The Boston Sight PD Prosthetic Device for daily wear is indicated for the management of a distorted corneal surface that: 1. precludes satisfactory spectacle lens correction 2. demonstrates significant improved rigid contact lens corrected vision 3. is incapable of wearing traditional corneal lenses because of the inability to achieve adequate lens centration/stability and/or tolerance to physical contact with a lens. Causes of corneal distortion include corneal degeneration (e.g. keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration), corneal dystrophy (e.g. lattice dystrophy, Reis-Bucklers dystrophy), and scarring from surgery (e.g. corneal transplant, LASIK, radial keratotomy), infection, or trauma. The BostonSight PD Prosthetic Device for daily wear is also indicated for therapeutic use in eyes with ocular surface disease from dry eye (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, limbal stem cell deficiency (e.g.Stevens-Johnson syndrome, chemical and thermal burns, radiation), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for distorted cornea or ocular surface disease, the BostonSight PD Prosthetic Device may incidentally provide correction of refractive error. The BostonSight PD Prosthetic Device may be cleaned and disinfected using a chemical (not heat) care system.
Device Story
BostonSight PD is a daily-wear, lathe-cut, rigid gas permeable (RGP) prosthetic contact lens. Device vaults over cornea, resting on scleral conjunctiva; creates tear reservoir between lens and cornea. Reservoir masks optical distortions; protects ocular surface from adverse environments (e.g., dysfunctional eyelids). Prescribed by eye care professionals for patients with corneal distortion or ocular surface disease. Lens provides physical protection and potential refractive correction. Patient self-administers daily; requires chemical (non-heat) cleaning/disinfection. Benefits include improved vision, ocular surface protection, and symptom relief for chronic conditions.
Clinical Evidence
No new clinical data provided. Safety and effectiveness supported by reference to predicate devices (K153066) and previously cleared RGP contact lenses (K033594, K022128, K071266) utilizing the same fluoro-silicone acrylate materials.
Technological Characteristics
Lathe-cut RGP lens; materials: roflufocon D, roflufocon E, oprifocon A, hexafocon B (Group 3 Fluoro Silicone Acrylate). Water content <1%. Diameter 18.0-26.0 mm. UV blocking >95-98%. Chemical disinfection only.
Indications for Use
Indicated for patients with distorted corneal surfaces (degeneration, dystrophy, post-surgical scarring, infection, trauma) or ocular surface disease (dry eye, limbal stem cell deficiency, skin disorders, neurotrophic keratitis, corneal exposure) requiring expanded tear reservoir and environmental protection. Suitable for patients unable to wear traditional corneal lenses or achieve satisfactory spectacle correction.
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
Boston EQUALENS II (oprifocon A) Rigid Gas Permeable Contact Lenses (K022128)
Boston XO2 (hexafocon B) Daily Wear Contact Lens (K071266)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping to create a sense of depth and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 25, 2016
BostonSight % Mr. Bret Andre Principal Consultant EyeReg Consulting, Inc. 6119 Canter Ln. West Linn, OR 97068
Re: K161461
Trade/Device Name: BostonSight PD Prosthetic Device Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid Gas Permeable Contact Lens Regulatory Class: Class II Product Code: HOD Dated: May 20, 2016 Received: May 26, 2016
Dear Mr. Andre:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K161461
Device Name BostonSight PD Prosthetic Device
Indications for Use (Describe)
The Boston Sight PD Prosthetic Device for daily wear is indicated for the management of a distorted corneal surface that:
- 1. precludes satisfactory spectacle lens correction
- 2. demonstrates significant improved rigid contact lens corrected vision
3. is incapable of wearing traditional corneal lenses because of the inability to achieve adequate lens centration/stability and/or tolerance to physical contact with a lens
Causes of corneal distortion include corneal degeneration (e.g. keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration), corneal dystrophy (e.g. lattice dystrophy, Reis-Bucklers dystrophy), and scarring from surgery (e.g. corneal transplant, LASIK, radial keratotomy), infection, or trauma.
The BostonSight PD Prosthetic Device for daily wear is also indicated for therapeutic use in eyes with ocular surface disease from dry eye (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, limbal stem cell deficiency (e.g.Stevens-Johnson syndrome, chemical and thermal burns, radiation), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for distorted cornea or ocular surface disease, the BostonSight PD Prosthetic Device may incidentally provide correction of refractive error.
