K173371 · Carl Zeiss Meditec, Inc. · HNO · Apr 13, 2018 · Ophthalmic
Device Facts
Record ID
K173371
Device Name
VisuMax Femtosecond Laser
Applicant
Carl Zeiss Meditec, Inc.
Product Code
HNO · Ophthalmic
Decision Date
Apr 13, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4370
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The VisuMax Femtosecond Laser is cleared for the following indications for use: · In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea; · In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea; · In the creation of a lamellar cut/resection of the cornea for lamellar keratoplasty; · In the creation of a cut/incision for penetrating keratoplasty and corneal harvesting; · In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea to create tunnels for placement of corneal ring segments.
Device Story
Ophthalmic surgical femtosecond laser; creates corneal incisions via precise micro-photodisruptions. System uses ultrashort laser pulses delivered through disposable applanation lens while eye is fixated under low vacuum. Operated by clinicians in surgical settings. Software modifications enable user-selected tunnel cut parameters via graphical user interface (GUI) for corneal ring segment placement. Output is precise tissue resection; assists surgeons in performing corneal procedures; benefits patients by enabling minimally invasive corneal surgeries.
Clinical Evidence
Bench testing only. Accuracy and repeatability of corneal tunnel lateral and depth dimensions verified using non-contact optical techniques in ex vivo corneas and corneal model materials. Cut quality evaluated in ex vivo corneas. All performance test acceptance criteria met.
Technological Characteristics
Femtosecond laser; 1043 nm wavelength; 500 kHz repetition rate; 0.375 uJ max pulse energy. Scanned pattern photodisruption. Complies with IEC 60601-1, 60601-1-2, 60601-1-6, 60601-2-22, 60825-1, 62304, and 62366.
Indications for Use
Indicated for patients undergoing ophthalmic surgery requiring corneal incisions, including LASIK flap creation, lamellar keratoplasty, penetrating keratoplasty, corneal harvesting, and tunnel creation for corneal ring segments.
Regulatory Classification
Identification
A keratome is an AC-powered or battery-powered device intended to shave tissue from sections of the cornea for a lamellar (partial thickness) transplant.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 13, 2018
Carl Zeiss Meditec, Inc. Todd Otani Sr. Clinical Research Scientist 5160 Hacienda Drive Dublin, CA 94568
Re: K173371
Trade/Device Name: VisuMax Femtosecond Laser Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HOF, HNO Dated: March 13, 2018 Received: March 14, 2018
Dear Todd Otani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours,
Kesia Alexander
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173371
Device Name VisuMax Femtosecond Laser
Indications for Use (Describe)
The VisuMax Femtosecond Laser is cleared for the following indications for use:
· In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea;
· In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea;
· In the creation of a lamellar cut/resection of the cornea for lamellar keratoplasty;
· In the creation of a cut/incision for penetrating keratoplasty and corneal harvesting;
· In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea to create tunnels for placement of corneal ring segments.
| Type of Use ( <i>Select one or both, as applicable</i> ) |
|----------------------------------------------------------|
|----------------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(a).
### SUBMITTER'S NAME, ADDRESS, TELEPHONE, CONTACT PERSON AND DATE SUMMARY PREPARED
| a. | Applicant: | Carl Zeiss Meditec AG<br>Goeschwitzer Strasse 51-52<br>7740 Jena<br>Germany |
|----|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| b. | Contact Person: | Todd Otani, O.D.<br>Carl Zeiss Meditec, Inc.<br>Zeiss Group<br>5160 Hacienda Drive<br>Dublin, California 94568, USA<br>t.otani@zeiss.com<br>Tel: (925) 557-4513<br>Fax: (925) 557-4259 |
| c. | Date: | 2018-04-12 |
### NAME OF DEVICE, TRADE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | VisuMax Femtosecond Laser |
|-------------------------|----------------------------------------------------|
| Common/Usual Name: | Femtosecond Laser |
| Classification Name: | Laser, Ophthalmic<br>Keratome, AC-powered |
| Classification Code(s): | 21 CFR 886.4390; 79 HQF<br>21 CFR 886.4370; 79 HNO |
#### PREDICATE DEVICE(S)
| PREDICATE DEVICE | MANUFACTURER | 510(K) CLEARANCE |
|------------------------|-----------------------|------------------|
| VisuMax Laser Keratome | Carl Zeiss Meditec | K100253 |
| WaveLight FS200 | Alcon Novartis | K141476 |
| IntraLase iFS Laser | Abbott Medical Optics | K141852 |
#### DEVICE DESCRIPTION
The VisuMax Femtosecond Laser is an ophthalmic surgical femtosecond laser intended for use in patients requiring corneal incisions. The cutting action of the VisuMax laser is achieved through precise individual micro-photodisruptions of tissue, created by tightly focused ultrashort pulses which are delivered through a disposable applanation lens while fixating the eye under very low vacuum.
