ArthrexVIP Web Portal

K193523 · Arthrex, Inc. · LLZ · Apr 30, 2020 · Radiology

Device Facts

Record IDK193523
Device NameArthrexVIP Web Portal
ApplicantArthrex, Inc.
Product CodeLLZ · Radiology
Decision DateApr 30, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

The Arthrex VIP Web Portal is intended for use as a software interface of imaging information from a medical scamer such as a CT scanner. It is also intended as software for displaying/editing implant placement and surgical treatment options that were generated in the OrthoVis Desktop Software by trained Arthrex technicians. The Arthrex VIP Web Portal is intended for use with the Arthrex Glenoid Intelligent Reusable Instrument System (Arthrex Glenoid IRIS) and with the Arthrex OrthoVis Preoperative Plan. It is indicated for use with the following glenoid implant lines: Arthrex Univers II and Arthrex Univers Apex, Keeled or Pegged Glenoid components, the Vault Lock Glenoid Component, as well as the Univers Revers and Modular Glenoid System (MGS) Baseplate components.

Device Story

Web-based software portal for preoperative planning of total shoulder joint replacement; facilitates image upload from CT scanners; enables surgeon review/modification of implant placement plans generated by Arthrex technicians using OrthoVis Desktop Software. Surgeon logs into portal to visualize implant orientation, glenoid bone preparation instruments, and screw trajectories; approves or modifies plans. Approved plans used by Arthrex technicians to produce patient-specific instrumentation (Arthrex Glenoid IRIS). Supports multiple glenoid implant lines including Univers II, Univers Apex, Vault Lock, and Univers Revers/MGS. Streamlines communication between surgeon and planning facility; improves surgical planning accuracy for shoulder arthroplasty.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated through non-clinical software verification and validation, including regression testing, unit testing, code reviews, integration testing, and dimensional verification.

Technological Characteristics

Web-based software interface; facilitates image transfer and preoperative planning visualization. Features include 3D visualization of glenoid implants, bone preparation instruments, and screw trajectories. Operates as a client-server system for surgeon-technician collaboration. No hardware components; software-only device.

Indications for Use

Indicated for patients requiring total shoulder joint replacement. Used by surgeons to review, edit, and approve preoperative plans for glenoid implant placement, including Univers II, Univers Apex, Vault Lock, Univers Revers, and Modular Glenoid System (MGS) components.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. April 30, 2020 Arthrex Inc. Heli Chambi Infantas Sr. Regulatory Affairs Associate 1370 Creekside Blvd NAPLES, FL 34108 Re: K193523 Trade/Device Name: ArthrexVIP Web Portal Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: March 28, 2020 Received: March 31, 2020 Dear Heli Chambi Infantas: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Michael C. Owens Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) #### K193523 Device Name ArthrexVIP Web Portal #### Indications for Use (Describe) The Arthrex VIP Web Portal is intended for use as a software interface of imaging information from a medical scamer such as a CT scanner. It is also intended as software for displaying/editing implant placement and surgical treatment options that were generated in the OrthoVis Desktop Software by trained Arthrex technicians. The Arthrex VIP Web Portal is intended for use with the Arthrex Glenoid Intelligent Reusable Instrument System (Arthrex Glenoid IRIS) and with the Arthrex OrthoVis Preoperative Plan. It is indicated for use with the following glenoid implant lines: Arthrex Univers II and Arthrex Univers Apex, Keeled or Pegged Glenoid components, the Vault Lock Glenoid Component, as well as the Univers Revers and Modular Glenoid System (MGS) Baseplate components. | <span style="font-size: 10pt;"> <input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------| |-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Type of Use (Select one or both, as applicable) {3}------------------------------------------------ # K193523 | | 510(k) Summary | |--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------| | 510(k) SPONSOR / MANUFACTURER: | Arthrex Inc.<br>1370 Creekside Blvd<br>Naples, FL 34108 | | CONTACT PERSON: | Heli F. Chambi Infantas<br>Sr. Regulatory Affairs Associate<br>1-239-598-4302, ext. 73187<br>Heli.Chambiinfantas@arthrex.com | | TRADE NAME: | ArthrexVIP Web Portal | | DATE PREPARED: | 16-Dec-2019 | | COMMON NAMES: | Image processing system and preoperative software<br>for simulating / evaluating implant placement and<br>surgical treatment options | | Product | Product<br>Code | Regulation and<br>Classification Name | Device<br>Class | |--------------------------|-----------------|-------------------------------------------------------------------|-----------------| | ArthrexVIP Web<br>Portal | LLZ | 21 CFR 892.2050<br>Picture Archiving and<br>Communications System | II | # Predicate Devices: K162697: ArthrexVIP Web Portal # Device Description: The ArthrexVIP Web Portal is composed of software intended for use to facilitate upload of medical images, preoperative planning, and plan approval of placement and orientation of total shoulder joint replacement components. Each surgeon user's uploaded images are associated with specific cases and associated with that surgeon's profile. Uploaded images can be downloaded from the portal by Arthrex technicians and used to create preoperative plans (see 510(k) K151568) in the OrthoVis Desktop Software. The surgeon user is then able to login to the ArthrexVIP Web Portal to review the preoperative plan and either approve or modify the location and/or orientation of the joint replacement component. The approved plan is then downloaded by Arthrex technicians for production (see 510(k) K151500 and K151568) as part of the Arthrex Glenoid IRIS device. {4}------------------------------------------------ #### Intended Use and Indications: The ArthrexVIP Web Portal is intended for use as a software interface and for the transfer of imaging information from a medical scanner such as a CT scanner. It is also intended as software for displaying/editing implant placement and surgical treatment options that were generated in the OrthoVis Desktop Software by trained Arthrex technicians. The ArthrexVIP Web Portal is intended for use with the Arthrex Glenoid Intelligent Reusable Instrument System (Arthrex Glenoid IRIS) and with the Arthrex OrthoVis Preoperative Plan. It is indicated for use with the following glenoid implant lines: Arthrex Univers II and Arthrex Univers Apex, Keeled or Pegged Glenoid components, the Vault Lock Glenoid Component, as well as the Univers Revers and Modular Glenoid System (MGS) Baseplate components. #### Basis of Substantial Equivalence: The modified ArthrexVIP Web Portal has the same intended use and indications for use as the predicate ArthrexVIP Web Portal (K162697) with the exception of "COS technicians" now being "Arthrex technicians" to reflect the change in manufacturer from Custom Orthopaedic Solutions (COS) to Arthrex and the addition of the Modular Glenoid System baseplate to the indicated glenoid implant lines. The subject device is a version of the predicate device with changes to the following: - Addition of backside seating feature - Addition of the Univers Revers Modular Glenoid System (MGS) implant models - Addition of the Univers Revers Universal Glenoid (UG) implant models - Addition of the glenosphere and inlay implant models compatible with MGS and UG - Addition of glenoid bone preparation instrument models for visualization - Addition of the Univers Revers screw trajectory models for visualization # Non-Clinical Testing The following testing was performed to demonstrate substantial equivalency of the ArthrexVIP Web Portal to the OrthoVis Web Portal: - . Software verification and validation - Regression testing - Unit testing - Code reviews and checks - Integration testing - Dimensional verification # Clinical Testing Clinical testing was not necessary to determine substantial equivalence to the predicate.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...