
#### Intended Use:
The NEWDEAL® TTC Plates are intended for use in arthrodesis of the ankle joint and distal tibia, fractures, osteotomies, fusions and replantations of small bones including the foot and ankle.
# Testing and Test Results:
The results of performance tests demonstrate that the TTC Plates have mechanical properties compatible with the predicate devices and intended use.
#### Conclusion
The Newdeal TTC Plates are substantially equivalent to commercially marketed devices, the Synthes Ankle Arthrodesis Plate, K022255 and the Synthes Modular Foot System - 2.7 mm Module, K010321.
The Newdeal TTC Plates do not raise any new issues of scientific technology, safety or effectiveness.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a seal for the Department of Health and Human Services (HHS) in the United States. The seal features the HHS logo, which is a stylized representation of three human figures. The logo is surrounded by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
TAY 17 2006
Newdeal SA c/o Ms. Judith O'Grady, R.N., M.S.N. Sr. VP. Regulatory Affairs Integra Lifesciences Corporation 311 Enterprise Drive Plainsboro, New Jersey 08536
Re: K060473
Trade/Device Name: TTC Plate Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS Dated: May 11, 2006 Received: May 12, 2006
Dear Ms. O'Grady:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 - Ms. Judith O'Grady, R.N., M.S.N.
comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Herbert Hoover hp
Mark N. Melkerson, M.S. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
# 510(k) Number (if known): K060473
Device Name: TTC Plates
# Indications For Use:
The NEWDEAL® TTC Plates are intended for arthrodesis of the ankle joint and distal tibia, fractures, osteotomies, fusions and replantations of small bones in the floor and ankle.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hubert Lemerno
Page 1 of 1
sion Sign-Division of General, Restorative, and Neurological Devices
510(k) Number K060473
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.