{4}------------------------------------------------
## INDICATIONS FOR USE
The VisuMax Femtosecond Laser is cleared for the following indications for use:
- . In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea;
- . In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea;
- . In the creation of a lamellar cut/resection of the cornea for lamellar keratoplasty:
- . In the creation of a cut/incision for penetrating keratoplasty and corneal harvesting;
- . In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea to create tunnels for placement of corneal ring segments.
## TECHNOLOGICAL CHARACTERISTICS COMPARISON
The subject VisuMax Femtosecond Laser is essentially the same device as the VisuMax Laser Keratome predicate device (K100253) with the addition of software modifications made to enable the expansion of the indications for use and minor modifications made to meet manufacturability and service requirements.
The VisuMax subject and the predicate laser systems share operating principle and mode of operation whereby the creation of tissue resections are produced by scanned patterns of femtosecond laser micro-photodisruptions in cornea.
The subject VisuMax Femtosecond Laser is substantially equivalent to the predicate laser systems presented in this 510(k) premarket notification in terms of indications for use, technological characteristics and fundamental scientific technology. No changes to the VisuMax Femtosecond Laser system specifications, system design, or hardware control system were made to implement the expanded indication for use. The laser parameters in the subject device are identical to those of the VisuMax predicate device (K100253).
As seen in the Technological Comparison table below, the subject device is identical to the primary predicate device with the exception of the software modifications implemented to enable the new indication for use. Differences between the VisuMax and the secondary predicate devices are limited to the maximum repetition rate and the maximum pulse energy and do not represent a difference in the fundamental technology of this device. Testing confirmed that the VisuMax performance is equivalent to that of the predicate devices.
Software modifications implemented to support the expanded indications (i.e. creation of tunnel cuts for the placement of corneal ring segments), allow the VisuMax to produce tunnel cuts equivalent to those produced by predicate devices K141852 and to allow the user based tunnel cut parameter selection on the graphical user interface (GUI).
{5}------------------------------------------------
The tunnel cutting functionality was shown to be substantially equivalent to the functionality of the predicate devices through a comparison of the relevant technical specifications, characteristics and features for the corneal ring segment cutting procedure.
{6}------------------------------------------------
| 510(k) clearance | K173371 (proposed) | K100253 (primary predicate) | K141852 (secondary predicate) | K141476 (secondary predicate) |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | VisuMax Femtosecond Laser | VisuMax Laser Keratome | iFS Laser System | WaveLight FS200 |
| Manufacturer | Carl Zeiss Meditec | Carl Zeiss Meditec | Abbott Medical Optics | Alcon Novartis |
| Indications for<br>Use | The VisuMax Femtosecond Laser is<br>indicated for the following:<br>• In the creation of a corneal flap in<br>patients undergoing LASIK surgery<br>or other treatment requiring initial<br>lamellar resection of the cornea;<br>• In patients undergoing surgery or<br>other treatment requiring initial<br>lamellar resection of the cornea;<br>• In the creation of a lamellar<br>cut/resection of the cornea for<br>lamellar keratoplasty;<br>• In the creation of a cut/incision for<br>penetrating keratoplasty and<br>corneal harvesting<br>• In patients undergoing surgery or<br>other treatment requiring initial<br>lamellar resection of the cornea to<br>create tunnels for placement of<br>corneal ring segments. | The VisuMax Laser Keratome is<br>indicated for the following:<br>• In the creation of a corneal flap in<br>patients undergoing LASIK surgery<br>or other treatment requiring initial<br>lamellar resection of the cornea;<br>• In patients undergoing surgery or<br>other treatment requiring initial<br>lamellar resection of the cornea;<br>• In the creation of a lamellar<br>cut/resection of the cornea for<br>lamellar keratoplasty;<br>• In the creation of a cut/incision for<br>penetrating keratoplasty and<br>corneal harvesting | The iFS Laser System is an<br>ophthalmic surgical laser designed<br>for use as an ophthalmic surgical<br>laser indicated for use as follows:<br>• In patients undergoing surgery or<br>other treatment requiring initial<br>lamellar resection of the cornea<br>• In patients undergoing surgery or<br>other treatment requiring initial<br>lamellar resection of the cornea to<br>create tunnels for placement of<br>corneal ring segments<br>• In lamellar keratoplasty and corneal<br>harvesting<br>• In the creation of a corneal flap in<br>patients undergoing LASIK surgery<br>or other treatment requiring initial<br>lamellar resection of the cornea<br>• In the creation of a lamellar<br>cut/resection of the cornea for<br>lamellar keratoplasty<br>• In the creation of lamellar<br>cut/resection of the cornea for<br>lamellar keratoplasty and in the<br>creation of penetrating cut/incision<br>for penetrating keratoplasty. | The WaveLight FS200 Laser System<br>is an ophthalmic surgical laser<br>indicated for use:<br>• In the creation of a corneal flap in<br>patients undergoing LASIK surgery<br>or other surgery or treatment<br>requiring initial lamellar resection<br>of the cornea.<br>• In patients undergoing surgery or<br>other treatment requiring initial<br>lamellar resection of the cornea to<br>create tunnels for placement of<br>corneal ring segments.<br>• In the creation of a lamellar<br>cut/resection of the cornea for<br>lamellar keratoplasty.<br>• In the creation of a penetrating<br>cut/incision for penetrating<br>keratoplasty and for corneal<br>harvesting.<br>• In patients undergoing ophthalmic<br>surgery or other treatment requiring<br>arcuate cuts/incisions in the cornea.<br>• In patients undergoing ophthalmic<br>surgery or other treatment requiring<br>pocket cuts/incisions in the cornea. |
