BOSTON X02 (HEXAFOCON B) RIGID GAS PERMEABLE CONTACT LENS
K071266 · Bausch & Lomb, Inc. · HQD · Aug 15, 2007 · Ophthalmic
Device Facts
Record ID
K071266
Device Name
BOSTON X02 (HEXAFOCON B) RIGID GAS PERMEABLE CONTACT LENS
Applicant
Bausch & Lomb, Inc.
Product Code
HQD · Ophthalmic
Decision Date
Aug 15, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5916
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
BOSTON XO2 (hexafocon B) Gas Permeable Contact Lens is indicated for the daily wear correction of refractive ametropia (myopia, hyperopia, astigmatism, and presbyopia) in aphakic and not-aphakic persons with non-diseased eves. BOSTON XO2 may be prescribed in otherwise non-diseased eves that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK) surgery. BOSTON XO2 is also indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in nondiseased eyes. Note: To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule.
Device Story
Fluoro silicone acrylate copolymer rigid gas permeable (RGP) contact lens; daily wear; corrects refractive errors and manages irregular corneal conditions; used in orthokeratology for temporary myopia reduction. Lens material contains UV absorbers and color additives. Prescribed by eye care professionals; patient self-applied. Clinical performance evaluated against predicate Boston XO; demonstrated equivalent vision quality and tolerance. Benefits include vision correction and management of corneal irregularities.
Clinical Evidence
Controlled, multi-center clinical study comparing Boston XO2 to Boston XO in adapted GP wearers. Evaluated safety (adverse events, slit lamp findings) and efficacy (vision quality, comfort, symptoms). No adverse events reported; slit lamp findings < grade 2. Vision and tolerance attributes (comfort, dryness, irritation) were at least as good as the predicate.
Technological Characteristics
Fluoro silicone acrylate copolymer RGP lens material. Contains UV absorbers (Uvinul D-49 or MHB) and color additives (D&C Green No. 6, C.I. Solvent Yellow No. 18, D&C Violet No. 2, D&C Red No. 17). Biocompatibility tested per ISO 10993 (cytotoxicity, ocular irritation, systemic toxicity). Bioburden stability tested per ISO 11987.
Indications for Use
Indicated for daily wear correction of myopia, hyperopia, astigmatism, and presbyopia in aphakic/non-aphakic patients with non-diseased eyes. Also indicated for irregular corneal conditions (keratoconus, pellucid marginal degeneration, post-penetrating keratoplasty, post-refractive surgery) and orthokeratology for temporary reduction of myopia up to 5.00 diopters.
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
{1}------------------------------------------------
K071266
# 510(K) SUMMARY BOSTON XO2 (hexafocon B) Rigid Gas Permeable Contact Lenses for Daily Wear
AUG 15 2007
#### 1. Applicant's Name and Address
***
Bausch & Lomb, Inc. 1400 North Goodman Street Rochester, NY 14609
Debra Ketchum 2. Contact Person Manager, Global Regulatory Affairs Bausch & Lomb, Inc. 1400 North Goodman Street Rochester, NY 14609 (585) 338-8638 debra.ketchum@bausch.com
#### ર્જ ldentification of Device
| Common Name: | contact lens-rigid gas permeable (hydrophobic) |
|------------------------|-----------------------------------------------------|
| Trade Name: | Boston XO2 (hexafocon B) Daily Wear Contact<br>Lens |
| Classification: | Class II ophthalmic (21CFR 886.5916) |
| Device classification: | Class II (21 CFR 886.5916) |
| Product Codes: | HQD, MUW |
Predicate Devices:
The predicate device, Boston XO (hexafocon A) was selected to address material use and design (aspheric, bifocal, toric, irregular corneas, and orthkeratology).
The BOSTON XO2 is substantially equivalent to the currently marketed BOSTON XO cleared in 510(k) Premarket Notification Nos. K000795 on May 26, 2006, K001960 on August 28, 2000, K053124 on January 30, 2006 and K011945 on September 12, 2001. The difference between the two devices is a change in the component ratios.