The BostonSight PD Prosthetic Device may be cleaned and disinfected using a chemical (not heat) care system.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg class="bi bi-check-square-fill" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2zm10.03 4.97a.75.75 0 0 0-1.08.022L7.477 9.417 5.384 7.323a.75.75 0 0 0-1.06 1.06L6.97 11.03a.75.75 0 0 0 1.079-.02l3.992-4.99a.75.75 0 0 0-.01-.052z" fill-rule="evenodd"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
| <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K161461 Applicant information: July 22nd, 2016 Date Prepared: Name: BostonSight Address 464 Hillside Avenue, Suite 205 Needham, MA 02494 Contact Person: Gene P. Guselli President and CEO Phone number: 781-726-7337 EyeReg Consulting, Inc. Consultant/Correspondent: Bret Andre Phone number (503) 372-5226 Fax number (503) 419-4475
## Device Information:
| Device Classification: | Class II |
|------------------------|-----------------------------------------|
| Classification Number: | HQD |
| Classification Name: | Lenses, Rigid Gas Permeable, Daily Wear |
| Trade Name: | BostonSight PD Prosthetic Device |
{5}------------------------------------------------
# Predicate Devices:
The BostonSight PD Prosthetic Device is substantially equivalent to the following predicate device(s)
- "BostonSight IC Corneal & Scleral Lens" . - Primary Predicate Manufactured by Boston Foundation for Sight 510(k) number; K153066
- "Boston" Scleral (itafluorofocon B) RGP Contact Lens for Daily Wear" . – Reference Predicate Manufactured by Boston Foundation for Sight PMA number: P860022/S40
# Device Description:
The BostonSight PD Prosthetic Device is a daily wear, prosthetic device for the ocular surface lathe cut from one of the following fluoro-silicone acrylate materials:
- · roflufocon D (supplied by Contamac Ltd.)
- · roflufocon E (supplied by Contamac Ltd.)
- · oprifocon A (supplied by Bausch & Lomb, Inc.)
- · hexafocon B (supplied by Bausch & Lomb, Inc.)
The BostonSight PD Prosthetic Device is designed to vault over the cornea and rest on the conjunctiva overlying the sclera, resulting in a tear reservoir between the back surface of the prosthetic device and the corneal surface. The tear reservoir masks optical distortions from an irregular corneal surface, and in combination with the device itself, protects the ocular surface from an adverse external environment, including but not limited to dysfunctional evelids and margins. The design parameters are customized to allow for tear exchange underneath the device.
The BostonSight PD Prosthetic Device may be shipped "dry" or "wet" in a polypropylene contact lens case. The primary container for shipping the BostonSight PD Prosthetic Device is the Bausch & Lomb Frequent Replacement Contact Lens Case, with clearance under 510(k) K896685. When shipped "wet". The BostonSight PD Prosthetic Device manufactured from material supplied by Bausch & Lomb, inc. may be packaged and shipped in Boston Advance Comfort Formula Conditioning Solution (K974466) or Boston SIMPLUS Multi-Action solution (K024289). The BostonSight PD Prosthetic Device manufactured from material supplied by Contamac, Ltd. may be packaged and shipped "wet" in in the OPTIMUM by Lobob Cleaning and Disinfecting Storage solution, with clearance under 510(k) K014162.