| Operating<br>principle | Femtosecond laser photodisruption | Femtosecond laser photodisruption | Femtosecond laser photodisruption | Femtosecond laser photodisruption |
| Mechanism of<br>action | Cutting surfaces created by scanned<br>patterns of femtosecond laser micro-<br>photodisruptions in cornea. | Cutting surfaces created by scanned<br>patterns of femtosecond laser micro-<br>photodisruptions in cornea. | Cutting surfaces created by scanned<br>patterns of femtosecond laser micro-<br>photodisruptions in cornea. | Cutting surfaces created by scanned<br>patterns of femtosecond laser micro-<br>photodisruptions in cornea. |
| Wavelength | 1043 nm | 1043 nm | 1053 nm | 1030 nm |
| Repetition rate | 500 kHz | 500 kHz | 60 kHz | 200 kHz |
| Pulse energy, max | 0.375 uJ | 0.375 uJ | 2.50 uJ | 2.40 uJ |
## TECHNOLOGICAL COMPARISON VISUMAX FEMTOSECOND LASER & PREDICATE DEVICES
{7}------------------------------------------------
## PERFORMANCE DATA
Performance data supporting substantial equivalence of the modified VisuMax Femtosecond Laser and the predicate devices included the following:
- . The accuracy and repeatability of corneal tunnel lateral dimensions were verified using a non-contact optical technique in ex vivo corneas. Tunnel cuts were made in a series of corneal models using a range of cut parameters spanning the VisuMax tunnel cut parameter ranges. Test acceptance criteria for all cut length dimension and cut angular dimensions were met.
- . The accuracy and repeatability of corneal tunnel depth dimensions were verified using a non-contact optical technique in a corneal model material. Cut depth measurements in the model material validated by comparing depth measurements made in both ex vivo corneas and the model material. Tunnel cuts were made in a series of corneal models using a range of cut parameters spanning the VisuMax tunnel cut parameter ranges. Test acceptance criteria for all cut depth dimensions were met.
- . Cut quality for the tunnel cuts was evaluated in ex vivo corneas for ease of tissue separation following the creation of the access and tunnel cuts. All corneas tested in this manner were judged to be of good cut quality, meeting the performance test acceptance criteria.
In summary, performance testing demonstrated that the tunnel cut features for the expanded indications for use by the VisuMax subject device are equivalent to the predicate devices in terms of cut dimension accuracy, cut repeatability and cut quality.
The VisuMax Femtosecond Laser also underwent medical electrical equipment testing and was found to be in compliance with the following applicable international safety standards:
- . IEC 60601-1 -- Medical Electric equipment, Part 1: General Requirements for Safety
- . IEC 60601-1-2 -- Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility
- IEC 60601-1-6 -- Medical electrical equipment Part 1-6: General requirements for safety -● Collateral Standard: Usability
- . IEC 60601-2-22 -- Medical Electrical Equipment, Part 2-22: Particular Requirements for basic Safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
- . IEC 60825-1 -- Safety of Laser Products, Part 1: Equipment Classification and Requirements
- IEC 62304 -- Medical Device Software - Software Life Cycle Processes
- IEC 62366 -- Medical devices Application of usability engineering to medical . devices
{8}------------------------------------------------
#### SOFTWARE VERIFICATION AND VALIDATION
The software for the VisuMax Femtosecond Laser is considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injurv or death to the patient or operator.
Software verification and validation testing were conducted and documentation was provided as recommended by FDA Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Verification and validation testing of the VisuMax software has been performed for both the system software and all associated software modules. The software verification and validation testing results demonstrate that the VisuMax Femtosecond Laser meets all requirements for performance, accuracy, functionality and safety for the modifications proposed in this 510(k) premarket notification.
### CONCLUSION
The purpose of this 510(k) Premarket Notification is to obtain clearance for an additional new indication for use for the creation of tunnel cuts for the placement of corneal ring segments for the VisuMax Femtosecond Laser. The VisuMax Femtosecond Laser is essentially the same as the predicate VisuMax Laser Keratome, with software modifications that enable the new indication for use. These modifications result in a VisuMax Femtosecond Laser that is substantially equivalent to the other predicate devices included in this 510(k) Premarket Notification. The supportive information provided in this 510(k) Premarket Notification provides reasonable assurance that the VisuMax Femtosecond Laser is safe and effective for its intended use and that it is substantially equivalent to the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.