{2}------------------------------------------------
| K000795, K001960, K043124 and K011945: Boston XO (hexafocon A) | | |
|----------------------------------------------------------------|----------------|----------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | Clearance Date | Device Description |
| K000795 | 5/26/2000 | Boston XO RGP Daily<br>Wear Contact Lenses |
| K001960 | 8/28/2000 | Boston XO RGP for Daily<br>Wear Orthokeratology<br>Contact Lenses |
| K053124 | 1/30/2006 | Boston XO, Boston EO,<br>Boston ES RGP Daily<br>Wear Contact Lenses<br>(Post Surgical) |
| K011945 | 9/12/2001 | Boston RGP Contact<br>Lenses wet shipped and<br>stored in Boston Advance<br>Comfort Formula<br>Conditioning Solution up<br>to 30 days. |
#### Description of device 4.
BOSTON XO2 (hexafocon B) Rigid Gas Permeable Contact Lens material is a fluoro silicone acrylate copolymer rigid gas permeable contact lens material. The material is available in blue, ice blue, green, or violet, yellow and red tint. The lenses may contain an ultraviolet light absorber, Uvinul D-49 or MHB. The blue and ice blue tinted lenses contain the color additive D&C Green No. 6; the green tinted lenses contain the color additives D&C Green No. 6 and C.I. Solvent Yellow No. 18; the violet tinted lenses contain the color additive D&C Violet No. 2. the red tinted lenses contain the color additive D&C Red No. 17; and the yellow lenses contain the color additive C.I. Solvent Yellow No.18. All of those color additives are listed in the Code of Federal Regulations (21 CFR 74.3206).
{3}------------------------------------------------
#### 5. Intended use
BOSTON XO2 (hexafocon B) Gas Permeable Contact Lens is indicated for the daily wear correction of refractive ametropia (myopia, hyperopia, astigmatism, and presbyopia) in aphakic and not-aphakic persons with non-diseased eves.
BOSTON XO2 may be prescribed in otherwise non-diseased eves that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK) surgery.
BOSTON XO2 is also indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in nondiseased eyes. Note: To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule.
#### 6. Description of Safety and Substantial Equivalence
A series of non-clinical laboratory testing and clinical testing was performed to demonstrate the safety and effectiveness of the BOSTON XO2 contact lens material. A summary of results from the non-clinical and clinical tests is provided below.
- 6.1. Lens Compatibility/Cycling Study
The compatibility of BOSTON XO2 material with several GP contact lens care regimens was assessed in this study. A lens cvcling study was completed with each contact lens care regimen. Initial and final data were compared. There were no significant changes to lens parameters after 30 complete cycles.
- 6.2. Determination of Total Extractables Substances (ISO 10340:1995(E))
Total Extractables testing was conducted on three lots of BOSTON XO2. The total amount of material extracted from BOSTON XO2 was less than that extracted from BOSTON XO the predicate device.
- 6.3. Determination of Extractable Color Additives and UV Blockers in Boston XO2 Material (ISO 10340:1995(E))
Determination of Extractable color additives and UV blockers testing was conducted on Boston XO2 material. These results are similar to those obtained when extracting Boston XO, the predicate device.
{4}------------------------------------------------
- 6.4. Cytotoxicity Study Using The ISO Agar Overlay Method (ISO 10993-5)
This in-vitro assay demonstrated the biocompatibility of this material. The test sample is placed on an agar surface directly overlaying a arowing L-929 cell monolayer to determine if the material induces a cvtopathic effect. The material did not induce cytotoxicity.
- 6.5. ISO Ocular Irritation Study (ISO 10993-10)
USP 0.9% Sodium Chloride Injection and sesame oil extracts of the Boston XO2 GP lens material was evaluated for the potential to produce an irritating effect on the ocular tissue of the rabbits. According to the criteria of the study protocol, the material is considered to be a nonirritant.
- 6.6. USP and ISO Systemic Toxicity Test (ISO 10993-11)
This study was designed to evaluate acute systemic toxicity of the BOSTON XO2 material extract following a single intravenous or intraperitoneal injection into mice. All observations of the mice were normal. The test samples meet the test requirements.
- 6.7. Bioburden of Lenses and Stability of Bioburden of Lenses Over time: Evaluation of the BOSTON XO2 material (ISO 11987:1997)
A 30-day shelf-life study for Boston XO2 lenses stored in Boston Advance Comfort Formula Conditioning Solution and Boston SIMPLUS Multi-Action solution was conducted. The average bioburden levels for BOSTON XO2 lenses stored in lens cases with Boston SIMPLUS Multi-Action Solution and Boston Advance Comfort Formula Conditioning Solution after 30 days were within the acceptable bioburden limit of < 100 cfullens.
#### Clinical data: 7.
The clinical investigation was conducted in accordance with the United States FDA regulations (21 CFR Parts 50, 54, 56 and 812), ICH Good Clinical Practices, ISO 14155-2 and the Declaration of Helsinki.