{6}------------------------------------------------
| | ROFLUFOCON D | ROFLUFOCON E | OPRIFOCON A | HEXAFOCON B |
|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|------------------------------------------------------------------------|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Refractive Index | 1.4333 | 1.4332 | 1.4230 | 1.4240 |
| Light Transmission<br>(clear) | >97% | >97% | >95% | >95% |
| Light Transmission<br>(tinted) | >90% | >90% | >90% | >83% |
| Water Content | <1% | <1% | <1% | <1% |
| Dynamic Contact<br>Angle (Receding) | 3° | 6° | 56° | 40° |
| Specific Gravity | 1.166 | 1.155 | 1.24 | 1.19 |
| Modulus | 697 MPa | 77 MPa | 1300 MPa | 1160 MPa |
| Shore D Hardness | 75 | 77 | 81 | 78 |
| Oxygen<br>Permeability<br>(Dk)<br>ISO/FATT Method | $100 x 10^{-11}$<br>(cm2/sec) (ml O2/ml x<br>mm Hg @ 35°C) | $125 x 10^{-11}$<br>(cm2/sec) (ml O2/ml x<br>mm Hg @ 35°C) | $85 x 10^{-11}$ (cm2/sec)<br>(ml O2/ml x mm Hg @<br>35°C) | $141 x 10^{-11}$<br>(cm2/sec) (ml O2/ml x<br>mm Hg @ 35°C) |
| contain one or more<br>of the following<br>color additives<br>conforming to:<br>21 CFR Part 73 &<br>74, Subpart D | D & C Green No.<br>6, FD & C Red<br>No. 17,<br>CI Solvent<br>Yellow 18 | D & C Green No.<br>6, FD & C Red<br>No. 17,<br>CI Solvent<br>Yellow 18 | D&C Green No.6<br>and D&C Yellow<br>No.10 | D&C Green No.<br>6; C.I. Solvent<br>Yellow No. 18;<br>D&C Violet No.<br>2; D&C Red No.<br>17; C.I. Solvent<br>Yellow No.18 |
| UV Light Blocking<br>(UVB - 280nm –<br>315nm; UVA 316nm<br>- 380nm) | >98% UVB<br>>95% UVA | >98% UVB<br>>95% UVA | >95% UVB<br>>97% UVA | >95% UVB<br>>97% UVA |
The physical properties of the BostonSight PD Prosthetic Device manufactured from the various materials are as follows:
The parameters for the BostonSight PD Prosthetic Device are as follows:
- ન્ન્સ Chord Diameter:
- 米 Center Thickness:
- 米 Base Curve:
- 米 Spherical Powers:
18.0 mm to 26.0 mm 0.05mm to 0.60 mm 5.0 mm to 9.0 mm -25.00 Diopters to +35.00 Diopters
# Indication for Use:
The Boston Sight PD Prosthetic Device for daily wear is indicated for therapeutic use for the management of a distorted corneal surface that:
- 1. precludes satisfactory spectacle lens correction
- 2. demonstrates significant improved rigid contact lens corrected vision
- 3. is incapable of wearing traditional corneal lenses because of the inability to achieve adequate lens centration/stability and/or tolerance to physical contact with a lens
Causes of corneal distortion include corneal degeneration (e.g. keratoconus, keratoglobus, pellucid marginal degeneration, Salzmann's nodular degeneration), corneal dystrophy (e.g, lattice dystrophy, Reis-Bucklers dystrophy), and scarring from surgery (e.g. corneal transplant, LASIK, radial keratotomy), infection, or trauma.
{7}------------------------------------------------
The BostonSight PD Prosthetic Device for daily wear is also indicated for therapeutic use in eyes with ocular surface disease from dry eye (e.g. ocular Graft-versus-Host disease, Sjögren's syndrome, dry eye syndrome), limbal stem cell deficiency (e.g.Stevens-Johnson syndrome, chemical and thermal burns, radiation), disorders of the skin (e.g. atopy, ectodermal dysplasia), neurotrophic keratitis (e.g. Herpes simplex, Herpes zoster, Familial Dysautonomia), and corneal exposure (e.g. anatomic, paralytic) that might benefit from the presence of an expanded tear reservoir and protection against an adverse environment. When prescribed for therapeutic use for distorted cornea or ocular surface disease, the BostonSight PD Prosthetic Device may incidentally provide correction of refractive error.
The BostonSight PD Prosthetic Device may be cleaned and disinfected using a chemical (not heat) care system.
# Description of Safety:
- ~ Non-Clinical Studies ~
Non-clinical testing to demonstrate the safety and effectiveness of contact lenses manufactured from roflufocon D, roflufocon E, oprifocon A, and hexafocon B has been addressed by reference to the predicate device, BostonSight IC Corneal & Scleral Lens (K153066).
~ Clinical Studies ~
Clinical testing to demonstrate the safety and effectiveness of contact lenses manufactured from roflufocon D, roflufocon E, oprifocon A, and hexafocon B has been addressed by reference to the following:
- OPTIMUM GP(roflufocon D, roflufocon E) Daily Wear Contact Lenses K033594 ●
- . Boston EQUALENS II (oprifocon A) Rigid Gas Permeable Contact Lenses - K022128
- . Boston XO2 (hexafocon B) Daily Wear Contact Lens - K071266
# Substantial Equivalence:
The BostonSight PD Prosthetic Device is substantially equivalent to the Boston" Scleral (itafluorofocon B) RGP Contact Lens for Daily Wear (predicate device) in the following key areas:
- · Device Design
- · Therapeutic indications for use
The BostonSight PD Prosthetic Device is substantially equivalent to the BostonSight IC Corneal & Scleral Lens (predicate device) in the following key areas:
- Components/Materials/Formulation
- · Manufacturing Process
- Manufacturing Facility
- Final Packaging & Wet Shipping
{8}------------------------------------------------
The BostonSight PD Prosthetic Device is substantially equivalent to the predicate device as depicted in the following table, and does not raise different questions of safety and effectiveness than the predicate device identified previously.
The following table depicts the pre-clinical characteristics of the BostonSight PD Prosthetic Device, as well as the predicate device(s).
| | BostonSight PD Prosthetic<br>Device<br>Subject Device | BostonSight IC Corneal<br>& Scleral Lens<br>(K153066)<br>Predicate Device | Boston® Scleral<br>(itafluorofocon B)<br>RGP Contact Lens<br>(P860022/S40)<br>Predicate Device |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | Indicated for therapeutic use in eyes with<br>ocular surface disease from dry eye (e.g.<br>ocular Graft-versus-Host disease,<br>Sjögren's syndrome, dry eye syndrome),<br>limbal stem cell deficiency (e.g.Stevens-<br>Johnson syndrome, chemical and thermal<br>burns, radiation), disorders of the skin<br>(e.g. atopy, ectodermal dysplasia),<br>neurotrophic keratitis (e.g. Herpes<br>simplex, Herpes zoster, Familial<br>Dysautonomia), and corneal exposure<br>(e.g. anatomic, paralytic) that might<br>benefit from the presence of an expanded<br>tear reservoir and protection against an<br>adverse environment. When prescribed<br>for therapeutic use for distorted cornea or<br>ocular surface disease, the BostonSight<br>PD Prosthetic Device may incidentally<br>provide correction of refractive error. | Indicated for the correction of<br>refractive error in aphakic and not<br>aphakic persons. The lenses may<br>be prescribed in otherwise non-<br>diseased eyes that require a rigid<br>gas permeable lens for the<br>management of irregular corneal<br>conditions such as keratoconus,<br>pellucid marginal degeneration or<br>following penetrating keratoplasty<br>or refractive (e.g. LASIK)<br>surgery. | Indicated for eyes having<br>significantly reduced vision<br>due to the presence of a<br>distorted corneal surface that:<br>1) precludes satisfactory<br>spectacle lens correction 2)<br>demonstrates significantly<br>improved rigid contact lens<br>corrected vision 3) is<br>incapable of wearing<br>traditional corneal lenses<br>because of the inability to<br>achieve adequate lens<br>centration/stability and/or<br>tolerance to physical contact<br>with a lens. Furthermore,<br>eyes suffering from certain<br>ocular surface disorders may<br>benefit from the physical<br>protection and the saline bath<br>provided by a scleral lens. |
| Device and<br>Classification | Class II<br>Lenses, Rigid Gas Permeable, Daily<br>Wear<br>HQD | Class II<br>Lenses, Rigid Gas Permeable,<br>Daily Wear<br>HQD | Class II<br>Lenses, Rigid Gas<br>Permeable, Daily Wear<br>HQD<br>(reclassified from Class II<br>to Class II in 1994) |
| Production Method | Lathe-cut | Lathe-cut | Lathe-cut |
| FDA Group # | Group # 3 Fluoro Silicone Acrylate | Group # 3 Fluoro Silicone<br>Acrylate | Group # 3 Fluoro Silicone<br>Acrylate |
| Water Content | <1% | <1% | <1% |
| UV Absorber/Blocker<br>available | YES | YES | YES |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.