The objective of this controlled, multi-center clinical study was to evaluate the safety and efficacy of the BOSTON XO2 (hexafocon B) Gas Permeable (GP) contact lens compared to the BOSTON XO (hexafocon A) GP contact lens when worn by adapted GP contact lens wearers on a daily wear basis.
{5}------------------------------------------------
Safety Conclusions:
The use of the BOSTON XO2 (hexafocon B) gas permeable contact lens was well tolerated. There were no adverse events related to the BOSTON XO2 product.
The results of this evaluation demonstrated that treatment with the BOSTON XO2 (hexafocon B) qas permeable contact lens is safe when used as directed.
Efficacy Conclusions:
The vision provided by the BOSTON XO2 (hexafocon B) contact lens is the same as vision with the BOSTON XO (hexafocon A) contact lens. Other Symptoms/Complaint measures associated with lens tolerance (Comfort, End of Day Comfort, Burning/Stinging Upon Insertion, Irritation, Itching, Dryness, and Redness) ratings demonstrated that the BOSTON XO2 lens was at least as good as the BOSTON XO, overall, and at all follow-up visits for these attributes.
Other symptoms/complaints measures associated with lens tolerance (Comfort, End of Day Comfort, Burning/Stinging Upon Insertion, Irritation, Itching, Dryness, and Redness) ratings demonstrated that the BOSTON XO2 article was at least as good as the BOSTON XO, Overall, and at all follow-up visits for these attributes.
Discussion and Overall Conclusions:
The results of this study demonstrated that there were no major complications associated with the use of the BOSTON XO2 (hexafocon B) GP contact lens. This was supported by the fact that there were no adverse events, no test eves required medical treatment and there were no slit lamp findings greater than grade 2.
Vision measures were acceptable for the BOSTON XO2 (hexafocon B) GP contact lenses and were generally similar to the BOSTON XO contact lens. Symptoms/Complaints for various performance attributes associated with lens tolerance were also similar.
Overall, the investigators were satisfied with the vision quality that their subjects were able to obtain, the clinical performance of the lens met their expectations, and their subjects were successful wearing the BOSTON XO2 (hexafocon B) GP contact lenses.
{6}------------------------------------------------
### Conclusions drawn from studies 8. Substantial Equivalence:
Information provided in this 510(k) establishes that the BOSTON XO2 GP Contact Lens is equivalent in optical, chemical, and physical properties of the predicate device and does not raise any questions of safety and effectiveness. The clinical evaluation demonstrated safe and effective lens performance and equivalence with the predicate device.
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 15 2007
Bausch & Lomb, Inc. c/o Debra Ketchum Manager, Global Regulatory Affairs 1400 North Goodman Street Rochester NY 14609-3547
Re: K071266
Trade/Device Name: Boston XO2 (hexafocon B) Daily Wear Gas Permeable Contact Lens Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid gas permeable contact lens Regulatory Class: Class II Product Code: HQD, MUW Dated: July 19, 2007 Received: July 23, 2007
Dear Ms. Ketchum:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roved your be determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for associated in the each as 1976, the enactment date of the Medical Device Amendments, or to conninered prior to rialy 20, 2012 - 12:25 pm accordance with the provisions of the Federal Food, Drug, de vices that have been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, therefore, mains of the Act include requirements for annual registration, listing of general obtaturing province, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elassined (600 as a votrols. Existing major regulations affecting your device can may or subject to back as assisted Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I lease be advised that I Dri bristian that your device complies with other requirements of the Act that I Dri has made a averal regulations administered by other Federal agencies. You must or any I cuena statuted and registments, including, but not limited to: registration and listing (21 Comply with an the 110 - 210 - 21 - 21 - 2011; good manufacturing practice requirements as set forth in the quality systems (21 CFR Part 820); and if applicable, the electronic forth in the quality by sections (Sections 531-542 of the Act); 21 CFR 1000-1050.
{8}------------------------------------------------
## Page 2 - Debra Ketchum
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M.B. Egeland, MD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
1071266
510(k) Number (if known): __
Device Name:
Boston XO2 (hexafocon B)
## Indication for Use
The BOSTON XO2 Gas Permeable Contact Lens is indicated for the daily wear correction of refractive ametropia (myopia, astigmatism, and presbyopia) in aphakic and notaphakic persons with non-diseased eyes.
The lens may be prescribed in otherwise non-diseased eyes that require a rigid contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g. LASIK) surgery.
The lens is also indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. Note: To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | |
|------------------|--|
|------------------|--|
OR
| Over-the-counter-use | |
|----------------------|--|
|----------------------|--|
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises
| 510(k) Number | K071266 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.